NA 04.033.270SRecalls & reported problems
Synthes GmbH · 12MM / TI CANN FRN / GT 400MM / RIGHT - STERILE · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886982271815), received between Oct 8, 2019 and Jul 7, 2022 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
- Device-Device Incompatibility1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- Nonunion/Delayed-union Bone Fracture1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJul 7, 2022Break; Break
IT WAS REPORTED THAT ON (B)(6) 2022, THE REMOVAL OF BROKEN NAIL DUE TO NONUNION. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH NO DELAY. NO FRAGMENTS WERE GENERATED. THERE IS NO FURTHER INFORMATION. THIS REPORT IS FOR ONE (1) 12 / TI CANN FRN / GT 400 / RIGHT - SILE. THIS IS REPORT 1 OF 5 FOR COMPLAINT (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 14950160 ↗MalfunctionOct 15, 2021Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT FOR A SURGERY. DURING THE SURGERY, WHILE ATTEMPTING TO INSERT RECON SCREWS INTO AN UNKNOWN FEMORAL RECON NAIL (FRN), THE SURGEON HIT THE NAIL WITH THE DRILL BIT. THE NAIL WAS UNSTABLE BECAUSE THE SURGEON DID NOT PIN BOTH TROCARS BEFORE REMOVING GUIDE WIRE TO DRILL THE FIRST HOLE. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THIS COMPLAINT INVOLVES AN UNKNOWN NUMBER OF DEVICES. THIS REPORT IS FOR (1) 12MM / TI CANN FRN / GT 400MM / RIGHT - STERILE. THIS REPORT IS 1 OF 10 OF (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12641953 ↗InjuryOct 8, 2019Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN OPEN REDUCTION INTERNAL FIXATION (ORIF) PROCEDURE DUE TO A RIGHT SIDE MIDSHAFT FEMUR FRACTURE ON (B)(6) 2019, THE SURGEON HAD DIFFICULTY INSERTING THE GREATER TROCHANTER FEMORAL RECON NAIL. DURING THE SURGERY, THE CANAL WAS REAMED 2.5 MM OVER THE NAIL SIZE BUT THE SURGEON STILL COULD NOT INSERT THE NAIL. SUBSEQUENTLY, THE ENTRY POINT WAS EVALUATED AND APPEARED TO BE PERFECT. THE NAIL WAS HAMMERED DOWN, BUT IT WAS APPROXIMATELY ONE-THIRD OF THE WAY DOWN, AND NOT REACHING THE ISTHMUS BEFORE THE NAIL WOULD IMPINGE ON BOTH THE LATERAL AND MEDIAL CORTEX. HOWEVER, IT APPEARED ON X-RAY AS IF THE DISTAL THIRD PORTION OF THE NAIL WAS TOO STRAIGHT TO GET THROUGH THE ANGLE NECESSARY WITH A GREATER TROCHE ENTRY. EVENTUALLY, THE SURGEON REMOVED THE NAIL AND USED A SMALLER DIAMETER (11 MM) NAIL. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A THIRTY (30) MINUTE SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. CONCOMITANT DEVICES REPORTED; UNKNOWN INSERTION HANDLE (PART# /LOT# UNKNOWN; QUANTITY: UNKNOWN), UNKNOWN DRIVING CAP (PART# /LOT# UNKNOWN; QUANTITY: UNKNOWN), UNKNOWN FEMORAL NAIL (PART# /LOT# UNKNOWN; QUANTITY: 1). THIS COMPLAINT INVOLVES ONE (1) DE
Patient problems: No Code Available
MAUDE report 9168744 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page