NA 04.033.966SRecalls & reported problems
Synthes GmbH · 9 MM / TI CANN FRN / GT 360MM / RIGHT - STERILE · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886982272225), received between Dec 18, 2018 and May 6, 2024 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Break1
- Difficult to Advance1
- Material Twisted/Bent1
Patient problems
- No Code Available2
- Nonunion/Delayed-union Bone Fracture2
- Failure of Implant1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryMay 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2023, THE PATIENT UNDERWENT ORIF WITH A NAIL IMPLANT FOR RIGHT FEMORAL SHAFT FRACTURE. ON APRIL 10, 2024, THE SALES REP RECEIVED A REPORT FROM THE AGENCY AND THE SURGEON THAT A PATIENT WITH A RIGHT FEMORAL SHAFT FRACTURE WHO WAS OSTEOSYNTHESIZED USING THE PRODUCT BECAME A FALSE JOINT. (B)(6) 2024, THE NAIL IMPLANT REMOVAL AND REFIXATION ENTAILING PLATE WERE EXPECTED TO BE PERFORMED. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) 9 MM / TI CANN FRN / GT 360MM / RIGHT - STERILE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 19256043 ↗InjuryDec 14, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE PATIENT UNDERWENT ORIF SURGERY FOR A FEMORAL DIAPHYSEAL FRACTURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. AFTER THE ORIGINAL PROCEDURE ON (B)(6) 2023, IT WAS CONFIRMED THAT THE DISTAL FRACTURE LINE BECAME A PSEUDOARTICULATION DESPITE A GOOD RECOVERY POSITION. NAIL REVISION SURGERY WAS SCHEDULED FOR (B)(6) 2023. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 9 MM / TI CANN FRN / GT 360MM / RIGHT ¿ STERILE. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 18330244 ↗InjuryFeb 10, 2020Break; Break
IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT UNDERWENT REVISION SURGERY DUE TO BROKEN FEMORAL RECON NAIL (FRN). ORIGINALLY, THE PATIENT HAD AN IMPLANTATION ON AN UNKNOWN DATE. POST-SURGERY AFTER A MONTH, THE LEG WAS NOT HEALED. WHEN THE NAIL BROKE AND THE LEG WAS STILL BROKEN. THE IMPLANT WAS SUCCESSFULLY EXPLANTED WITHOUT ANY ISSUE. THERE WAS NO SURGICAL DELAY. THE PATIENT WAS STABLE AFTER THE PROCEDURE. PROCEDURE AND PATIENT OUTCOME ARE UNKNOWN. CONCOMITANT DEVICES REPORTED: UNKNOWN LOCKING SCREWS (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # UNKNOWN), UNKNOWN RECON SCREWS (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY# UNKNOWN), UNKNOWN END CAP (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) 9 MM / TI CANN FRN / GT 360MM / RIGHT - STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Failure of Implant; No Code Available
MAUDE report 9687296 ↗InjuryJul 22, 2019Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT REVISION SURGERY. THE PATIENT PUT WEIGHT ON THE IMPLANT WHICH CAUSED FEMORAL RECON NAIL GREATER TROCHANTER TO BEND DISTALLY, AS CLAIMED BY THE SURGEON PATIENT WAS NON COMPLIANT. THERE WERE NO BROKEN IMPLANTS. THERE WAS A SURGICAL DELAY OF UNKNOWN NUMBER OF MINUTES AS THE SURGEON HAD TO MANIPULATE THE LEG TO BEND THE NAIL BACK TO ORIGINAL SHAPE FOR REMOVAL. PROCEDURE OUTCOME IS SUCCESSFUL. PATIENT STATUS IS STABLE. CONCOMITANT DEVICE REPORTED: UNKNOWN LOCKING SCREWS (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1), UNKNOWN RECON SCREWS (PART# UNKNOWN, LOT # UNKNOWN, QUANTITY # UNKNOWN). THIS REPORT IS FOR ONE (1) 9 MM TI CANNULATED FEMORAL RECON NAIL. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Code Available
MAUDE report 8812850 ↗MalfunctionDec 18, 2018Difficult to Advance; Difficult to Advance
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING AN UNKNOWN SURGERY, THE SURGEON HAD DIFFICULTY INSERTING THE TITANIUM CANNULATED FEMUR RECON NAIL. THE SURGEON COULD NOT ADVANCE UNTIL HE REAMED THE NAIL FURTHER DISTAL TO THE LESSER TROCHANTER RESULTING IN IT GETTING STUCK. THE SURGEON USED AN UNKNOWN 14MM REAMER TO SEAT THE NAIL FLUSH BUT DID NOT WANT TO PULL IT OUT AGAIN AND REAM FURTHER. THE SURGEON DID NOT WANT TO SPLIT THE FEMUR WHILE TRYING TO HAMMER IT IN FURTHER. IT WAS FURTHER REPORTED THAT THE TAPER OF THE DRIVING END EXTENDS FAR INTO THE DIAPHYSIS TO ACCOMMODATE THE RECON SCREW FEATURES, REQUIRES EXCESSIVE DISTAL REAMING BEYOND THE STANDARD FOR A STARTING REAMER. IT IS AN UNKNOWN IF THERE WAS SURGICAL DELAY. PROCEDURE AND THE PATIENT OUTCOME ARE UNKNOWN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8176250 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page