NA 03.404.025SRecalls & reported problems
Synthes GmbH · 14.5MM REAMER HEAD FOR RIA 2 STERILE · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982274083), received between Nov 4, 2020 and Dec 11, 2020 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Nov 19, 2020Class IITerminated
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed
Z-0580-2021 · Root cause: Use error · Recall details · FDA record ↗
Most reported problems
Device problems
- Break2
- Appropriate Device Problem Term/Code Not Available1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 11, 2020Break; Appropriate Device Problem Term/Code Not Available; Break; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT ON (B)(6) 2020, DURING AN UNKNOWN PROCEDURE, THE SURGEON WAS USING THE REMER IRRIGATOR ASPIRATOR (RIA) SYSTEM AND THE ASSEMBLY BROKE OFF DURING THE REAMING PROCESS. THE SURGEON WAS UNABLE TO COMPLETE HARVESTING AND OBTAINED NO GRAFT FOR USE. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. THE SURGERY ENDED FINE AND THE PATIENT WAS STABLE. THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS FOR (1) 14.5MM REAMER HEAD FOR RIA 2 STERILE. THIS REPORT IS 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10993525 ↗MalfunctionNov 4, 2020Break; Break
IT WAS REPORTED THAT ON (B)(6) 2020, WHILE REAMING WITH A RIA2 FOR BONE GRAFT, THE REAMER HEAD FOR RIA 2 TIP BROKE WHILE TAKING THE REAMER BACK OUT AFTER THE FIRST PASS. NO UNDUE TORQUE NOR ANGULATION WAS APPLIED SINCE IT WAS AN INTACT FEMUR. IT IS UNKNOWN IF THERE WAS SURGICAL DELAY. PATIENT AND PROCEDURE OUTCOME IS UNKNOWN. THIS REPORT IS FOR ONE (1) 14.5MM REAMER HEAD FOR RIA 2 STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10786893 ↗
Other REAMER devices
- QUICKSET 2440-00-551Depuy (Ireland) · REAMER1,150 reports
- QUICKSET 2440-00-549Depuy (Ireland) · REAMER1,113 reports
- QUICKSET 2440-00-553Depuy (Ireland) · REAMER1,105 reports
- QUICKSET 2440-00-547Depuy (Ireland) · REAMER1,073 reports
- QUICKSET 2440-00-555Depuy (Ireland) · REAMER1,065 reports
- QUICKSET 2440-00-557Depuy (Ireland) · REAMER1,031 reports
- QUICKSET 2440-00-545Depuy (Ireland) · REAMER947 reports
- QUICKSET 2440-00-552Depuy (Ireland) · REAMER888 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page