NA 03.168.003Recalls & reported problems
Synthes GmbH · DIRECT MEASURING DEVICE FOR 3.2MM GUIDE WIRES · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10886982274502), received between Apr 24, 2018 and Mar 3, 2026 — classified as 2 injuries, 9 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Incorrect Measurement3
- Device-Device Incompatibility2
- Contamination /Decontamination Problem1
- Crack1
- Device Operates Differently Than Expected1
- Improper or Incorrect Procedure or Method1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Injury2
- Insufficient Information1
- No Patient Involvement1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 3, 2026Incorrect Measurement
IT WAS REPORTED THAT THE DIRECT MEASUR-DEVICE F/GUIDEWIRES Ø3.2 HAD CONCERNS REGARDING INCONSISTENT MEASUREMENT RESULTS ACROSS DIFFERENT PROCEDURES. THE SURGEON NOTED FRUSTRATION WITH THE DEVICE'S VARYING MEASUREMENT METHODS, WHICH REQUIRE MANUAL ADJUSTMENTS AND CAN LEAD TO CONFUSION AND IMPLANT WASTAGE DURING OPERATIONS. THE REPORTED FAILURE INVOLVED INCORRECT MEASUREMENT FUNCTIONALITY, BUT THERE WAS NO PATIENT HARM OR SIGNIFICANT DELAY ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24497449 ↗MalfunctionMar 3, 2026Incorrect Measurement
IT WAS REPORTED THAT THE DIRECT MEASUR-DEVICE F/GUIDEWIRES Ø3.2 HAD CONCERNS REGARDING INCONSISTENT MEASUREMENT RESULTS ACROSS DIFFERENT PROCEDURES. THE SURGEON NOTED FRUSTRATION WITH THE DEVICE'S VARYING MEASUREMENT METHODS, WHICH REQUIRE MANUAL ADJUSTMENTS AND CAN LEAD TO CONFUSION AND IMPLANT WASTAGE DURING OPERATIONS. THE REPORTED FAILURE INVOLVED INCORRECT MEASUREMENT FUNCTIONALITY, BUT THERE WAS NO PATIENT HARM OR SIGNIFICANT DELAY ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24497443 ↗MalfunctionJan 15, 2026Break
DEPTH GAUGE IS BROKEN AT THE END AND CANNOT PROVIDE ACCURATE MEASUREMENT FOR THE SURGEON. THERE WAS NO PATIENT OR USER HARM. THERE WAS NO SIGNIFICANT DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24085803 ↗MalfunctionAug 4, 2025Break; Break
IT WAS REPORTED DURING A PROCEDURE THE FEMORAL NECK SYSTEM (FSN) MISSED THE TARGETED COMPLETELY; THE DRILL WAS SET TO 95MM BUT OVER DRILLED TO 100MM. THERE WAS NO IMPACT ON THE PATIENT. IT WAS NOTED THERE WAS RUST ON THE FSN ITEMS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22701535 ↗MalfunctionJul 29, 2025Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE THERE WAS RUST ON FEMORAL NECK SYSTEM (FNS) ITEMS, AND NEED REPLACING DUE TO FNS DRILL BIT HITTING. THE X2 CANNULATED SCREWS BROKE WHILE INSERTING IN FEMORAL NECK. THE SCREW (03.354.460) BROKE IN FEMORAL NECK. IT SNAPPED IN HALF WHILE INSERTING. THE FNS MISSED THE TARGETER COMPLETELY AND WHENEVER WE SET THE DRILL TO 95MM IT OVER DRILLED TO 100MM. IT KEPT SLIDING FOR SOME REASON. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS RELATED TO (B)(4) WHICH CAPTURES ADDITIONAL REPORTED DEVICES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22651253 ↗MalfunctionMar 19, 2024Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE THE TIP OF DEPTH GAUGE BROKEN OFF. THE PATIENT OUTCOME WAS UNKNOWN. THIS REPORT IS FOR ONE DIRECT MEASURING DEVICE 3.2 GUIDE WIRES. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 18932193 ↗MalfunctionApr 26, 2023Crack; Contamination /Decontamination Problem; Device-Device Incompatibility; Crack; Contamination /Decontamination Problem; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, INSTRUMENT WAS UNABLE TO STERILIZE/ASSEMBLE. CHIPPING AND CRACKING. UNABLE TO USE. NEED REPLACEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR ONE (1) DIRECT MEASURING DEVICE 3.2 GUIDE WIRES THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16818666 ↗MalfunctionJan 26, 2023Incorrect Measurement; Incorrect Measurement
THERE WAS A 15-MINUTE DELAY IN RELATION TO THE EVENT. THE SURGERY WAS COMPLETED SUCCESSFULLY, AND THE PATIENT STATUS/OUTCOME WAS SATISFACTORY. NO OTHER MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16246953 ↗InjuryMay 20, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING FEMORAL NECK SYSTEM (FNS) PROCEDURE THE SURGEON PENETRATED THE FEMORAL HEAD WITH THE TRIPLE REAMER. THE SURGEON MEASURED IT WITH 87MM AND SET THE REAMER TO 80MM AND THE REAMER STILL PENETRATED TO THE HEAD. THE TRIPLE REAMER WAS ADVANCED FARTHER INTO FEMORAL HEAD THAN INTENDED AND POSSIBILITY EXISTS. THE REAMER WAS POSSIBLY SET TO A LENGTH THAT WAS TOO LONG BUT THAT WAS NOT CONFIRMED AND DID NOT OCCUR AT THE TIME OF THE PROCEDURE. ACCORDING TO THE SURGEON, PATIENT IS PAIN FREE AND WANTS TO WEIGHT BEAR. THE SURGEON IS TREATING THE PATIENT CONSERVATIVELY FOR NOW BUT PLANS TO ALLOW WEIGHT BEARING SOON. THERE WAS NO SURGICAL DELAY. PATIENT AND PROCEDURE OUTCOME WERE UNKNOWN. CONCOMITANT DEVICE REPORTED: BOLT FOR FEMORAL NECK SYSTEM 90MM LENGTH-STERILE (PART NUMBER 04.168.290S, LOT UNKNOWN, QUANTITY 1), COMPLETE OPENING DRILL BIT/ REAMER ASSEMBLY (PART NUMBER 03.168.004, LOT UNKNOWN, QUANTITY 1). THIS REPORT INVOLVES ONE (1) DIRECT MEASURING DEVICE FOR 3.2MM GUIDE WIRES. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: Injury
MAUDE report 10074299 ↗MalfunctionNov 21, 2019Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PRE-OPERATIVELY ON (B)(6) 2019 AN INSERT FOR FEMORAL NECK SYSTEM (FNS) INSERTION HANDLE WAS STUCK AND UNRETRIEVABLE FROM A DIRECT MEASURING DEVICE FOR 3.2MM GUIDE WIRES. THE SURGICAL TECHNICIAN INCORRECTLY ASSEMBLED (WITHOUT FORCE) THE PIECES AND THEY BECAME COLD-WELDED TOGETHER. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) DIRECT MEASURING DEVICE FOR 3.2MM GUIDE WIRES. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 9357223 ↗InjuryApr 24, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected
DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE SURGEON USED THE FEMORAL NECK SYSTEM TO FIX A FEMORAL NECK FRACTURE. THE DESIRED IMPLANT LENGTH WAS NOT AVAILABLE IN THE HOSPITAL, SO HE DECIDED TO USE A LONGER IMPLANT. THE OPERATING ROOM (OR) NURSE SET UP THE REAMER TO DRILL FOR THE CHOSEN LONGER IMPLANT LENGTH AND NOT FOR THE ORIGINALLY SHORTER MEASURED LENGTH. REAMER WAS STOPPED AFTER HAVING PENETRATED THE HIP JOINT. CONCOMITANT DEVICE: FEMORAL NECK SYSTEM (PART 04.168.100S, LOT UNKNOWN, QUANTITY 1). THIS IS REPORT 1 OF 1 FOR (B)(4) .
Patient problems: Injury
MAUDE report 7454469 ↗
FDA clearance and classification
The FDA regulates product code HTJ (gauge, depth) under 21 CFR 888.4300 as a Class I device. 21 CFR 888.4300 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 03.168.003
Has the NA 03.168.003 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886982274502), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 03.168.003?
The FDA's MAUDE database lists 11 adverse event reports for this device, received between Apr 24, 2018 and Mar 3, 2026. In those reports, the most common specific device problems are “Break” (4), “Incorrect Measurement” (3) and “Device-Device Incompatibility” (2); the most common patient problem is “Injury” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 03.168.003?
None of the reports are classified as deaths, 2 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 03.168.003 FDA approved?
Devices of this type (product code HTJ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.
Who makes the NA 03.168.003, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 03.168.003). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other GAUGE, DEPTH devices
- NA 319.006Synthes GmbH · GAUGE, DEPTH892 reports
- NA 03.133.080Synthes GmbH · GAUGE, DEPTH169 reports
- NA 03.010.072Synthes GmbH · GAUGE, DEPTH129 reports
- MULTILOC 03.019.017Synthes GmbH · GAUGE, DEPTH116 reports
- NA 319.01Synthes GmbH · GAUGE, DEPTH111 reports
- NA 319.004Synthes GmbH · GAUGE, DEPTH82 reports
- QUICKSET 2274-60-000Depuy (Ireland) · GAUGE, DEPTH68 reports
- NA 319.091Synthes GmbH · GAUGE, DEPTH64 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page