Skip to content

NA 04.233.140SRecalls & reported problems

Synthes GmbH · RFNA / 11MM / 400MM STANDARD BEND / STERILE · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982290687), received between Apr 14, 2022 and Aug 18, 2025 — classified as 2 injuries.

2 other recalls cover devices cleared under the same FDA submission (K201346, K233696); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Bone Fracture(s)1
  • Hematoma1
  • Pulmonary Embolism1
  • Unspecified Infection1
  • Wound Dehiscence1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SUBJECT ID: (B)(6) STUDY NO: (B)(4). CLINICAL ADVERSE EVENT RECEIVED FOR LEFT FEMORAL NECK FRACTURE - PERI-IMPLANT DEVICE AND PROCEDURE(RELATEDNESS). DEVICE RELATED: POSSIBLE, PROCEDURE RELATED: NOT RELATED. DATE OF EVENT: 05 JUN 2025. DATE OF IMPLANT: (B)(6) 2024. DATE OF REVISION: NO INFORMATION PROVIDED. DEVICE LOCATION: LEFT. TREATMENT/IMPACT: REQUIRED IN-PATIENT HOSPITALIZATION OR PROLONGATION OF EXISTING HOSPITALIZATION, RESULTED IN MEDICAL OR SURGICAL INTERVENTION TO PREVENT LIFE-THREATENING ILLNESS OR INJURY OR PERMANENT IMPAIRMENT TO A BODY STRUCTURE OR A BODY FUNCTION, SURGICAL INTERVENTION NOT AT THE FRACTURE SITE (DEEP HARDWARE REMOVAL AND OPEN REDUCTION INTERNAL FIXATION FEMORAL NECK FRACTURE: BOTH 4.5 CORTICAL SCREWS WERE REMOVED. THEY WERE OBVIOUSLY BENT. REPLACED THEM WITH 6.5 MM CANCELLOUS SCREWS. GUIDEWIRES FOR PLANNED PLACEMENT OF TWO 6.5 CANNULATED SCREWS. CHECKED THE GUIDEWIRES ON MULTIPLE FLUOROSCOPIC VIEWS AND FOUND TO BE IN GOOD POSITION ALIGNMENT. WE USED A DRILL TO DRILL THE LATERAL CORTEX. WE THEN ADVANCED BOTH SCREWS AFTER MEASURING. SCREWS HAD GOOD TO FINGER BITE. REMOVE THE GUIDEWIRES) DEPUY SYNTHES PRODUCTS USED: CATALOG NUMBER: 04.233.140S LOT NUMBER

    Patient problems: Bone Fracture(s)

    MAUDE report 22825991 ↗
  • InjuryApr 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2022, THE PATIENT CAME TO OPERATING ROOM FOR REMOVAL OF RFNA DUE TO INFECTION. THE SURGEON VOICED THAT AFTER NAIL WAS ORIGINALLY INSERTED PATIENT HAD RECEIVED HEPARIN FOR A SIGNIFICANT PE AND DEVELOPED A LARGE HEMATOMA AT SURGICAL SITE WHICH ULTIMATELY LEAD TO THE WOUND BREAKING DOWN. HE BELIEVES THIS IS WHAT LEAD TO THE PATIENT HAVING AN INFECTION. RFNA WAS EXPLANTED AND ANTIBIOTIC SPACER WAS PLACED.

    Patient problems: Hematoma; Pulmonary Embolism; Unspecified Infection; Wound Dehiscence

    MAUDE report 14111402 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K201346, K233696. One submission often covers many models, so check the product and lot details in each record.

  • Z-0229-2026 · Sep 26, 2025 · Open, Classified
    LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail A
  • Z-0228-2026 · Sep 26, 2025 · Open, Classified
    LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail A

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page