NA 04.045.056Recalls & reported problems
Synthes GmbH · LOCKING SCREW FOR IM NAIL Ø 5MM/ 56MM/ XL25 · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982292919), received between Mar 24, 2022 and Aug 28, 2024 — classified as 2 injuries.
The most frequently reported specific device problem is “Migration” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K201336, K201346); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Migration2
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Insufficient Information1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryAug 28, 2024Migration; Migration
IT WAS REPORTED THAT ON JULY 30, 2024 A PATIENT UNDERWENT THE REMOVAL PROCEDURE DUE TO INTRAMEDULLARY SCREW BACK OUT OF RETROGRADE FEMORAL NAIL ADVANCED (RFNA). NO FURTHER INFORMATION IS PROVIDED. THIS REPORT IS FOR ONE (1) LOCKING SCREW FOR IM NAIL Ø 5MM/ 56MM/ XL25. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 20091465 ↗InjuryMar 24, 2022Migration; Migration
IT WAS REPORTED THAT ON (B)(6) 2022 THE PATIENT UNDERWENT A REVISION OF RETROGRADE FEMORAL NAILING-ADVANCED SYSTEM (RFNA) DUE TO THREE (3) SCREWS BACKING OUT IN THE DISTAL PORTION OF THE NAIL. ALL SCREWS WERE REMOVED. A DISTAL FEMUR PLATE WAS ADDED FOR FIXATION OF THE FRACTURE AND THE NAIL WAS LEFT IN AS WELL. NO FURTHER INFORMATION PROVIDED. THIS REPORT IS FOR ONE (1) LOCKING SCREW FOR IM NAIL Ø 5MM/ 56MM/ XL25. THIS IS REPORT 1 OF 3 FOR COMPLAINT (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13892255 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K201336, K201346. One submission often covers many models, so check the product and lot details in each record.
- Z-0229-2026 · Sep 26, 2025 · Open, ClassifiedLOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail A
- Z-0228-2026 · Sep 26, 2025 · Open, ClassifiedLOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail A
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.