NA 04.045.060Recalls & reported problems

Synthes GmbH · LOCKING SCREW FOR IM NAIL Ø 5MM/ 60MM/ XL25 · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886982292933), received between Jun 17, 2021 and May 26, 2025 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K201336, K201346); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Device-Device Incompatibility1
  • Migration1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • Nonunion/Delayed-union Bone Fracture1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS IMPLANTED WITH A RETROGRADE FEMORAL NAIL ADVANCED (RFNA) ON (B)(6) 2025. AN OPEN REDUCTION AND RETROGRADE INTRAMEDULLARY NAILING OF THE LEFT FEMUR SURGERY OCCURRED ON (B)(6) 2025. POST-OPERATIVELY X-RAYS SHOWED SIGNIFICANT BACK-OUT AND THE SCREW WAS PROMINENT. THE PATIENT HAD LIMITED FRACTURE HEALING AT THAT POINT, SO THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025, FOR SCREW EXCHANGE AND PLACEMENT OF A LOCKING NUT THAT THREADS ONTO THE SCREW. LATER IT WAS DISCOVERED THAT THE LOCKING NUT AND SCREW WERE DISENGAGED, AND THE SCREW WAS PARTIALLY BACKED OUT. ON (B)(6) 2025, ANOTHER SURGERY WAS PERFORMED FOR LEFT FEMUR REMOVAL OF HARDWARE, REVISION OF IM NAILING. POST-OPERATIVE FOLLOW-UP INDICATE THAT THE PATIENT APPEARS STABLE, AND IMAGING SHOWED NO LOOSENING OR FAILURES AT THE PRESENT TIME; THE FRACTURE ALIGNMENT REMAINS STABLE. THIS REPORT CAPTURES THE THIRD REMOVAL EVENT THAT OCCURRED ON (B)(6) 2025. THE FIRST REMOVAL EVENT THAT OCCURRED ON (B)(6) 2025, IS CAPTURED ON RELATED COMPLAINT (B)(4). THE SECOND REMOVAL EVENT THAT OCCURRED ON (B)(6) 2025, IS CAPTURED ON RELATED COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 22084194 ↗
  • MalfunctionJan 5, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2022, THE PATIENT UNDERWENT A TNA PROCEDURE WITH SUPRAPATELLAR APPROACH. THE PROXIMAL ANTERIOR TO POSTERIOR 0 DEGREE SCREW MISSED THE NAIL WHEN ATTEMPTING TO INSERT IT USING THE AIMING ARM AND THE TRIPLE GUIDE SLEEVES. THE SURGEON DRILLED THROUGH THE SLEEVES, BUT THE BIT DID NOT ALIGN WITH THE SCREW HOLE AND WENT MEDIAL TO THE NAIL. THE SCREW THAT WAS INSERTED FOLLOWED THE SAME PATH AND MISSED THE NAIL. THE SCREW WAS REMOVED, AND THE SURGEON DID NOT USE THAT SCREW HOLE. THERE WAS NO SURGICAL DELAY, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE LOCKING SCREW FOR IM NAIL Ø 5MM/ 60MM/ XL25. THIS IS REPORT 1 OF 3 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16102752 ↗
  • InjuryJan 28, 2022Migration; Migration

    IT WAS REPORTED THAT ON UNKNOWN DATE, PATIENT UNDERWENT AN ORTHOPEDIC PROCEDURE. IT WAS NOTED THAT THE DISTAL SCREW BACKING OUT OF RFNA NAIL. AN X-RAY TAKEN FIVE (5) WEEKS POST OPERATION SHOWED THE DISTAL SCREW BACKED OUT. IT IS UNKNOWN IF HOW THE PROCEDURE WAS COMPLETED. THERE WAS PATIENT CONSEQUENCE. THIS REPORT IS FOR ONE (1) LOCKING SCREW FOR IM NAIL Ø 5MM/ 60MM/ XL25. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13385247 ↗
  • InjuryJun 17, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6)2021, PATIENT UNDERWENT A REVISION DUE TO LOSS OF FIXATION. WE WERE MADE AWARE OF A REVISION OF IMPLANTS THAT'S SHOULD BEEN IMPLANTED ON (B)(6)2021. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. IT IS UNKNOWN IF THERE WAS SURGICAL DELAY REPORTED. PATIENT OUTCOME WAS UNKNOWN. THIS COMPLAINT INVOLVES (7) DEVICES. THIS REPORT IS FOR (1) LOCKING SCREW FOR IM NAIL 5/ 60/ XL25. THIS REPORT IS 4 OF 7 FOR (B)(4).

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 12021136 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K201336, K201346. One submission often covers many models, so check the product and lot details in each record.

  • Z-0228-2026 · Sep 26, 2025 · Open, Classified
    LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail A
  • Z-0229-2026 · Sep 26, 2025 · Open, Classified
    LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail A

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →