Skip to content

NA 04.045.334Recalls & reported problems

Synthes GmbH · LOW PROF LCKNG SCREW F/IM NAIL Ø 5.0MM / L 34MM / XL25 · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982293534), received between Sep 6, 2022 and Jan 23, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K201336, K201346); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Device-Device Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Nonunion/Delayed-union Bone Fracture1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 23, 2024Break; Break

    (B)(4) IS CREATED TO CAPTURE THE POST OP EVENT STATING THAT, IT WAS REPORTED THAT ON (B)(6) 2024, THE PATIENT CAME IN FOR A REVISION SURGERY OF A NONUNION OF A PROXIMAL THIRD TIBIA FRACTURE. PRIMARY SURGERY WAS TREATED WITH A TIBIAL NAIL ADVANCED. SURGEON REMOVED THE TIBIAL NAIL AND 4 LOW PROFILE INTERLOCK SCREWS. ONE OF THE PROXIMAL SCREWS WAS BROKEN. (B)(4) WAS FOR THE INTRA OP EVENT: SURGEON OPTED TO USE A RIA2 FOR BONE GRAFTING, WHICH WAS TAKEN FROM THE FEMUR. INITIALLY THERE WAS AN ISSUE WITH SUCTION OCCURRING THROUGH THE RIA2. RIA2 PARTS WERE DISSEMBLED AND PUT BACK TOGETHER. THERE WAS STILL SOME ISSUE WITH WATER SPRAYING OUT THE BACK END OF THE RIA2 SHAFT. SURGEON OPTED TO CONTINUE WITH PROCEDURE TO GET BONE GRAFT. WHILE REAMING THE REAMER HEAD BROKE OFF IN THE PATIENTS FEMORAL CANAL, THE 4 PRONGS THAT ATTACH THE REAMER HEAD TO THE SHAFT WERE RETAINED IN THE PATIENT. AFTER DISASSEMBLING THE RIA2 AFTER THE CASE, IT WAS NOTED THAT THE YELLOW KEYED SEAL DID NOT HAVE ANY BLACK SEAL ON IT. THE SECOND YELLOW SEAL INCLUDED IN THE PACKAGE ALSO DID NOT HAVE A BLACK SEAL. THE RIA2 OBTAINED SOME GRAFT, BUT SURGEON HAD TO ALSO GET BONE GRAFT FROM THE PROXIMAL TIBIA. ADDITIONAL XRAYS, AD

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 18565686 ↗
  • MalfunctionSep 6, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2022 AT THE TIME OF USING A TIBIAL NAIL AND ONE OF THE LOCKING SCREWS MISSED THE NAIL THROUGH AIMING JIG. THERE WAS SURGICAL DELAY DUE TO THE REPORTED EVENT BUT THE SURGERY WAS SUCCESSFULLY COMPLETED. PATIENT OUTCOME IS NORMAL. THIS REPORT IS FOR ONE (1) LOW PROF LCKNG SCR 5.0 / L 34 / XL25 THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15361452 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K201336, K201346. One submission often covers many models, so check the product and lot details in each record.

  • Z-0229-2026 · Sep 26, 2025 · Open, Classified
    LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail A
  • Z-0228-2026 · Sep 26, 2025 · Open, Classified
    LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail A

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page