NA 04.045.042SRecalls & reported problems
Synthes GmbH · LOCKING SCREW FOR IM NAIL Ø 5MM/ 42MM/ XL25/ STERILE · Class 2 · Implant
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982294432), received between Oct 20, 2023 and Oct 20, 2023 — classified as 1 injury.
2 other recalls cover devices cleared under the same FDA submission (K201336, K201346); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Nonunion/Delayed-union Bone Fracture1
- Unspecified Infection1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryOct 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM AUSTRALIA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, A BROKEN TNA NEEDED REMOVING AND RE-NAILING. THE NAIL HAD BROKEN AT THE DISTAL END. THE NAIL WAS ORIGINALLY IMPLANTED ON (B)(6) 2023, AND WAS REMOVED ON (B)(6) 2023 AND RE-NAILED WITH A 10X360MM TNA. ALL PREVIOUS NAIL ITEMS WERE REMOVED. THE REVISION PROCEDURE WAS PERFORMED BECAUSE OF THE BROKEN IMPLANT, THE FRACTURE HAD NOT HEALED. THE PATIENT HAD SIGNIFICANT DISTAL BONE LOSS AND A MASQUELET PROCEDURE WAS COMPLETED. THE PATIENT HAD BORNE WEIGHT AFTER BONE GRAFT HAD BEEN INSERTED, AND THIS IS WHEN THE NAIL BROKE. REAMINGS SENT OFF DURING THE CASE CAME BACK POSITIVE FOR INFECTION. THE REVISION PROCEDURE WAS COMPLETED SUCCESSFULLY AND THE PATIENT IS PROGRESSING WELL. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A LOCKING SCREW FOR IM NAIL Ø 5MM/ 42MM/ XL25/ STERILE. THIS IS REPORT 4 OF 7 FOR (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture; Unspecified Infection
MAUDE report 17975198 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K201336, K201346. One submission often covers many models, so check the product and lot details in each record.
- Z-0229-2026 · Sep 26, 2025 · Open, ClassifiedLOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Adva
- Z-0228-2026 · Sep 26, 2025 · Open, ClassifiedLOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Adva
Other NAIL, FIXATION, BONE devices
- NA 03.043.024Synthes GmbH · NAIL, FIXATION, BONE234 reports
- NA 03.010.440Synthes GmbH · NAIL, FIXATION, BONE70 reports
- TRIGEN 71676627Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall2 reports
- TRIGEN 71676628Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall0 reports
- NA 03.043.018Synthes GmbH · NAIL, FIXATION, BONE33 reports
- TRIGEN 71675383Smith & Nephew, Inc. · NAIL, FIXATION, BONE29 reports
- NA 03.010.441Synthes GmbH · NAIL, FIXATION, BONE27 reports
- NA 03.010.435Synthes GmbH · NAIL, FIXATION, BONE24 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.