Skip to content

NA 03.043.023Recalls & reported problems

Synthes GmbH · COMPRESSION SCREW/ MEDIUM · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982297228), received between Jan 17, 2023 and Jan 17, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device-Device Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 17, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2022, DURING A INTRA-OP CASE WHILE UTILIZING THE COMPRESSION DEVICE IN OUR ANS TNA PARAPATELLAR SET, THE COMPRESSION DEVICE WOULD NOT DISENGAGE FROM THE CONNECTING SCREW. THE SURGEON PROCEEDED TO REMOVE THE TWO DEVICES TOGETHER AS A UNIT AND SUBSEQUENTLY THREAD IN A NEW CONNECTING SCREW FROM THE TRAY. MAYBE A 10 SECOND DELAY OR ADDITION TO SURGICAL TIME. HOWEVER, WE STERILE PROCESSING REPORTED THAT THE COMPRESSION DEVICE AND CONNECTING SCREW WERE STILL COMPLETELY THREADED TOGETHER AND AS SUCH, LEFT OUT OF OUR INSTRUMENT SET. THEY TRIED SOAKING THE INSTRUMENTS TO LOOSEN THEM, CLAMPING THEM TO GET BETTER TORQUE FOR DISENGAGEMENT, ETC. TO NO AVAIL. THE LOT #'S IS NOT AVAILABLE AS THE COMPRESSION DEVICE IS STILL INSIDE AND THUS NOT VISIBLE, THE CONNECTING SCREW THAT HAD SOME SORT OF CLAMP ON IT DURING THE ATTEMPT TO DISENGAGE SUSTAINED DAMAGE WHICH SKEWS THE VISIBILITY OF THE LOT #. PATIENT CONSEQUENCES OR SURGERY RESULTS ARE UNKNOWN. THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS FOR ONE (1) COMPRESSION SCREW/ MEDIUM THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16189375 ↗

Other INSTRUMENT, COMPRESSION devices

All INSTRUMENT, COMPRESSION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page