Skip to content

NA 04.043.015SRecalls & reported problems

Synthes GmbH · TIBIAL NAIL-ADVANCED / 8MM 285MM / STERILE · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982297440), received between Jun 21, 2022 and Dec 7, 2023 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K201336); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Unspecified Infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE REPORT FROM BELGIUM REPORTS AN EVENT AS FOLLOWS: SUBJECT LEU-081 AE1: POSTOPERATIVE ADVERSE EVENT: "FRACTURE-RELATED INFECTION (FRI)" IS THE AE RELATED TO THE TIBIAL FRACTURE AND/OR TREATMENT OF TIBIAL FRACTURE? "YES" IS THE AE RELATED TO THE DEVICE? ¿YES" DPS IMPLANT WAS USED. THIS REPORT IS FOR A TIBIAL NAIL-ADVANCED / 8MM 285MM / STERILE. THIS IS REPORT 1 OF 10 FOR (B)(4).

    Patient problems: Unspecified Infection

    MAUDE report 18282715 ↗
  • MalfunctionJun 21, 2022Material Deformation; Material Deformation

    DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, DURING A PROCEDURE FOR ORIF WITH A TNA FOR THE FRACTURE OF THE DIAPHYSEAL TIBIA, THE SURGEON INSERTED THE NAIL TO THE OPTIMAL DEPTH TO BEGIN DRILLING FOR A PROXIMAL LOCKING SCREW. THE DRILL BIT INTERFERED WITH THE NAIL HOLE WHILE DRILLING, IN BOTH ML HOLES. AFTER INVESTIGATING, IT WAS DISCOVERED THAT THE INTERFERENCE OCCURRED BEHIND THE HOLE, SO THE SURGEON DRILLED AGAIN WHILE ADJUSTING BY HAND. THE DRILL BIT INTERFERED WITH THE NAIL HOLE, BUT THE SURGEON WAS ABLE TO DRILL AND INSERT SCREWS. WHILE REMOVING THE INSERTION HANDLE, THE CONNECTING SCREW CAME OFF WITH A RASPING SOUND AND THE SURGEON CONCLUDED THAT NAIL DEFLECTION HAD OCCURRED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH LESS THAN A 30 MINUTES DELAY. THIS REPORT IS FOR A TIBIAL NAIL-ADVANCED / 8MM 285MM / STERILE. THIS IS REPORT 1 OF 2 FOR (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14758462 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K201336. One submission often covers many models, so check the product and lot details in each record.

  • Z-0229-2026 · Sep 26, 2025 · Open, Classified
    LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail A
  • Z-0228-2026 · Sep 26, 2025 · Open, Classified
    LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail A

Other NAIL, FIXATION, BONE devices

All NAIL, FIXATION, BONE devices →

Sources and updates