NA 04.043.150SRecalls & reported problems
Synthes GmbH · TIBIAL NAIL-ADVANCED / 9MM 390MM / STERILE · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886982297662), received between Jan 27, 2022 and Jul 23, 2024 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K201336); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility2
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Failure of Implant1
- Nonunion/Delayed-union Bone Fracture1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 23, 2024Detachment of Device or Device Component; Device-Device Incompatibility; Detachment of Device or Device Component; Device-Device Incompatibility
IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2024, THE POLY INLAY IN THE TIBIAL NAIL WAS DISLODGED AND CAUSED A BLOCKAGE WITHIN THE NAIL. THIS MADE IT UNABLE TO PASS THE BALL TO GUIDE WIRE DOWN THE CANNULATION OF THE NAIL. SURGERY WAS SUCCESSFUL COMPLETED WITH NO DELAY. THIS REPORT IS FOR A TIBIAL NAIL-ADVANCED / 9MM 390MM / STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19806528 ↗InjuryMay 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SUBJECT (B)(6). AE1: POSTOPERATIVE ADVERSE EVENT: " NONUNION". IS THE AE RELATED TO THE TREATMENT OF FRACTURE? "YES".. IS THE AE RELATED TO THE DEVICE? "YES". DPS IMPLANT WAS USED. PATIENT UNDERWENT REVISION SURGERY. PATIENT IS RECOVERING. THIS REPORT IS FOR A TIBIAL NAIL-ADVANCED / 9MM 390MM / STERILE. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 19253886 ↗InjurySep 29, 2022Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PATIENT UNDERWENT REVISION SURGERY FOR A BROKEN TIBIAL NAIL THAT HAD BEEN IMPLANTED THREE MONTHS PRIOR. THE PATIENT HAD HYPERTROPHIC NONUNION, AND THE NAIL WAS BROKEN THROUGH THE DISTAL/PROXIMAL SCREW HOLE OF THE ANS TIBIAL NAIL 9-390MM. THE SURGEON REMOVED ALL THE SCREWS INTACT AND UTILIZED AN EXTRACTION HOOK TO REMOVE THE ENTIRE NAIL INCLUDING THE DISTAL PORTION BELOW THE NAIL BREAKAGE. IT WAS A DIFFICULT PROCEDURE. THE NAIL WAS REPLACED WITH A TIBIAL NAIL EX10X390MM WITH MULTIPLE SCREWS AND BLOCKING SCREWS. IT WAS REPORTED THAT THE PATIENT WEIGHED 300 POUNDS AND HAD MULTIPLE COMORBIDITIES. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE TIBIAL NAIL-ADVANCED / 9MM 390MM / STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Failure of Implant
MAUDE report 15501499 ↗MalfunctionJan 27, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2021 DURING A TIBIA PROCEDURE, THE SURGEON IMPLANTED THE TIBIAL NAIL IN QUESTION. THE NAIL WAS INSERTED PER NORMAL TECHNIQUE AND WAS THEN SUBSEQUENTLY REMOVED TO PERFORM ADDITIONAL INTRAMEDULLARY REAMING. UPON REMOVAL, WE NOTICED THE DISTAL POLY INLAY HAD PARTIALLY POPPED OUT OF THE END OF THE NAIL. THE SURGEON WAS ABLE TO PUSH IT BACK IN MANUALLY AND RE-INSERT THE NAIL AND COMPLETE THE PROCEDURE SUCCESSFULLY. THERE WAS AN UNSPECIFIC AMOUNT OF SURGICAL DELAY. THERE IS NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR ONE (1) TIBIAL NAIL-ADVANCED / 9MM 390MM / STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13366780 ↗
FDA clearance and classification
Cleared through 510(k) K201336 on Sep 3, 2020 (decision: Substantially Equivalent). Submitted by Synthes (USA) Products, LLC as “DePuy Synthes Tibial Nail Advanced, Depuy Synthes Locking Screws Medullary Nails, 4.0 and 5.0mm”. The FDA review took 106 days. FDA record ↗
The FDA regulates product code JDS (nail, fixation, bone) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K201336. One submission often covers many models, so check the product and lot details in each record.
- Z-0229-2026 · Sep 26, 2025 · Open, ClassifiedLOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Adva
- Z-0228-2026 · Sep 26, 2025 · Open, ClassifiedLOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Adva
Common questions about the NA 04.043.150S
Has the NA 04.043.150S been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886982297662), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the NA 04.043.150S?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jan 27, 2022 and Jul 23, 2024. In those reports, the most common specific device problems are “Device-Device Incompatibility” (2), “Break” (1) and “Detachment of Device or Device Component” (1); the most common patient problems are “Failure of Implant” (1) and “Nonunion/Delayed-union Bone Fracture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 04.043.150S?
None of the reports are classified as deaths, 2 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 04.043.150S FDA approved?
It was cleared through 510(k) K201336 on Sep 3, 2020. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the NA 04.043.150S MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the NA 04.043.150S, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 04.043.150S). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other NAIL, FIXATION, BONE devices
- NA 03.043.024Synthes GmbH · NAIL, FIXATION, BONE233 reports
- NA 03.010.440Synthes GmbH · NAIL, FIXATION, BONE69 reports
- TRIGEN 71676627Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall2 reports
- TRIGEN 71676628Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall0 reports
- NA 03.043.018Synthes GmbH · NAIL, FIXATION, BONE33 reports
- TRIGEN 71675383Smith & Nephew, Inc. · NAIL, FIXATION, BONE29 reports
- NA 03.010.441Synthes GmbH · NAIL, FIXATION, BONE27 reports
- NA 03.010.435Synthes GmbH · NAIL, FIXATION, BONE23 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page