NA 04.043.435SRecalls & reported problems
Synthes GmbH · TIBIAL NAIL-ADVANCED / 12MM 345MM / STERILE · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886982298089), received between Jul 18, 2022 and Aug 15, 2023 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (3 reports).
2 other recalls cover devices cleared under the same FDA submission (K201336); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility3
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Failure of Implant1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 15, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2023, THE PIN HAD FALLEN OUT OF THE TNA DRIVING CAP. DURING THE PROCEDURE, A TNA NAIL WAS INSERTED, BUT THEN REMOVED TO DIAL IN THE REDUCTION. WHEN THE NAIL WAS EXTRACTED, IT WAS NOTICED THAT THE POLY INSERTED DISTALLY HAD MOVED OUTSIDE OF THE NAIL, THE SURGICAL TECH WAS NOT ABLE TO PUT BACK INTO THE NAIL, AND A NEW NAIL WAS SELECTED. THIS NAIL WAS ALSO EXTRACTED FOR REPOSITIONING AND THE SLEEVE HAD MOVED AS WELL, BUT SURGEON DECIDED TO MOVE FORWARD AND DRILL THROUGH THE POLY INSERT. ADDITIONALLY, THE SURGEON WAS HAVING ISSUES INSERTING THE NAIL THROUGH THE SUPRAPATELLAR PROTECTION SLEEVE, SO IT WAS REMOVED AND THE NAIL WAS INSERTED WITHOUT THE SLEEVE. DURING THE SURGERY, THE SURGEON USED A DRILL BIT IN A ¿BLOCKING¿ FASHION TO REDIRECT THE NAIL, BUT THE DRILL BIT WAS UNABLE TO REMAIN STABLE BECAUSE OF THE COMMINUTION. THEN A BLOCKING SCREW WAS USED. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO ADVERSE PATIENT IMPACT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A TIBIAL NAIL-ADVANCED / 12MM 345MM / STERILE. THIS IS REPORT 2 OF 4 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17548591 ↗MalfunctionAug 15, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2023, THE PIN HAD FALLEN OUT OF THE TNA DRIVING CAP. DURING THE PROCEDURE, A TNA NAIL WAS INSERTED, BUT THEN REMOVED TO DIAL IN THE REDUCTION. WHEN THE NAIL WAS EXTRACTED, IT WAS NOTICED THAT THE POLY INSERTED DISTALLY HAD MOVED OUTSIDE OF THE NAIL, THE SURGICAL TECH WAS NOT ABLE TO PUT BACK INTO THE NAIL, AND A NEW NAIL WAS SELECTED. THIS NAIL WAS ALSO EXTRACTED FOR REPOSITIONING AND THE SLEEVE HAD MOVED AS WELL, BUT SURGEON DECIDED TO MOVE FORWARD AND DRILL THROUGH THE POLY INSERT. ADDITIONALLY, THE SURGEON WAS HAVING ISSUES INSERTING THE NAIL THROUGH THE SUPRAPATELLAR PROTECTION SLEEVE, SO IT WAS REMOVED AND THE NAIL WAS INSERTED WITHOUT THE SLEEVE. DURING THE SURGERY, THE SURGEON USED A DRILL BIT IN A ¿BLOCKING¿ FASHION TO REDIRECT THE NAIL, BUT THE DRILL BIT WAS UNABLE TO REMAIN STABLE BECAUSE OF THE COMMINUTION. THEN A BLOCKING SCREW WAS USED. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO ADVERSE PATIENT IMPACT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A TIBIAL NAIL-ADVANCED / 12MM 345MM / STERILE. THIS IS REPORT 1 OF 4 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17548589 ↗InjuryJul 19, 2022Break; Break
IT WAS REPORTED ON (B)(6) 2022 THAT A PATIENT WAS INVOLVED, ONE OF THE TIBIAL NAIL TNA, THE DISTAL POLYMER INLAY CAME OUT INSIDE THE PATIENT. THEY WERE ABLE TO RETRIEVED THE INLAY. THIS REPORT IS FOR ONE (1) TIBIAL NAIL-ADVANCED / 12 345 / STERILE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Failure of Implant
MAUDE report 15056181 ↗MalfunctionJul 18, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED ON UNKNOWN DATE THAT, THE PATIENT HAD AN EXISTING 4.5MM LCP PROXIMAL TIBIA PLATE. THE INTENT WAS TO TENSION THE LATERAL SIDE TO REDUCE THE MALUNION/NON UNION THEN NAIL THE TIBIA AND AUGMENT THE MEDIAL SIDE WITH A BUTTRESS PLATE. UPON PLACEMENT OF THE ATD ON THE EXISTING LATER PLATE, THE HOOK FROM THE ATD SWIVEL BROKE OFF. WE RETRIEVED THE ITEM FROM THE PATIENT, NO RESIDUAL FRAGMENTS WERE LEFT IN THE PATIENTS. WE USED A NEW ATD AND IT WORKED FINE. WHEN NAILING THE TIBIA THE POLY INSIDE THE TNA NAIL BECAME DAMAGED WHEN TRYING TO REMOVE THE GOLD "TEAR DROP" ZIMMER GUIDE ROD (47-2490-097-00) AFTER INSERTION OF THE NAIL. WE REALIZED IT WHEN WE WERE ROTATING THE TNA SUPRAPATELLAR HANDLE ATTACHED TO THE NAIL BUT THE HOLES IN THE NAIL ON X RAY WERE NOT MOVING. SO WE REMOVED THE NAIL AND HAD TO REAM UP AND USE THE NEXT SIZE DIAMETER NAIL. WE ONLY HAD ONE EACH OF THE TNA SIZES. THERE WERE A SURGICAL DELAY OF 20 MINUTES. PROCEDURE WAS COMPLETED SUCCESSFULLY. PATIENT OUTCOME IS STABLE. THIS REPORT IS FOR A TIBIAL NAIL-ADVANCED / 12MM 345MM / STERILE. THIS IS REPORT 2 OF 5 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15042727 ↗
FDA clearance and classification
Cleared through 510(k) K201336 on Sep 3, 2020 (decision: Substantially Equivalent). Submitted by Synthes (USA) Products, LLC as “DePuy Synthes Tibial Nail Advanced, Depuy Synthes Locking Screws Medullary Nails, 4.0 and 5.0mm”. The FDA review took 106 days. FDA record ↗
The FDA regulates product code JDS (nail, fixation, bone) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K201336. One submission often covers many models, so check the product and lot details in each record.
- Z-0228-2026 · Sep 26, 2025 · Open, ClassifiedLOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Adva
- Z-0229-2026 · Sep 26, 2025 · Open, ClassifiedLOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Adva
Common questions about the NA 04.043.435S
Has the NA 04.043.435S been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886982298089), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the NA 04.043.435S?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jul 18, 2022 and Aug 15, 2023. In those reports, the most common specific device problems are “Device-Device Incompatibility” (3) and “Break” (1); the most common patient problem is “Failure of Implant” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 04.043.435S?
None of the reports are classified as deaths, 1 as an injury and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 04.043.435S FDA approved?
It was cleared through 510(k) K201336 on Sep 3, 2020. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the NA 04.043.435S MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the NA 04.043.435S, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 04.043.435S). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other NAIL, FIXATION, BONE devices
- NA 03.043.024Synthes GmbH · NAIL, FIXATION, BONE233 reports
- NA 03.010.440Synthes GmbH · NAIL, FIXATION, BONE69 reports
- TRIGEN 71676627Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall2 reports
- TRIGEN 71676628Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall0 reports
- NA 03.043.018Synthes GmbH · NAIL, FIXATION, BONE33 reports
- TRIGEN 71675383Smith & Nephew, Inc. · NAIL, FIXATION, BONE29 reports
- NA 03.010.441Synthes GmbH · NAIL, FIXATION, BONE27 reports
- NA 03.010.435Synthes GmbH · NAIL, FIXATION, BONE23 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page