Radial Head Replacement System Implant is a one-piece… recallZ-1072-2022 · Synthes (USA) Products LLC · initiated Apr 7, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes (USA) Products LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
DePuy Synthes Radial Head Replacement System Contraindications added to Electronic Instructions for Use (e-IFU): Absolute contraindications include: Infection, sepsis, osteomyelitis.
FDA root cause category: Labeling Change Control
What the company is doing
DePuy Synthesis issued Urgent: MEDICAL DEVICE CORRECTION (Notification) on 4/14/22. Letter states reason for recall, health risk and action to take: " Download and review the updated version of the e-IFU from the following website: https://www.e-ifu.com/search-document-metadata/radial%20head%20replacement%20system Forward this notice to all attending surgeons in your facility using this device. " Please complete, sign, and return the Business Reply Form within three (3) business days email to [email protected] (Please reference FA 2097468 in the subject line). Note: Product is NOT being removed from the field and does not need to be returned.
Recall details
- Recall number
- Z-1072-2022
- Status
- Open, Classified
- Initiated by the company
- Apr 7, 2022
- Recalling firm
- Synthes (USA) Products LLC
- Product code
- KWI
- FDA submissions
- K183618
- Distribution
- US Nationwide distribution.
Product description
Radial Head Replacement System Implant is a one-piece, stemmed radial head replacement. The radial head is available in diameters of 19, 22 and 25mm. The radial stems range from 5.5 to 8.5mm in diameter and from 21 24mm in length.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
All serial numbers. Material GTIN Product 03.405.000S 10886982300829 Radial Head Replacement Instr Kit; 09.405.250S 10886982300614 Rad Head w/Head 22 shaft 5.5 HH10 +0; 09.405.253S 10886982300621 Rad Head w/Head 22 shaft 5.5 HH10 +3; 09.405.256S 10886982300638 Rad Head w/Head 22 shaft 5.5 HH10 +6; 09.405.260S 10886982300645 Rad Head w/Head 22 shaft 6.5 HH10 +0; 09.405.263S 10886982300652 Rad Head w/Head 22 shaft 6.5 HH10 +3; 09.405.266S 10886982300669 Rad Head w/Head 22 shaft 6.5 HH10 +6; 09.405.270S 10886982300676 Rad Head w/Head 22 shaft 7.5 HH10 +0; 09.405.273S 10886982300683 Rad Head w/Head 22 shaft 7.5 HH10 +3; 09.405.276S 10886982300690 Rad Head w/Head 22 shaft 7.5 HH10 +6; 09.405.280S 10886982300706 Rad Head w/Head 22 shaft 8.5 HH10 +0; 09.405.283S 10886982300713 Rad Head w/Head 22 shaft 8.5 HH10 +3; 09.405.286S 10886982300720 Rad Head w/Head 22 shaft 8.5 HH10 +6; 09.405.560S 10886982300737 Rad Head w/Head 25 shaft 6.5 HH11 +0; 09.405.563S 10886982300744 Rad Head w/Head 25 shaft 6.5 HH11 +3; 09.405.566S 10886982300751 Rad Head w/Head 25 shaft 6.5 HH11 +6; 09.405.570S 10886982300768 Rad Head w/Head 25 shaft 7.5 HH11 +0; 09.405.573S 10886982300775 Rad Head w/Head 25 shaft 7.5 HH11 +3; 09.405.576S 10886982300782 Rad Head w/Head 25 shaft 7.5 HH11 +6; 09.405.580S 10886982300799 Rad Head w/Head 25 shaft 8.5 HH11 +0; 09.405.583S 10886982300805 Rad Head w/Head 25 shaft 8.5 HH11 +3; 09.405.586S 10886982300812 Rad Head w/Head 25 shaft 8.5 HH11 +6; 09.405.950S 10886982300522 Rad Head w/Head 19 shaft 5.5 HH9 +0; 09.405.953S 10886982300539 Rad Head w/Head 19 shaft 5.5 HH9 +3; 09.405.956S 10886982300546 Rad Head w/Head 19 shaft 5.5 HH9 +6; 09.405.960S 10886982300553 Rad Head w/Head 19 shaft 6.5 HH9 +0; 09.405.963S 10886982300560 Rad Head w/Head 19 shaft 6.5 HH9 +3; 09.405.966S 10886982300577 Rad Head w/Head 19 shaft 6.5 HH9 +6; 09.405.970S 10886982300584 Rad Head w/Head 19 shaft 7.5 HH9 +0; 09.405.973S 10886982300591 Rad Head w/Head 19 shaft 7.5 HH9 +3; 09.405.976S 10886982300607 Rad Head w/Head 19 shaft 7.5 HH9 +6.
Devices named in this recall
31 devices in the FDA device registry whose identifier appears in the recall's code information.
- NA 09.405.966SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall2 reports
- NA 03.405.000SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall2 reports
- NA 09.405.250SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall1 report
- NA 09.405.263SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall1 report
- NA 09.405.570SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall1 report
- NA 09.405.583SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall1 report
- NA 09.405.950SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.953SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.956SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.960SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.963SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.970SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.973SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.976SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.253SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.256SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.260SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.266SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.270SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.273SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.276SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.280SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.283SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.286SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.560SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.563SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.566SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.573SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.576SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.580SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
- NA 09.405.586SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page