NA 03.424.060Recalls & reported problems
Synthes GmbH · 2.4 VOLT(TM) DEPTH GAUGE 0 TO 60MM · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10886982335920), received between Jan 22, 2025 and Mar 4, 2026 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Material Deformation3
- Tear, Rip or Hole in Device Packaging2
Patient problems
- No Clinical Signs, Symptoms or Conditions9
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 4, 2026Material Deformation
IT WAS REPORTED THAT THE DEPTH GAUGE WAS BENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24512693 ↗MalfunctionNov 17, 2025Break; Break
ADDITIONAL INFORMATION RECEIVED STATES THAT THIS WAS A POST-OPERATIVE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23574232 ↗MalfunctionOct 17, 2025Break; Break
IT WAS REPORTED THAT THE VOLT DEPTH GAUGE BROKE AT THE METAL TIP COLOR JUNCTION AND NEEDS TO BE REPLACED. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23329727 ↗MalfunctionOct 17, 2025Break; Break
IT WAS REPORTED THAT THE VOLT DEPTH GAUGE BROKE AT THE METAL TIP COLOR JUNCTION AND NEEDS TO BE REPLACED. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23329723 ↗MalfunctionOct 17, 2025Break; Break
IT WAS REPORTED THAT THE VOLT DEPTH GAUGE BROKE AT THE METAL TIP COLOR JUNCTION AND NEEDS TO BE REPLACED. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23329717 ↗MalfunctionOct 17, 2025Break; Break
IT WAS REPORTED THAT THE VOLT DEPTH GAUGE BROKE AT THE METAL TIP COLOR JUNCTION AND NEEDS TO BE REPLACED. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23329712 ↗MalfunctionMay 1, 2025Break; Break
IT WAS REPORTED THAT THERE WERE BROKEN DEPTH GAUGES IN SPD AND HANDLE. THIS EVENT HAPPENED DURING STERILE PROCESSING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21945352 ↗MalfunctionJan 22, 2025Tear, Rip or Hole in Device Packaging; Material Deformation; Tear, Rip or Hole in Device Packaging; Material Deformation
ADDITIONAL INFORMATION RECEIVED STATES THAT WAS NO ALLEGATION AGAINST THE LOCKING SCREWS. THIS WAS NOTICED WHEN THE ITEMS WERE BEING TAKEN OUT OF THE BOX TO BE PUT INTO THE SET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21206672 ↗MalfunctionJan 22, 2025Tear, Rip or Hole in Device Packaging; Material Deformation; Tear, Rip or Hole in Device Packaging; Material Deformation
ADDITIONAL INFORMATION RECEIVED STATES THAT WAS NO ALLEGATION AGAINST THE LOCKING SCREWS. THIS WAS NOTICED WHEN THE ITEMS WERE BEING TAKEN OUT OF THE BOX TO BE PUT INTO THE SET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21206649 ↗
FDA clearance and classification
The FDA regulates product code HTJ (gauge, depth) under 21 CFR 888.4300 as a Class I device. 21 CFR 888.4300 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 03.424.060
Has the NA 03.424.060 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886982335920), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 03.424.060?
The FDA's MAUDE database lists 9 adverse event reports for this device, received between Jan 22, 2025 and Mar 4, 2026. In those reports, the most common specific device problems are “Break” (6), “Material Deformation” (3) and “Tear, Rip or Hole in Device Packaging” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 03.424.060?
None of the reports are classified as deaths, 0 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 03.424.060 FDA approved?
Devices of this type (product code HTJ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.
Who makes the NA 03.424.060, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 03.424.060). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other GAUGE, DEPTH devices
- NA 319.006Synthes GmbH · GAUGE, DEPTH892 reports
- NA 03.133.080Synthes GmbH · GAUGE, DEPTH169 reports
- NA 03.010.072Synthes GmbH · GAUGE, DEPTH129 reports
- MULTILOC 03.019.017Synthes GmbH · GAUGE, DEPTH116 reports
- NA 319.01Synthes GmbH · GAUGE, DEPTH111 reports
- NA 319.004Synthes GmbH · GAUGE, DEPTH82 reports
- QUICKSET 2274-60-000Depuy (Ireland) · GAUGE, DEPTH68 reports
- NA 319.091Synthes GmbH · GAUGE, DEPTH64 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page