Skip to content

NA SD03.127.004Recalls & reported problems

Synthes GmbH · VA DRILL SLEEVE 3.5 FOR DRILL BIT 2.5MM · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982342089), received between Jul 23, 2024 and Jul 23, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility2
  • Material Twisted/Bent2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 23, 2024Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT BOLT WAS BENT WHEN OPENED. AIMING ARM MISSED NAIL. LOCKING SCREW DID NOT LOCK INTO PLATE AFTER BENT. CALIBRATION WAS OFF FOR TWO DEPTH GAUGE AND TWO (2) 3.5MM VARIABLE ANGLE (VA) DRILL SLEEVES WERE NOT ABLE TO THREAD INTO HANDLE, UNSURE IF THREADS WERE STRIPPED. THERE WAS NO PATIENT CONSEQUENCES REPORTED. THERE WAS NO SURGICAL DELAY. THIS REPORT IS FOR ONE (1) VA DRILL SLEEVE 3.5 FOR DRILL BIT 2.5MM. THIS IS REPORT 6 OF 6 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19808871 ↗
  • MalfunctionJul 23, 2024Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT BOLT WAS BENT WHEN OPENED. AIMING ARM MISSED NAIL. LOCKING SCREW DID NOT LOCK INTO PLATE AFTER BENT. CALIBRATION WAS OFF FOR TWO DEPTH GAUGE AND TWO (2) 3.5MM VARIABLE ANGLE (VA) DRILL SLEEVES WERE NOT ABLE TO THREAD INTO HANDLE, UNSURE IF THREADS WERE STRIPPED. THERE WAS NO PATIENT CONSEQUENCES REPORTED. THERE WAS NO SURGICAL DELAY. THIS REPORT IS FOR ONE (1) VA DRILL SLEEVE 3.5 FOR DRILL BIT 2.5MM. THIS IS REPORT 5 OF 6 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19808836 ↗

Other Guide, surgical, instrument devices

All Guide, surgical, instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page