NA SD03.043.031Recalls & reported problems
Synthes GmbH · TNA CONNECTING TUBE ROTATING TNA AIMING ARM · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982344366), received between Oct 1, 2024 and Feb 17, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K201336); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 17, 2026Device-Device Incompatibility
IT WAS REPORTED THAT CONNECT F/EXPERT HAN F/03.008.007, TNA CONNECTING TUBE TNA ARM, AND AIMING ARM ASSEMBLY TNA ARM WERE INVOLVED IN AN EVENT WHERE THE STRIKE CAP AND TNA CONNECTING TUBE BECAME WEDGED TOGETHER, MAKING IT IMPOSSIBLE FOR STAFF TO DISASSEMBLE THE DEVICES. ATTEMPTS TO REMOVE THE COMPONENTS PLACED SIGNIFICANT STRESS ON THE PATIENT, LEADING TO A CHANGE IN THE REMOVAL APPROACH. ALL THREE PIECES REMAINED STUCK TOGETHER AFTER THE PROCEDURE. THERE WAS NO REPORTED PATIENT CONSEQUENCE OR CLINICAL SYMPTOMS AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24365754 ↗MalfunctionOct 1, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THERE WERE ISSUES WITH SOME DEVICES. THE AIMING ARM IS INTENDED TO BE USED TO TARGET THE SCREWS IN THE PROXIMAL PORTION OF THE NAIL. HOWEVER, THE AIMING ARM MISSED THE HOLE IN THE TNA NAIL. PROCEDURE AND PATIENT OUTCOME WERE UNKNOWN. THIS REPORT IS FOR A TNA CONNECTING TUBE ROTATING TNA AIMING ARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20344557 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K201336. One submission often covers many models, so check the product and lot details in each record.
- Z-0229-2026 · Sep 26, 2025 · Open, ClassifiedLOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Adva
- Z-0228-2026 · Sep 26, 2025 · Open, ClassifiedLOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Adva
Other NAIL, FIXATION, BONE devices
- NA 03.043.024Synthes GmbH · NAIL, FIXATION, BONE234 reports
- NA 03.010.440Synthes GmbH · NAIL, FIXATION, BONE70 reports
- TRIGEN 71676627Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall2 reports
- TRIGEN 71676628Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall0 reports
- NA 03.043.018Synthes GmbH · NAIL, FIXATION, BONE33 reports
- TRIGEN 71675383Smith & Nephew, Inc. · NAIL, FIXATION, BONE29 reports
- NA 03.010.441Synthes GmbH · NAIL, FIXATION, BONE27 reports
- NA 03.010.435Synthes GmbH · NAIL, FIXATION, BONE24 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.