NA SD480.110Recalls & reported problems
Synthes GmbH · TI PATIENT SPECIFIC PLATE MANDIBLE/ANGLE/2.0MM THICK · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10887587005324), received between Mar 6, 2018 and Jul 18, 2022 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Inadequacy of Device Shape and/or Size2
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Defective Device1
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Code Available2
- Unspecified Infection2
- Bone Fracture(s)1
- Nonunion/Delayed-union Bone Fracture1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 18, 2022Defective Device; Defective Device
DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT PRIOR TO A MANDIBULAR RECONSTRUCTION PROCEDURE ON 22-JUNE-2022, IT WAS CONFIRMED BY THE SURGEON THAT THE CUSTOM-MADE DEVICES DID NOT CONFORM TO THE PATIENT¿S CT DATA. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A TI PATIENT SPECIFIC PLATE MANDIBLE/ANGLE/2.0MM THICK. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15040995 ↗MalfunctionMar 23, 2021Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE HUB OF THE CEREBASE CATHETER BROKE WHERE THE ORANGE AND YELLOW PIECE MEET. A BALLAST CATHETER WAS USED TO COMPLETE THE PROCEDURE. IT IS UNKNOWN IF THERE IS SURGICAL DELAY REPORTED. THERE WERE NO FRAGMENTS GENERATED. THIS REPORT INVOLVES ONE (1) TI PATIENT SPECIFIC PLATE MANDIBLE/ANGLE/2.0MM THICK. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11554093 ↗InjuryMar 1, 2019No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT MANDIBLE RESECTION. THE PLATE AND SCREWS WERE PLACED, AND A TITANIUM PATIENT SPECIFIC PLATE MANDIBLE WAS REMOVED. THE PATIENT WAS NOTED TO HAVE POOR BONE QUALITY. THERE WAS NO MALFUNCTION OF THE PLATE OR SCREWS. THE PATIENT HAD A FRACTURE AROUND THE PLATE AND SCREWS AND PROBABLE OSTEOMYELITIS. ONLY THE BONE AROUND THE HARDWARE FAILED. IT IS UNKNOWN IF THERE WAS SURGICAL DELAY. PROCEDURE AND PATIENT OUTCOME ARE UNKNOWN. THIS REPORT IS FOR ONE (1) TI PATIENT SPECIFIC PLATE MANDIBLE/ANGLE/2.0MM THICK. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: Bone Fracture(s); Unspecified Infection
MAUDE report 8382490 ↗InjuryJul 17, 2018Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT MULTIPLE PATIENT SPECIFIC IMPLANTS MALFUNCTIONED. THE SURGERY WAS PERFORMED ON (B)(6) 2018. THE PROCEDURE WAS PERFORMED TO CORRECT HEMIFACIAL MICROSOMIA OF THE PATIENT. THERE WAS NOTHING WRONG WITH THE DEVICE BUT THE CASE WAS VERY COMPLICATED AND THE DEVICE DID NOT LINE UP WITH THE SURGEON'S PREFERENCE. THERE WAS NO DEFECT IN THE DEVICE. BOTH PLATES NEED TO BE SHORTER AS THEY EXTENDED TOO FAR TOWARD THE ANGLE AND THE BODY; THE PLATES WERE TOO THICK. THESE PLATES ULTIMATELY WERE NOT IMPLANTED. THERE WAS NO DELAY IN THE PROCEDURE AND THE PATIENT OUTCOME WAS FINE. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 7693676 ↗InjuryJul 17, 2018Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT MULTIPLE PATIENT SPECIFIC IMPLANTS MALFUNCTIONED. THE SURGERY WAS PERFORMED ON (B)(6) 2018. THE PROCEDURE WAS PERFORMED TO CORRECT HEMIFACIAL MICROSOMIA OF THE PATIENT. THERE WAS NOTHING WRONG WITH THE DEVICE BUT THE CASE WAS VERY COMPLICATED AND THE DEVICE DID NOT LINE UP WITH THE SURGEON'S PREFERENCE. THERE WAS NO DEFECT IN THE DEVICE. BOTH PLATES NEED TO BE SHORTER AS THEY EXTENDED TOO FAR TOWARD THE ANGLE AND THE BODY; THE PLATES WERE TOO THICK. THESE PLATES ULTIMATELY WERE NOT IMPLANTED. THERE WAS NO DELAY IN THE PROCEDURE AND THE PATIENT OUTCOME WAS FINE. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 7693649 ↗InjuryMar 6, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A SURGEON TOOK OUT A TRUMATCH MANDIBULAR PLATE AND SEVEN LOCKING SCREWS FROM THE PATIENT ON (B)(6) 2018 DUE AN INFECTION AND NON-UNION. THE PATIENT WAS IMPLANTED WITH ONE (1) TRUMATCH MANDIBULAR PLATE AND (7) SEVEN LOCKING SCREWS ON (B)(6) 2017. THE PATIENT EXPERIENCED NON-UNION AND INFECTION POST-SURGERY AND THE SURGEON REMOVED ALL THE HARDWARE. THE HARDWARE REMOVAL WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. THE PATIENT STATUS IS REPORTED TO BE FINE POST-SURGERY. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture; Unspecified Infection
MAUDE report 7318763 ↗
Other PLATE, BONE devices
- MatrixNEURO 04.503.104.01Synthes GmbH · PLATE, BONE252 reports
- RapidSorb 806.004.02SSynthes GmbH · PLATE, BONE146 reports
- TruMatch SD980.001Materialise NV · Plate, Bone1 recall21 reports
- Matrix 04.511.225.01Synthes GmbH · PLATE, BONE62 reports
- NA 400.834Synthes GmbH · PLATE, BONE58 reports
- TruMatch SD980.016Materialise NV · Plate, Bone55 reports
- ProPlan CMF SD900.101Materialise NV · Plate, Bone1 recall1 report
- TruMatch SD980.013Materialise NV · Plate, Bone1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page