MatrixMANDIBLE 03.503.408Recalls & reported problems
Synthes GmbH · MATRIXMANDIBLE 1.5MM DRILL BIT J-LATCH W/8MM STOP/50MM · Class 2
All models of MatrixMANDIBLE (79)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10887587012889), received between Dec 28, 2018 and Sep 16, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
6 other recalls cover devices cleared under the same FDA submission (K043310); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Failure to Cut4
- Overheating of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Insufficient Information2
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2025Break; Break
DURING THE PROCEDURE, WHILE THE SPECIALIST WAS PERFORMING THE DRILLING, THE DRILL BIT BROKE. THE DRILL BIT ALONG WITH ITS FRAGMENTS WERE REMOVED. ANOTHER DRILL BIT WITH THE SAME CHARACTERISTICS WAS PASSED TO THE DOCTOR, THUS SUCCESSFULLY COMPLETING THE SURGERY. THERE WAS NO HARM TO THE PATIENT AND THERE WERE NO ALTERATIONS IN SURGICAL TIMES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23066891 ↗MalfunctionFeb 29, 2024Break; Failure to Cut; Break; Failure to Cut
THE SURGERY WAS COMPLETED SUCCESSFULLY. NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED AND THE PATIENT OUTCOME WAS REPORTED TO BE STABLE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 18807715 ↗MalfunctionFeb 29, 2024Break; Failure to Cut; Break; Failure to Cut
THE SURGERY WAS COMPLETED SUCCESSFULLY. NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED AND THE PATIENT OUTCOME WAS REPORTED TO BE STABLE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 18807713 ↗MalfunctionOct 4, 2021Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE DRILL BIT WAS BROKEN. IT IS UNKNOWN IF THERE WAS SURGICAL OR PATIENT INVOLVEMENT. NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR ONE (1) MATRIXMANDIBLE 1.5MM DRILL BIT J-LATCH W/8MM STOP/50MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12569835 ↗MalfunctionFeb 4, 2021Failure to Cut; Failure to Cut
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2021, THE 1.5MM DRILL BIT WAS USED TO FIX THE 2.0 SCREWS TO THE PLATE, HOWEVER 1.5MM DRILL BIT WAS DULL. ANOTHER DRILL BIT OF THE SAME DIAMETER WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO SURGICAL DELAY REPORTED AND THERE WAS NO HARM OR ADVERSE EVENT TO THE PATIENT. CONCOMITANT DEVICES REPORTED: UNKNOWN SCREWS (PART #: UNKNOWN, LOT #: UNKNOWN, QUANTITY: UNKNOWN). THIS REPORT IS FOR ONE (1) MATRIXMANDIBLE 1.5MM DRILL BIT J-LATCH W/8MM STOP/50MM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11282067 ↗MalfunctionJan 17, 2020Failure to Cut; Failure to Cut
UPDATE: THIS REPORT IS 1 OF 6 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9604480 ↗MalfunctionDec 28, 2018Overheating of Device; Overheating of Device
DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2018, DURING THE INTER-MOLAR DISTRACTION OSTEOGENESIS (IMDO) PROCEDURE, TWO (2) TRUMATCH MANDIBLE GUIDE WERE HEATING UP DURING PRE-DRILLING. THE CYLINDER AND GUIDES WERE HEATING UP DUE TO THE DRILL. IT WAS NOTED THREE (3) DRILL BITS WERE TOUCHING THE CYLINDER BECAUSE OF THE ANGULATION OF THE GUIDE. THERE WAS NO SURGICAL DELAY. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT AND THERE WAS NO ISSUE WITH THE PLATES OR SCREWS. CONCOMITANT DEVICES REPORTED: MANDIBLE PLATE (PART/LOT UNKNOWN, QUANTITY UNKNOWN), MANDIBLE SCREW (PART/LOT UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR A DRILL BIT. THIS IS REPORT 1 OF 3 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8202839 ↗
FDA clearance and classification
Cleared through 510(k) K043310 on Jan 5, 2005 (decision: Substantially Equivalent). Submitted by Synthes (Usa) as “Synthes (USA) Electric Pen Drive System”. The FDA review took 35 days. FDA record ↗
The FDA regulates product code DZI (drill, bone, powered) under 21 CFR 872.4120 as a Class II device. 21 CFR 872.4120 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K043310. One submission often covers many models, so check the product and lot details in each record.
- Z-0681-2017 · Nov 1, 2016 · TerminatedSynthes 0.76 mm Drill Bit/Stryker J-Latch/with 5mm Stop/44.5 mm Non-Sterile
- Z-2784-2016 · Jun 24, 2016 · TerminatedAdaptor for Small Battery Drive and Small Battery Drive II. Product Usage: The Light Adaptor for Small Battery Drive (SBD) /Small Battery Drive II (SBD II) is designed for use in
- Z-2783-2016 · Jun 24, 2016 · TerminatedLight Adaptor for Small Battery Drive and Small Battery Drive II.
- Z-2199-2016 · May 31, 2016 · TerminatedPart # 313.250, 1.1mm Drill Bit, Mini QC with 4mm Stop/44.5mm, lot # 4326989, mfg. 9/28/2001 Product Usage: The DePuy Synthes MR Safe Instruments and MR conditional Instruments ar
- Z-1313-2015 · Dec 19, 2014 · TerminatedSynthes Electric Pen Drive (EPD) System, The Synthes Electric Pen Drive is indicated for screw insertion, pin and wire replacement, cutting of bone and metal, drilling, decorticati
- Z-0615-2014 · Apr 15, 2013 · TerminatedSynthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicated for screw insertion, pin and wire replacement, cutting of bone and met
Common questions about the MatrixMANDIBLE 03.503.408
Has the MatrixMANDIBLE 03.503.408 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10887587012889), based on the FDA recall file dated Nov 27, 2026. However, 6 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the MatrixMANDIBLE 03.503.408?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Dec 28, 2018 and Sep 16, 2025. In those reports, the most common specific device problems are “Break” (4), “Failure to Cut” (4) and “Overheating of Device” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the MatrixMANDIBLE 03.503.408?
None of the reports are classified as deaths, 0 as injuries and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the MatrixMANDIBLE 03.503.408 FDA approved?
It was cleared through 510(k) K043310 on Jan 5, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the MatrixMANDIBLE 03.503.408, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand MatrixMANDIBLE, model 03.503.408). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other DRILL, BONE, POWERED devices
- NA 03.505.004Synthes GmbH · DRILL, BONE, POWERED69 reports
- Nobel Biocare® 37470Nobel Biocare AB · DRILL, BONE, POWERED1 recall0 reports
- MatrixMANDIBLE 03.503.476Synthes GmbH · DRILL, BONE, POWERED26 reports
- NA 310.510Synthes GmbH · DRILL, BONE, POWERED17 reports
- NA 317.82Synthes GmbH · DRILL, BONE, POWERED16 reports
- MatrixMANDIBLE 03.503.479Synthes GmbH · DRILL, BONE, POWERED11 reports
- NA 310.520Synthes GmbH · DRILL, BONE, POWERED9 reports
- NA 05.001.120Synthes GmbH · DRILL, BONE, POWERED8 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page