Synthes GmbH · SHAFT FOR 90° SCREWDRIVER · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Synthes right-angle screwdriver shaft is a part of a surgical bone screwdriver. Its 142 reports often describe shafts that did not fit other instrument parts, and some describe breaks, bent parts or pieces coming loose. There are no deaths and 3 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
154
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 154 adverse event reports that list this exact device (DI 10887587013299), received between Mar 3, 2017 and Sep 29, 2026 — classified as 3 injuries, 151 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (67 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 13 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility67
  • Break48
  • Material Twisted/Bent16
  • Detachment of Device or Device Component12
  • Device Slipped12
  • Contamination /Decontamination Problem8
  • Naturally Worn3
  • Appropriate Device Problem Term/Code Not Available2

Patient problems

  • No Clinical Signs, Symptoms or Conditions105
  • No Consequences Or Impact To Patient22
  • No Patient Involvement12
  • Insufficient Information5
  • No Code Available5
  • No Known Impact Or Consequence To Patient3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 25 most recent of 154 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Break; Device Slipped

    It was reported that no patient was involved. There were little pieces in the 90-degree screwdriver.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934704 ↗
  • MalfunctionSep 29, 2026Break; Device Slipped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934693 ↗
  • MalfunctionSep 28, 2026Break; Device Slipped

    It was reported that no patient was involved. There were little pieces in the 90-degree screwdriver.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923443 ↗
  • MalfunctionSep 28, 2026Break; Device Slipped

    It was reported that no patient was involved. There were little pieces in the 90-degree screwdriver.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923432 ↗
  • MalfunctionSep 28, 2026Break; Device Slipped

    It was reported that no patient was involved. There were little pieces in the 90-degree screwdriver.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923431 ↗
  • MalfunctionSep 28, 2026Break; Device Slipped

    It was reported that no patient was involved. There were little pieces in the 90-degree screwdriver.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923425 ↗
  • MalfunctionSep 28, 2026Break; Device Slipped

    It was reported that no patient was involved. There were little pieces in the 90-degree screwdriver.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923421 ↗
  • MalfunctionSep 28, 2026Break; Device Slipped

    It was reported that no patient was involved. There were little pieces in the 90-degree screwdriver.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923413 ↗
  • MalfunctionSep 28, 2026Break; Device Slipped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923361 ↗
  • MalfunctionSep 28, 2026Break; Device Slipped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923302 ↗
  • MalfunctionSep 24, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50890122 ↗
  • MalfunctionSep 15, 2026Break

    It was reported that the 90 degree screwdriver shafts are broken and 90 degree blades are broken.

    Patient problems: Insufficient Information

    MAUDE report 50771545 ↗
  • MalfunctionSep 15, 2026Break

    It was reported that the 90 degree screwdriver shafts are broken and 90 degree blades are broken.

    Patient problems: Insufficient Information

    MAUDE report 50771543 ↗
  • MalfunctionMar 6, 2026Break; Device-Device Incompatibility

    IT WAS REPORTED THAT SHAFT FOR 90 SCREWDRIVER AND MATRIXRIB SCRDRVR BLD SLF-RT 90 SCRWDRVR EXPERIENCED FAILURES IN THE USA. SPECIFICALLY, TWO 90-DEGREE COLLETS AND TWO 90-DEGREE BLADES WERE FOUND BROKEN. THE COLLETS WOULD NOT HOLD THE SCREWDRIVER BLADES, AND THE BLADES WOULD NOT HOLD THE RIB SCREWS AS INTENDED. THERE WAS NO PATIENT INVOLVEMENT OR CLINICAL CONSEQUENCE ASSOCIATED WITH THIS EVENT, AS THE ISSUE OCCURRED OUTSIDE OF PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24531304 ↗
  • MalfunctionMar 6, 2026Break; Device-Device Incompatibility

    IT WAS REPORTED THAT SHAFT FOR 90 SCREWDRIVER AND MATRIXRIB SCRDRVR BLD SLF-RT 90 SCRWDRVR EXPERIENCED FAILURES IN THE USA. SPECIFICALLY, TWO 90-DEGREE COLLETS AND TWO 90-DEGREE BLADES WERE FOUND BROKEN. THE COLLETS WOULD NOT HOLD THE SCREWDRIVER BLADES, AND THE BLADES WOULD NOT HOLD THE RIB SCREWS AS INTENDED. THERE WAS NO PATIENT INVOLVEMENT OR CLINICAL CONSEQUENCE ASSOCIATED WITH THIS EVENT, AS THE ISSUE OCCURRED OUTSIDE OF PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24531295 ↗
  • MalfunctionJul 28, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT WHEN REASSEMBLING THE 90-DEGREE SCREWDRIVER, THE STAFF TOLD ME THAT THE DRIVE SHAFT WAS BROKEN AND THE BALL BEARINGS WERE FALLING OUT. THEY DISPOSED OF THE BROKEN DRIVE SHAFT BUT KEPT THE OUTER SLEEVE. NOTHING BROKEN WAS NOTICED DURING THE SURGERY. THERE WAS NO ADVERSE EVENT OCCURRED AND NO PATIENT CONSEQUENCES AND NO REVISION SURGERY REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22629773 ↗
  • MalfunctionApr 25, 2025Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING PRODUCT TESTING AFTER THE SET CAME THROUGH THE WASH, IT WAS FOUND THAT THE GEARS IN THE GOLD HANDLE OF THE 90 SCREWDRIVER WERE GRINDING ALONG WITH THE INNER SHAFT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21909561 ↗
  • MalfunctionApr 25, 2025Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING PRODUCT TESTING AFTER THE SET CAME THROUGH THE WASH, IT WAS FOUND THAT THE GEARS IN THE GOLD HANDLE OF THE 90 SCREWDRIVER WERE GRINDING ALONG WITH THE INNER SHAFT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21909556 ↗
  • MalfunctionOct 15, 2024Contamination /Decontamination Problem; Device-Device Incompatibility; Contamination /Decontamination Problem; Device-Device Incompatibility

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024. THE MINIMALLY INVASIVE PLATE OSTEOSYNTHESIS (MIPO) SYSTEM DID NOT WORK PROPERLY IN SURGERY AND WAS STICKING. THE ASSISTANT, WHEN REVIEWING THE PIECES AFTER SURGERY, FOUND THAT THE MOVING PARTS WERE UNLUBRICATED AND WITH DRY BLOOD RESIDUES. AS THE SYSTEM CONTAINS TWO OF EACH ITEM, OTHER DEVICES WERE AVAILABLE FOR USE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20450800 ↗
  • MalfunctionJun 26, 2024Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE 90 DEGREE SCREWDRIVER SHAFT BROKE. THE PIECE THAT HOLDS THE BLADE BROKE OFF FROM THE ACTUAL SHAFT. PROCEDURE AND PATIENT OUTCOME ARE UNKNOWN. THIS REPORT IS FOR A SHAFT FOR 90° SCREWDRIVER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 19613707 ↗
  • MalfunctionMay 2, 2024Break; Break

    IT WAS REPORTED THAT ON UNKNOWN DATE THE 90 DEGREE SCREWDRIVER SHAFT WAS BROKE. THE PIECE THAT HOLDS THE BLADE BROKE OFF FROM THE ACTUAL SHAFT. THERE IS NO MORE INFORMATION PROVIDED. THIS REPORT IS FOR ONE (1) SHAFT FOR 90 SCREWDRIVER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 19232668 ↗
  • MalfunctionMar 25, 2024Break; Break

    IT WAS FURTHER REPORTED THAT THE EVENT OCCURRED POSTOPERATIVELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18967759 ↗
  • MalfunctionMar 14, 2024Mechanical Jam; Mechanical Jam

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT DURING A SAGITTAL SPLIT RAMUS OSTEOTOMY (SSRO) PERFORMED ON (B)(6) 2023, THE INSTRUMENTS IN QUESTION WERE ASSEMBLED AND ROTATED AS A SCREWDRIVER, HOWEVER THEY DID NOT ROTATE. OTHER INSTRUMENTS WERE ASSEMBLED AS A DRILL, AND THEY WERE USED AS A SCREWDRIVER INSTEAD OF THE INSTRUMENTS IN QUESTION. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN THIRTY (30) MINUTES DELAY. PATIENT OUTCOME IS REPORTED AS STABLE. NO FURTHER INFORMATION IS AVAILABLE. DURING MANUFACTURER'S INVESTIGATION OF THE RETURNED DEVICE IT WAS IDENTIFIED THAT A FOREIGN UNKNOWN SUBSTANCE WAS ON THE FLUTED TIP OF INNER SHAFT. THIS REPORT IS FOR ONE (1) SHAFT FOR 90° SCREWDRIVER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18907477 ↗
  • MalfunctionDec 14, 2023Break; Noise, Audible; Break; Noise, Audible

    IT WAS REPORTED THAT ON NOVEMBER 22, 2023 THAT, THE SALES REPRESENTATIVE HAVE A 90 DEGREE SCREWDRIVER SHAFT THAT BROKE. IT MAKES A SOUND WHEN ATTACHED TO THE PRO 2 DRIVER AND IS NOT WORKING PROPERLY. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE ((B)(6)) SHAFT FOR 90 SCREWDRIVER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18329354 ↗
  • MalfunctionNov 1, 2023Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT DURING SURGERY ON (B)(6), 2023, THE BEARING AT THE SHAFT TIP BROKE IN HALF. THE SHAFT IN QUESTION HAD BEEN USED SINCE IT HAD BEEN PURCHASED IN EARLY JULY. THERE WAS AN UNUSUAL NOISE DURING THE SURGERY, BUT SINCE THERE WERE TWO SHAFTS, THE REMAINING ONE WAS USED, AND THE OPERATION CONTINUED. IT WAS DISCOVERED THAT THE SHAFT WAS BROKEN AFTER WASHING. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY DELAY. THE PATIENT OUTCOME WAS REPORTED TO BE STABLE. THIS REPORT INVOLVES ONE SHAFT FOR 90° SCREWDRIVER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18046641 ↗

Other DRIVER, WIRE, AND BONE DRILL, MANUAL devices

All DRIVER, WIRE, AND BONE DRILL, MANUAL devices →