MatrixMIDFACE 04.503.206.01Recalls & reported problems

Synthes GmbH · TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM · Class 2 · Implant

All models of MatrixMIDFACE (40)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Synthes MatrixMIDFACE self-tapping screw is used to hold bone plates in place, mainly in the jaw and face. The 30 reports mostly describe screws breaking during surgery, and screws loosening or backing out of the jawbone so plates and screws had to be removed, sometimes with discomfort and localized swelling. There are 8 reports of injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
30
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 10887587018812), received between Jul 16, 2018 and Oct 30, 2024 — classified as 8 injuries, 22 malfunctions.

The most frequently reported specific device problem is “Break” (22 reports).

2 other recalls cover devices cleared under the same FDA submission (K050608, K083388); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break22
  • Migration8

Patient problems

  • No Clinical Signs, Symptoms or Conditions21
  • Discomfort8
  • Swelling/ Edema8
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 30, 2024Break; Break

    IT WAS REPORTED ON AN UNKNOWN DATE THAT THE MATRIX MIDFACE SCREW BROKE. IT IS ALSO UNKNOWN IF SURGERY WAS DELAYED, FRAGMENTS WERE GENERATED OR IF THE PROCEDURE WAS SUCCESSFULLY COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20567364 ↗
  • InjurySep 12, 2023Migration; Migration

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, TWO MANDIBLE PLATES WERE REMOVED FROM THE PATIENT. THE SCREWS, AND THEREFORE THE PLATES, HAD BECOME LOOSE. SOME OF THE SCREWS HAD BACKED OUT FROM THE MANDIBLE. THE WHOLE CONSTRUCT WAS REMOVED, INCLUDING TWO PLATES AND EIGHT SCREWS. THE PATIENT EXPERIENCED SOME MILD DISCOMFORT AND LOCALIZED SWELLING. WHEN THE PLATES AND SCREWS WERE REMOVED, BICORTICAL SCREWS WERE INSERTED TO FIXATE THE MANDIBLE. THE IMPLANTS HAD BEEN PLACED ON (B)(6) 2023. THE REMOVAL SURGERY WAS COMPLETED SUCCESSFULLY. IT IS FELT BY THE SURGEONS THEY ARE LIKELY GETTING AN ELEMENT OF THERMAL NECROSIS IN THE BONE DUE TO THE POWER TOOL (NONE SYNTHES PRODUCT) THEY ARE USING. THIS IS CAUSING THE LOSS OF BONE AND THEREFORE THE SCREWS ARE BACKING OUT. THIS REPORT INVOLVES ONE TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 3 OF 9 FOR (B)(4).

    Patient problems: Discomfort; Swelling/ Edema

    MAUDE report 17729057 ↗
  • InjurySep 12, 2023Migration; Migration

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, TWO MANDIBLE PLATES WERE REMOVED FROM THE PATIENT. THE SCREWS, AND THEREFORE THE PLATES, HAD BECOME LOOSE. SOME OF THE SCREWS HAD BACKED OUT FROM THE MANDIBLE. THE WHOLE CONSTRUCT WAS REMOVED, INCLUDING TWO PLATES AND EIGHT SCREWS. THE PATIENT EXPERIENCED SOME MILD DISCOMFORT AND LOCALIZED SWELLING. WHEN THE PLATES AND SCREWS WERE REMOVED, BICORTICAL SCREWS WERE INSERTED TO FIXATE THE MANDIBLE. THE IMPLANTS HAD BEEN PLACED ON (B)(6) 2023. THE REMOVAL SURGERY WAS COMPLETED SUCCESSFULLY. IT IS FELT BY THE SURGEONS THEY ARE LIKELY GETTING AN ELEMENT OF THERMAL NECROSIS IN THE BONE DUE TO THE POWER TOOL (NONE SYNTHES PRODUCT) THEY ARE USING. THIS IS CAUSING THE LOSS OF BONE AND THEREFORE THE SCREWS ARE BACKING OUT. THIS REPORT INVOLVES ONE TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 6 OF 9 FOR (B)(4).

    Patient problems: Discomfort; Swelling/ Edema

    MAUDE report 17729055 ↗
  • InjurySep 12, 2023Migration; Migration

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, TWO MANDIBLE PLATES WERE REMOVED FROM THE PATIENT. THE SCREWS, AND THEREFORE THE PLATES, HAD BECOME LOOSE. SOME OF THE SCREWS HAD BACKED OUT FROM THE MANDIBLE. THE WHOLE CONSTRUCT WAS REMOVED, INCLUDING TWO PLATES AND EIGHT SCREWS. THE PATIENT EXPERIENCED SOME MILD DISCOMFORT AND LOCALIZED SWELLING. WHEN THE PLATES AND SCREWS WERE REMOVED, BICORTICAL SCREWS WERE INSERTED TO FIXATE THE MANDIBLE. THE IMPLANTS HAD BEEN PLACED ON (B)(6) 2023. THE REMOVAL SURGERY WAS COMPLETED SUCCESSFULLY. IT IS FELT BY THE SURGEONS THEY ARE LIKELY GETTING AN ELEMENT OF THERMAL NECROSIS IN THE BONE DUE TO THE POWER TOOL (NONE SYNTHES PRODUCT) THEY ARE USING. THIS IS CAUSING THE LOSS OF BONE AND THEREFORE THE SCREWS ARE BACKING OUT. THIS REPORT INVOLVES ONE TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 5 OF 9 FOR (B)(4).

    Patient problems: Discomfort; Swelling/ Edema

    MAUDE report 17729054 ↗
  • InjurySep 12, 2023Migration; Migration

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, TWO MANDIBLE PLATES WERE REMOVED FROM THE PATIENT. THE SCREWS, AND THEREFORE THE PLATES, HAD BECOME LOOSE. SOME OF THE SCREWS HAD BACKED OUT FROM THE MANDIBLE. THE WHOLE CONSTRUCT WAS REMOVED, INCLUDING TWO PLATES AND EIGHT SCREWS. THE PATIENT EXPERIENCED SOME MILD DISCOMFORT AND LOCALIZED SWELLING. WHEN THE PLATES AND SCREWS WERE REMOVED, BICORTICAL SCREWS WERE INSERTED TO FIXATE THE MANDIBLE. THE IMPLANTS HAD BEEN PLACED ON (B)(6) 2023. THE REMOVAL SURGERY WAS COMPLETED SUCCESSFULLY. IT IS FELT BY THE SURGEONS THEY ARE LIKELY GETTING AN ELEMENT OF THERMAL NECROSIS IN THE BONE DUE TO THE POWER TOOL (NONE SYNTHES PRODUCT) THEY ARE USING. THIS IS CAUSING THE LOSS OF BONE AND THEREFORE THE SCREWS ARE BACKING OUT. THIS REPORT INVOLVES ONE TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 4 OF 9 FOR (B)(4).

    Patient problems: Discomfort; Swelling/ Edema

    MAUDE report 17729050 ↗
  • InjurySep 12, 2023Migration; Migration

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, TWO MANDIBLE PLATES WERE REMOVED FROM THE PATIENT. THE SCREWS, AND THEREFORE THE PLATES, HAD BECOME LOOSE. SOME OF THE SCREWS HAD BACKED OUT FROM THE MANDIBLE. THE WHOLE CONSTRUCT WAS REMOVED, INCLUDING TWO PLATES AND EIGHT SCREWS. THE PATIENT EXPERIENCED SOME MILD DISCOMFORT AND LOCALIZED SWELLING. WHEN THE PLATES AND SCREWS WERE REMOVED, BICORTICAL SCREWS WERE INSERTED TO FIXATE THE MANDIBLE. THE IMPLANTS HAD BEEN PLACED ON (B)(6) 2023. THE REMOVAL SURGERY WAS COMPLETED SUCCESSFULLY. IT IS FELT BY THE SURGEONS THEY ARE LIKELY GETTING AN ELEMENT OF THERMAL NECROSIS IN THE BONE DUE TO THE POWER TOOL (NONE SYNTHES PRODUCT) THEY ARE USING. THIS IS CAUSING THE LOSS OF BONE AND THEREFORE THE SCREWS ARE BACKING OUT. THIS REPORT INVOLVES ONE TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 7 OF 9 FOR (B)(4).

    Patient problems: Discomfort; Swelling/ Edema

    MAUDE report 17729048 ↗
  • InjurySep 12, 2023Migration; Migration

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, TWO MANDIBLE PLATES WERE REMOVED FROM THE PATIENT. THE SCREWS, AND THEREFORE THE PLATES, HAD BECOME LOOSE. SOME OF THE SCREWS HAD BACKED OUT FROM THE MANDIBLE. THE WHOLE CONSTRUCT WAS REMOVED, INCLUDING TWO PLATES AND EIGHT SCREWS. THE PATIENT EXPERIENCED SOME MILD DISCOMFORT AND LOCALIZED SWELLING. WHEN THE PLATES AND SCREWS WERE REMOVED, BICORTICAL SCREWS WERE INSERTED TO FIXATE THE MANDIBLE. THE IMPLANTS HAD BEEN PLACED ON (B)(6) 2023. THE REMOVAL SURGERY WAS COMPLETED SUCCESSFULLY. IT IS FELT BY THE SURGEONS THEY ARE LIKELY GETTING AN ELEMENT OF THERMAL NECROSIS IN THE BONE DUE TO THE POWER TOOL (NONE SYNTHES PRODUCT) THEY ARE USING. THIS IS CAUSING THE LOSS OF BONE AND THEREFORE THE SCREWS ARE BACKING OUT. THIS REPORT INVOLVES ONE TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 8 OF 9 FOR (B)(4).

    Patient problems: Discomfort; Swelling/ Edema

    MAUDE report 17729046 ↗
  • InjurySep 12, 2023Migration; Migration

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, TWO MANDIBLE PLATES WERE REMOVED FROM THE PATIENT. THE SCREWS, AND THEREFORE THE PLATES, HAD BECOME LOOSE. SOME OF THE SCREWS HAD BACKED OUT FROM THE MANDIBLE. THE WHOLE CONSTRUCT WAS REMOVED, INCLUDING TWO PLATES AND EIGHT SCREWS. THE PATIENT EXPERIENCED SOME MILD DISCOMFORT AND LOCALIZED SWELLING. WHEN THE PLATES AND SCREWS WERE REMOVED, BICORTICAL SCREWS WERE INSERTED TO FIXATE THE MANDIBLE. THE IMPLANTS HAD BEEN PLACED ON (B)(6) 2023. THE REMOVAL SURGERY WAS COMPLETED SUCCESSFULLY. IT IS FELT BY THE SURGEONS THEY ARE LIKELY GETTING AN ELEMENT OF THERMAL NECROSIS IN THE BONE DUE TO THE POWER TOOL (NONE SYNTHES PRODUCT) THEY ARE USING. THIS IS CAUSING THE LOSS OF BONE AND THEREFORE THE SCREWS ARE BACKING OUT. THIS REPORT INVOLVES ONE TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 9 OF 9 FOR (B)(4).

    Patient problems: Discomfort; Swelling/ Edema

    MAUDE report 17729043 ↗
  • InjurySep 1, 2023Migration; Migration

    THIS IS REPORT 1 OF 9 FOR COMPLAINT (B)(4).

    Patient problems: Discomfort; Swelling/ Edema

    MAUDE report 17669746 ↗
  • MalfunctionOct 4, 2022Break; Break

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, WHILE PLACING THE CORTICAL SCREWS IN QUESTION THEY WOULD FRAGMENT AND DISINTEGRATE. WHEN FORCE WAS APPLIED TO THE SCREWS, THEY WOULD BREAK APART WITHOUT PROVIDING ANY RESISTANCE AT ALL. NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR A TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 9 OF 9 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15538034 ↗
  • MalfunctionOct 4, 2022Break; Break

    DEVICE REPORT FROM URUGUAY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, WHILE PLACING THE CORTICAL SCREWS IN QUESTION THEY WOULD FRAGMENT AND DISINTEGRATE. WHEN FORCE WAS APPLIED TO THE SCREWS, THEY WOULD BREAK APART WITHOUT PROVIDING ANY RESISTANCE AT ALL. THIS REPORT IS FOR A TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 5 OF 9 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15538026 ↗
  • MalfunctionOct 4, 2022Break; Break

    DEVICE REPORT FROM URUGUAY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, WHILE PLACING THE CORTICAL SCREWS IN QUESTION THEY WOULD FRAGMENT AND DISINTEGRATE. WHEN FORCE WAS APPLIED TO THE SCREWS, THEY WOULD BREAK APART WITHOUT PROVIDING ANY RESISTANCE AT ALL. THIS REPORT IS FOR A TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 8 OF 9 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15538022 ↗
  • MalfunctionOct 4, 2022Break; Break

    DEVICE REPORT FROM URUGUAY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, WHILE PLACING THE CORTICAL SCREWS IN QUESTION THEY WOULD FRAGMENT AND DISINTEGRATE. WHEN FORCE WAS APPLIED TO THE SCREWS, THEY WOULD BREAK APART WITHOUT PROVIDING ANY RESISTANCE AT ALL. THIS REPORT IS FOR A TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 1 OF 9 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15538020 ↗
  • MalfunctionOct 4, 2022Break; Break

    DEVICE REPORT FROM URUGUAY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, WHILE PLACING THE CORTICAL SCREWS IN QUESTION THEY WOULD FRAGMENT AND DISINTEGRATE. WHEN FORCE WAS APPLIED TO THE SCREWS, THEY WOULD BREAK APART WITHOUT PROVIDING ANY RESISTANCE AT ALL. THIS REPORT IS FOR A TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 4 OF 9 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15538017 ↗
  • MalfunctionOct 4, 2022Break; Break

    DEVICE REPORT FROM URUGUAY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, WHILE PLACING THE CORTICAL SCREWS IN QUESTION THEY WOULD FRAGMENT AND DISINTEGRATE. WHEN FORCE WAS APPLIED TO THE SCREWS, THEY WOULD BREAK APART WITHOUT PROVIDING ANY RESISTANCE AT ALL. NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR A TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 3 OF 9 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15538014 ↗
  • MalfunctionOct 4, 2022Break; Break

    THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THE PATIENT IS IN GOOD CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15538013 ↗
  • MalfunctionOct 4, 2022Break; Break

    THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THE PATIENT IS IN GOOD CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15538010 ↗
  • MalfunctionOct 4, 2022Break; Break

    DEVICE REPORT FROM URUGUAY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, WHILE PLACING THE CORTICAL SCREWS IN QUESTION THEY WOULD FRAGMENT AND DISINTEGRATE. WHEN FORCE WAS APPLIED TO THE SCREWS, THEY WOULD BREAK APART WITHOUT PROVIDING ANY RESISTANCE AT ALL. THIS REPORT IS FOR A TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 7 OF 9 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15538009 ↗
  • MalfunctionOct 4, 2022Break; Break

    DEVICE REPORT FROM URUGUAY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, WHILE PLACING THE CORTICAL SCREWS IN QUESTION THEY WOULD FRAGMENT AND DISINTEGRATE. WHEN FORCE WAS APPLIED TO THE SCREWS, THEY WOULD BREAK APART WITHOUT PROVIDING ANY RESISTANCE AT ALL. THIS REPORT IS FOR A TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 7 OF 9 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15537870 ↗
  • MalfunctionOct 4, 2022Break; Break

    DEVICE REPORT FROM URUGUAY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, WHILE PLACING THE CORTICAL SCREWS IN QUESTION THEY WOULD FRAGMENT AND DISINTEGRATE. WHEN FORCE WAS APPLIED TO THE SCREWS, THEY WOULD BREAK APART WITHOUT PROVIDING ANY RESISTANCE AT ALL. THIS REPORT IS FOR A TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 3 OF 9 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15537867 ↗
  • MalfunctionOct 4, 2022Break; Break

    DEVICE REPORT FROM URUGUAY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, WHILE PLACING THE CORTICAL SCREWS IN QUESTION THEY WOULD FRAGMENT AND DISINTEGRATE. WHEN FORCE WAS APPLIED TO THE SCREWS, THEY WOULD BREAK APART WITHOUT PROVIDING ANY RESISTANCE AT ALL. THIS REPORT IS FOR A TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 8 OF 9 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15537859 ↗
  • MalfunctionOct 4, 2022Break; Break

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, WHILE PLACING THE CORTICAL SCREWS IN QUESTION THEY WOULD FRAGMENT AND DISINTEGRATE. WHEN FORCE WAS APPLIED TO THE SCREWS, THEY WOULD BREAK APART WITHOUT PROVIDING ANY RESISTANCE AT ALL. NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR A TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 9 OF 9 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15537857 ↗
  • MalfunctionOct 4, 2022Break; Break

    DEVICE REPORT FROM URUGUAY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT (B)(6) 2022, WHILE PLACING THE CORTICAL SCREWS IN QUESTION THEY WOULD FRAGMENT AND DISINTEGRATE. WHEN FORCE WAS APPLIED TO THE SCREWS, THEY WOULD BREAK APART WITHOUT PROVIDING ANY RESISTANCE AT ALL. NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR A TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 5 OF 9 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15537853 ↗
  • MalfunctionOct 4, 2022Break; Break

    DEVICE REPORT FROM URUGUAY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, WHILE PLACING THE CORTICAL SCREWS IN QUESTION THEY WOULD FRAGMENT AND DISINTEGRATE. WHEN FORCE WAS APPLIED TO THE SCREWS, THEY WOULD BREAK APART WITHOUT PROVIDING ANY RESISTANCE AT ALL. NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR A TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 6 OF 9 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15537852 ↗
  • MalfunctionOct 4, 2022Break; Break

    DEVICE REPORT FROM URUGUAY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, WHILE PLACING THE CORTICAL SCREWS IN QUESTION THEY WOULD FRAGMENT AND DISINTEGRATE. WHEN FORCE WAS APPLIED TO THE SCREWS, THEY WOULD BREAK APART WITHOUT PROVIDING ANY RESISTANCE AT ALL. NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR A TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM. THIS IS REPORT 1 OF 9 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15537851 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K050608, K083388. One submission often covers many models, so check the product and lot details in each record.

  • Z-2739-2020 · Jun 16, 2020 · Terminated
    MatrixMIDFACE Screws - Product Usage: The intended use for this device is in fixation of cranial bones and facial skeleton in procedures such as craniotomies, reconstructive proced
  • Z-2272-2015 · Jun 24, 2015 · Terminated
    Synthes TI Matrix Pre-Bent Maxillary Plate / 2mm ADVMNT/0.8mm THK/L and Synthes TI Matrix Pre-Bent Maxillary Plate / 2mm ADVMNT/0.8mm THK/R. The Synthes Matrix ORTHOGNATHIC Fixati

Other PLATE, BONE devices

All PLATE, BONE devices →