Skip to content

MatrixMIDFACE 04.503.355Recalls & reported problems

Synthes GmbH · TI MATRIXMIDFACE OBLIQUE L-PL 3X4 HOLES-LEFT/0.7MM THICK · Class 2 · Implant

All models of MatrixMIDFACE (40)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10887587020037), received between Oct 14, 2024 and Oct 14, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

1 other recall cover devices cleared under the same FDA submission (K050608); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Migration1

Patient problems

  • Obstruction/Occlusion1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 14, 2024Break; Migration; Break; Migration

    IT WAS REPORTED THAT ON (B)(6) 2020 THE PATIENT UNDERWENT NASAL RECONSTRUCTIVE SURGERY. ONE SYNTHES, L-SHAPE AND ONE SYNTHES, STRAIGHT 0.5MM, MULTI-HOLE TITANIUM MINI PLATES WERE USED IN HER SURGERY. ONE OF THE MINI-PLATES WAS SCREW FIXATED TO THE RIGHT NASAL BONE WITH THE MINI-PLATE EXTENDING SUBCUTANEOUSLY TO THE RIGHT EXTERNAL NASAL VALVE/NOSTRIL, FOR SUPPORT OF THE RIGHT INTERNAL AND EXTERNAL VALVES. THE OTHER MINI-PLATE WAS SUTURE-SECURED TO THE CARTILAGINOUS DORSAL AND CAUDAL NASAL SEPTUM, SUBMUCOSALLY, FOR SUPPORT OF THE NASAL DORSUM AND TIP. THE PATIENT DID WELL POSTOPERATIVELY UNTIL APPROXIMATELY (B)(6) 2020, WHEN EXCESSIVE MOBILITY OF THE RIGHT SIDE MINI-PLATE AT THE SCREW FIXATION SITES ON THE RIGHT NASAL BONE WAS DISCOVERED. REVISION SURGERY WITH RE-FIXATION OF THE RIGHT SIDE MINI-PLATE TO THE NASAL BONE WAS PERFORMED ON (B)(6) 2021. THE PATIENT DID WELL WITH NORMAL, COMFORTABLE BREATHING UNTIL (B)(6) 2024, WHEN THE RIGHT-SIDE NASAL OBSTRUCTION RECURRED DUE TO VALVE COLLAPSE. THE PATIENT SUBSEQUENTLY UNDERWENT DENTAL X-RAYS FOR A SEPARATE ISSUE THAT REVEALED MINI-PLATE DISCONTINUITIES. DETAILED X-RAYS OF THE NASAL BONES WERE OBTAINED ON (B)(6) 2024, SHOWING COMPLETE DIS

    Patient problems: Obstruction/Occlusion

    MAUDE report 20439695 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K050608. One submission often covers many models, so check the product and lot details in each record.

  • Z-2739-2020 · Jun 16, 2020 · Terminated
    MatrixMIDFACE Screws - Product Usage: The intended use for this device is in fixation of cranial bones and facial skeleton in procedures such as craniotomies, reconstructive proced

Other PLATE, BONE devices

All PLATE, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page