MatrixMANDIBLE 04.503.414.01Recalls & reported problems
Synthes GmbH · 2.0MM TI MATRIXMANDIBLE SCREW SELF-TAPPING 14MM · Class 2 · Implant
All models of MatrixMANDIBLE (79)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10887587020686), received between Mar 21, 2023 and Oct 25, 2023 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Entrapment of Device1
Patient problems
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 25, 2023Break; Break
IT WAS REPORTED THAT ON (B)(6) 2023, THE HEAD OF THE 14MM SCREW IN QUESTION BROKE ON THE LEFT NATIVE MANDIBLE DURING APPLICATION OF A PLATE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. ALL FRAGMENTS GENERATED WERE ABLE TO BE REMOVED WITHOUT ADDITIONAL MEDICAL INTERVENTION. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. THIS REPORT IS FOR A 2.0MM TI MATRIXMANDIBLE SCREW SELF-TAPPING 14MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18004786 ↗InjuryMar 21, 2023Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED THAT ON (B)(6) 2023, DURING INTRA-OP THE SCREW BROKE HALFWAY DOWN THE LENGTH OF THE THREADS WHILE BEING INSERTED INTO THE BONE. THE BONE WAS PRE-DRILLED WITH THE PROPER DIAMETER DRILL BIT (1.5MM) FULL THICKNESS OF BOTH CORTEX. THE CASE WAS NOT DELAYED BUT WE WERE NOT ABLE TO USE THE PLATE HOLE IN WHICH THE SCREW BROKE SINCE IT WAS EMBEDDED INTO THE BONE. PATIENT CONSEQUENCES ARE UNKNOWN. THIS REPORT IS FOR ONE (1) 2.0 TI MATRIXMANDIBLE SCREW STPING 14. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 16586017 ↗
Other PLATE, BONE devices
- MatrixNEURO 04.503.104.01Synthes GmbH · PLATE, BONE252 reports
- RapidSorb 806.004.02SSynthes GmbH · PLATE, BONE146 reports
- TruMatch SD980.001Materialise NV · Plate, Bone1 recall21 reports
- Matrix 04.511.225.01Synthes GmbH · PLATE, BONE62 reports
- NA 400.834Synthes GmbH · PLATE, BONE58 reports
- TruMatch SD980.016Materialise NV · Plate, Bone55 reports
- ProPlan CMF SD900.101Materialise NV · Plate, Bone1 recall1 report
- TruMatch SD980.013Materialise NV · Plate, Bone1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page