Skip to content

Matrix 04.511.206.01Recalls & reported problems

Synthes GmbH · 1.85MM TI MATRIX SCREW SELF-TAPPING/6MM · Class 2 · Implant

Adverse event reports
11
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10887587023458), received between Aug 24, 2018 and Jun 18, 2026 — classified as 5 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K083388); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Device-Device Incompatibility2
  • Material Twisted/Bent2
  • Break1
  • Device Slipped1
  • Difficult to Advance1
  • Failure to Advance1
  • Failure to Cut1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Nonunion/Delayed-union Bone Fracture4
  • Insufficient Information2
  • No Known Impact Or Consequence To Patient1
  • Pain1
  • Swelling/ Edema1
  • Unspecified Infection1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 18, 2026Device Slipped

    IT WAS REPORTED THAT ON (B)(6) 2022 THE PATIENT UNDERWENT A PROCEDURE, IMPLANTING PLATES AND SCREWS IN THE MANDIBLE AND MAXILLA. ON (B)(6) 2022, THE PATIENT WAS ADMITTED TO THE HOSPITAL DUE TO PAIN, SWELLING AND DIAGNOSED WITH INFECTION. ON (B)(6) 2022 THE PATIENT UNDERWENT AN I&D PROCEDURE. DURING THAT PROCEDURE, IT WAS DISCOVERED A SCREW WAS LOOSE IN THE RIGHT MANDIBLE. THE LOOSE SCREW WAS REMOVED. ADDITIONALLY, ONE PLATE AND THREE OTHER SCREWS WERE REMOVED, BUT THERE WAS NO SPECIFIC ALLEGATIONS AGAINST THOSE DEVICES. ALL OTHER IMPLANTS REMAINED.

    Patient problems: Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 25548422 ↗
  • MalfunctionSep 9, 2024Break; Break

    A METAL STRIP WAS PEELED OFF WHEN A SCREW INSERTING ON A PLATE TO MAXILLA BONE. IT IS UNKNOWN IF IT CAME FROM THE PLATE OR THE SCREW. BOTH DEVICES WERE STILL IMPLANTED IN THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY OR IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20169239 ↗
  • MalfunctionNov 9, 2023Material Twisted/Bent; Material Twisted/Bent

    THIS REPORT IS RELATED TO (B)(4), WHICH CAPTURES ADDITIONAL IMPACTED PRODUCTS.

    Patient problems: Insufficient Information

    MAUDE report 18102512 ↗
  • MalfunctionNov 9, 2023Material Twisted/Bent; Material Twisted/Bent

    DEVICE REPORT FROM BRAZIL REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON 06-OCTOBER-2023, THE SCREWS IN QUESTION WERE HAVING TROUBLE ATTACHING TO THE SCREWDRIVER. DIFFERENT SCREWDRIVERS WERE USED IN AN ATTEMPT TO DETERMINE IF THE ISSUE WAS CAUSED BY THE SCREW OR SCREWDRIVER. IT WAS DETERMINED THAT THE SCREW HEADS WERE NOT MAGNETIZED AS USUAL. THERE WAS NO REPORTED ADVERSE PATIENT IMPACT. THE SCREWS WERE USED DURING THE PROCEDURE AS ORIGINALLY INTENDED, THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 1.85MM TI MATRIX SCREW SELF-TAPPING/6MM. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18102508 ↗
  • InjuryAug 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT WILL UNDERGO REVISION SURGERY ON AN UNKNOWN DATE, DUE TO UNSTABLE FIXATION AND MALOCCLUSION. PATIENT ORIGINALLY UNDERWENT AN ORTHOGNATHIC PROCEDURE INVOLVING THE UPPER AND LOWER JAW ON (B0(6) 2023. IT HAS BEEN DETERMINED BY THE SURGEON AND ORTHODONTIST THAT THE MAXILLARY SURGERY IS STABLE, HOWEVER, THE MANDIBULAR PORTION IS NOT AND MUST BE REVISED. THIS REPORT IS FOR A 1.85MM TI MATRIX SCREW SELF-TAPPING/6MM. THIS IS REPORT 5 OF 9 FOR (B)(4). ADDITIONAL REPORTS ARE CAPTURED UNDER (B)(4).

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 17450994 ↗
  • InjuryAug 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT WILL UNDERGO REVISION SURGERY ON AN UNKNOWN DATE, DUE TO UNSTABLE FIXATION AND MALOCCLUSION. PATIENT ORIGINALLY UNDERWENT AN ORTHOGNATHIC PROCEDURE INVOLVING THE UPPER AND LOWER JAW ON (B)(6), 2023. IT HAS BEEN DETERMINED BY THE SURGEON AND ORTHODONTIST THAT THE MAXILLARY SURGERY IS STABLE, HOWEVER, THE MANDIBULAR PORTION IS NOT AND MUST BE REVISED. THIS REPORT IS FOR A 1.85MM TI MATRIX SCREW SELF-TAPPING/6MM. THIS IS REPORT 4 OF 9 FOR (B)(4). ADDITIONAL REPORTS ARE CAPTURED UNDER (B)(4).

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 17450954 ↗
  • InjuryAug 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT WILL UNDERGO REVISION SURGERY ON AN UNKNOWN DATE, DUE TO UNSTABLE FIXATION AND MALOCCLUSION. PATIENT ORIGINALLY UNDERWENT AN ORTHOGNATHIC PROCEDURE INVOLVING THE UPPER AND LOWER JAW ON (B)(6) 2023. IT HAS BEEN DETERMINED BY THE SURGEON AND ORTHODONTIST THAT THE MAXILLARY SURGERY IS STABLE, HOWEVER, THE MANDIBULAR PORTION IS NOT AND MUST BE REVISED. THIS REPORT IS FOR A 1.85MM TI MATRIX SCREW SELF-TAPPING/6MM. THIS IS REPORT 3 OF 9 FOR (B)(4).

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 17450915 ↗
  • InjuryAug 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT WILL UNDERGO REVISION SURGERY ON AN UNKNOWN DATE, DUE TO UNSTABLE FIXATION AND MALOCCLUSION. PATIENT ORIGINALLY UNDERWENT AN ORTHOGNATHIC PROCEDURE INVOLVING THE UPPER AND LOWER JAW ON (B)(6) 2023. IT HAS BEEN DETERMINED BY THE SURGEON AND ORTHODONTIST THAT THE MAXILLARY SURGERY IS STABLE, HOWEVER, THE MANDIBULAR PORTION IS NOT AND MUST BE REVISED. THIS REPORT IS FOR A 1.85MM TI MATRIX SCREW SELF-TAPPING/6MM. THIS IS REPORT 1 OF 9 FOR (B)(4).

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 17450753 ↗
  • MalfunctionApr 29, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6), 2022 DURING ORTHOGNATHIC SURGERY, DURING THE FIXATION OF MANDIBLE FOLLOWING A BILATERAL SAGITTAL SPLIT OSTEOTOMY PROCEDURE WHILE INSERTING THE SCREW INTO THE PLATE, IT PROTRUDED THROUGH THE INDICATED PLATE HOLE, WHILE ATTEMPTING TO SECURE THE PLATE TO THE BONE. THIS HAPPENED TO THE PLATE ON THE LEFT AND THE RIGHT SIDE OF MANDIBLE DURING FIXATION. PROCEDURE COMPLETED SUCCESSFULLY. THIS REPORT IS FOR ONE 1.85MM TI MATRIX SCREW SELF-TAPPING/6MM. THIS IS REPORT 2 OF 3 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14251774 ↗
  • MalfunctionApr 26, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE USA AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE WHEN TRYING TO IMPLANT THE PATIENT, THE MATRIX SCREW HEAD THAT WE ARE USING FELL/SLIPPED THROUGH THE HOLE IN THE PLATE. WE USED AN ALTERNATIVE FIXATING THE MANDIBLE. NO SURGICAL DELAY. THE PROCEDURE OUTCOME IS UNKNOWN. NO PATIENT CONSEQUENCES. CONCOMITANT DEVICE REPORTED: UNK - PLATES: MATRIXORTHOGNATHIC (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: 1). THIS COMPLAINT INVOLVES ONE (1) DEVICE 1.85MM TI MATRIX SCREW SELF-TAPPING/6MM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14206829 ↗
  • MalfunctionAug 24, 2018Failure to Advance; Failure to Cut; Difficult to Advance; Failure to Advance; Failure to Cut; Difficult to Advance

    IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2018, THE TITANIUM (TI) MATRIX SCREW SELF-TAPPING WOULD NOT ENGAGE INTO THE PRE-DRILLED HOLE. A MATRIX DRILL BIT J-LATCH WAS USED TO PUT IN THE SCREW. IT IS UNKNOWN IF IT WAS AN ISSUE WITH THE SCREW OR THE DRILL BIT. ANOTHER UNKNOWN DRILL BIT AND UNKNOWN SCREW WERE USED, AND THEY WORKED FINE. IT IS UNKNOWN IF THERE WAS SURGICAL DELAY OR WHAT THE PATIENT OUTCOME WAS. THIS REPORT IS FOR A 1.85MM TI MATRIX SCREW SELF-TAPPING/6MM. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 7813511 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083388. One submission often covers many models, so check the product and lot details in each record.

  • Z-2272-2015 · Jun 24, 2015 · Terminated
    Synthes TI Matrix Pre-Bent Maxillary Plate / 2mm ADVMNT/0.8mm THK/L and Synthes TI Matrix Pre-Bent Maxillary Plate / 2mm ADVMNT/0.8mm THK/R. The Synthes Matrix ORTHOGNATHIC Fixati

Other PLATE, BONE devices

All PLATE, BONE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.