Matrix 04.511.224.01Recalls & reported problems
Synthes GmbH · 1.85MM TI MATRIX SCREW SELF-DRILLING/4MM · Class 2 · Implant
All models of Matrix (80)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10887587023595), received between Jul 19, 2022 and Jul 19, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).
1 other recall cover devices cleared under the same FDA submission (K083388); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 19, 2022Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM CANADA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE SURGEONS WERE IMPLANTING MATRIX ORTHOGNATHIC 1.85MM SCREWS INTO BONE OF POOR QUALITY. THE SCREWS COULD NOT BE PROPERLY TIGHTENED, AND KEPT SPINNING IN THE BONE WITHOUT GAINING PURCHASE. THE SAME SITUATION OCCURRED WITH 1.85MM SELF-DRILLING, SELF-TAPPING, AND 2.1MM EMERGENCY SCREWS. AFTER 10 MINUTES OF DELAY, THE SURGEON DECIDED TO USE ANOTHER SYSTEM TO COMPLETE THE PROCEDURE. THIS REPORT IS FOR A 1.85MM TI MATRIX SCREW SELF-DRILLING/4MM. THIS IS REPORT 5 OF 5 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15051601 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K083388. One submission often covers many models, so check the product and lot details in each record.
- Z-2272-2015 · Jun 24, 2015 · TerminatedSynthes TI Matrix Pre-Bent Maxillary Plate / 2mm ADVMNT/0.8mm THK/L and Synthes TI Matrix Pre-Bent Maxillary Plate / 2mm ADVMNT/0.8mm THK/R. The Synthes Matrix ORTHOGNATHIC Fixati
Other PLATE, BONE devices
- MatrixNEURO 04.503.104.01Synthes GmbH · PLATE, BONE252 reports
- RapidSorb 806.004.02SSynthes GmbH · PLATE, BONE146 reports
- TruMatch SD980.001Materialise NV · Plate, Bone1 recall21 reports
- Matrix 04.511.225.01Synthes GmbH · PLATE, BONE62 reports
- NA 400.834Synthes GmbH · PLATE, BONE58 reports
- TruMatch SD980.016Materialise NV · Plate, Bone55 reports
- ProPlan CMF SD900.101Materialise NV · Plate, Bone1 recall1 report
- TruMatch SD980.013Materialise NV · Plate, Bone1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page