Synthes GmbH · 2.0MM IMF SCREW SELF-DRILLING 8MM · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10887587028644), received between Dec 21, 2017 and Jan 19, 2024 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Entrapment of Device3
  • Material Fragmentation1

Patient problems

  • Foreign Body In Patient2
  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 19, 2024Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT DURING AN OPEN REDUCTION/INTERNAL FIXATION FOR A MANDIBLE FRACTURE ON (B)(6) 2024, AN INTERMEDULLARY FIXATION (IMF) SCREW BROKE OFF INTO A PATIENT. ONLY A PORTION OF THE SCREW WAS ABLE TO BE REMOVED. THE SCREW DID NOT BREAK OFF AT THE HEAD, IT BROKE OFF IN THE MIDDLE OF THE SCREW BODY AND HAD TO BE DRILLED OUT AND ONLY A PORTION WAS ABLE TO BE REMOVED. THERE WAS A SURGICAL DELAY OF 45 MINUTES, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE PATIENT WILL BE MONITORED FOR CONSEQUENCES RELATED TO THE EVENT. THIS REPORT INVOLVES ONE 2.0MM IMF SCREW SELF-DRILLING 8MM. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 18543366 ↗
  • MalfunctionOct 10, 2023Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT ON (B)(6) 2023, DURING SURGERY FOR A LEFT MANDIBLE ANGLE FRACTURE, THE 2.0MM IMF (INTERMAXILLARY FIXATION) SCREW HEAD BROKE OFF WHILE BEING INSERTED INTO PATIENT. THE HEAD WAS REMOVED, AND THE THREADS OF THE SCREW REMAINED IN THE PATIENT¿S LOWER JAW. THE IMF SCREWS WERE BEING USED TO PUT THE PATIENT INTO OCCLUSION. THE THREADS OF THE REMAINING SCREW WERE FILED DOWN AND SMOOTHED OVER, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THIS REPORT INVOLVES ONE 2.0MM IMF SCREW SELF-DRILLING 8MM. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 17906893 ↗
  • MalfunctionMay 15, 2023Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT ON APRIL 28, 2023, IMF SCREW BROKE DURING INSERTION. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. PATIENT OUTCOME IS STABLE. NO FRAGMENTS GENERATED. THIS REPORT IS FOR ONE (1) 2.0MM IMF SCREW SELF-DRILLING 8MM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16934703 ↗
  • MalfunctionSep 27, 2021Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2021, PATIENT UNDERWENT A SURGICAL PROCEDURE DUE TO MANDIBLE FRACTURE. WHILE THE SCREW WAS BEING INSERTED AND TIGHTENED THE IMF SCREW BROKE IN HALF. FRAGMENTS WERE GENERATED AND SCREW WAS BURRED OUT. SURGEON NEEDED TO BURR SCREW OUT SINCE THE TOP OF THE SCREW WAS OFF. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF TEN(10) MINUTES. NO PATIENT CONSEQUENCE. CONCOITANT DEVICE REPORTED: UNK - SCREWDRIVER (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: 1). THIS REPORT IS FOR ONE (1) 2.0MM IMF SCREW SELF-DRILLING 8MM. THIS IS REPORT 1 OF 1 FOR COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12537304 ↗
  • MalfunctionDec 20, 2018Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2018, DURING CLOSED REDUCTION IN MANDIBLE, THE INTERMAXILLARY FIXATION (IMF) SCREW BROKE UPON INSERTION. THE BROKEN FRAGMENT WAS LEFT IN THE MANDIBLE. ANOTHER SCREW WAS INSERTED NEXT TO IT. THERE WAS NO SURGICAL DELAY. PROCEDURE AND PATIENT OUTCOME ARE UNKNOWN. THIS REPORT IS FOR ONE (1) 2.0MM IMF SCREW SELF-DRILLING 8MM. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8185114 ↗
  • MalfunctionJan 17, 2018Break; Break

    IT WAS REPORTED THAT DURING A SURGERY FOR FACIAL TRAUMA ON (B)(6) 2017, WHEN THE SURGEON WAS TRYING TO INSERT A 2.0 MM INTERMAXILLARY FIXATION (IMF) SCREW SELF-DRILLING 8 MM, THE TIP OF IT BROKE OFF IN THE PATIENT'S MANDIBLE. AS THE TIP OF SCREW WAS NOT ABLE TO BE REMOVED IT WAS LEFT IN PATIENT'S MANDIBLE AND THE OTHER HALF OF THE SCREW WAS REMOVED. ANOTHER SCREW WAS AVAILABLE FOR COMPLETION OF THE PROCEDURE. THERE WAS NO SURGICAL DELAY AND THE PATIENT WAS STABLE AFTER THE SURGERY. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7201504 ↗
  • MalfunctionDec 21, 2017Break; Material Fragmentation; Break; Material Fragmentation

    IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2017 A 2.0 MM IMF SCREW SELF DRILLING 8 MM BROKE WHEN THE SURGEON WAS TRYING TO INSERT IT IN THE MANDIBLE OF THE PATIENT. THE BROKEN FRAGMENT WAS REMOVED AND THE SURGERY WAS COMPLETED WITHOUT ANY DELAY. THE PATIENT WAS STABLE AFTER THE SURGERY. CONCOMITANT DEVICE REPORTED: SCREWDRIVER (PART # UNKNOWN, LOT # UNKNOWN QTY # UNKNOWN). THIS REPORT IS FOR ONE (1) 2.0MM IMF SCREW SELF-DRILLING 8MM. (B)(4).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7139095 ↗

Other SCREW, FIXATION, INTRAOSSEOUS devices

All SCREW, FIXATION, INTRAOSSEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page