Skip to content

Bivona ZT25LN25NGE261NRecalls & reported problems

Icu Medical, Inc. · ZT25LN25NGE261N Bivona DI: Non-Adjustable-Wired_Non-Sterile · Class 2

All models of Bivona (469)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10887709182049), received between Mar 18, 2026 and Mar 18, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult to Insert” (1 report).

1 other recall cover devices cleared under the same FDA submission (K923878); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Insert1
  • Suction Failure1

Patient problems

  • Hyperextension1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2026Difficult to Insert; Suction Failure

    IT WAS REPORTED THAT THE DEVICE DID NOT ALLOW THE 8FR SUCTION CATHETER TO PASS SMOOTHLY WITHOUT SOME RESISTANCE. WHILE THE DEVICE WAS IN USE ON A PATIENT, THE TUBE HAD ALREADY BEEN INSERTED, AND THE ISSUE WAS NOT NOTICED UNTIL AFTER PLACEMENT WHEN SUCTION WAS INITIATED TO CLEAR SECRETIONS. THE SUCTION CATHETER COULD NOT BE INSERTED PROPERLY. THE BEDSIDE TEAM WAS ABLE TO REPLACE THE TUBE WITH THE PATIENT¿S ORIGINAL TRACH, BUT THIS RESULTED IN AN UNPLANNED TRACH CHANGE. MEDICATION INTERVENTION WAS REQUIRED BECAUSE THE PATIENT HAD TO BE SWITCHED BACK TO THE ORIGINAL TRACH, AS THE CUSTOM TUBES WERE NOT PATENT ENOUGH FOR 8 FR SUCTION CATHETERS. THE ISSUE WAS EVENTUALLY RESOLVED, BUT IT TOOK A THIRD SET OF TUBES TO OBTAIN ONES THAT WERE PATENT. THE PATIENT HAD TO REMAIN STRAPPED DOWN WITH THE NECK HYPEREXTENDED WHILE WAITING FOR THE CUSTOM TUBES, AND BECAUSE THE TUBES WERE NOT PATENT, THE PATIENT STAYED IN THAT POSITION UNTIL THE THIRD SET ARRIVED AND WAS CONFIRMED USABLE. THE PATIENT¿S DATE OF BIRTH WAS (B)(6) 2025; THE PATIENT WAS 4 MONTHS OLD DURING THE EVENT, WEIGHED 5.74 KG, AND WAS OF NON-HISPANIC/OTHER RACE. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: Hyperextension

    MAUDE report 24630692 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K923878. One submission often covers many models, so check the product and lot details in each record.

  • Z-0260-2007 · Nov 17, 2006 · Terminated
    Custom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by Smiths Medical Critical Care; REF HU05JS55NNA112S.

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page