Skip to content

TUR Set 654301Recalls & reported problems

ICU Medical, Inc. · LATEX-FREE T-U-R Y-Set Nonvented Set for Transurethral Resection Procedures 96 inch (244 cm) For Flexible Irrigation Container Systems (previously Hospira product) · Class 1

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10887787006725), received between Mar 12, 2018 and Feb 2, 2026 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation6
  • Particulates3
  • Break1
  • Contamination1
  • Fluid/Blood Leak1

Patient problems

  • No Consequences Or Impact To Patient10
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 2, 2026Fluid/Blood Leak

    UPDATED INFORMATION: THE PIERCING PIN HAS BEEN REPORTED LEAKING WHEN CONNECTED TO THE 1000ML AND 3000ML FLUID BAGS WHICH THEY ORDER FROM ICU MEDICAL. PREVIOUSLY, IT WAS COMMUNICATED THAT THE LEAKAGE ISSUE WAS OCCURRING AT THE BAG SPIKE END OF THE TUBING. HOWEVER, THE LEAKAGE IS ACTUALLY OCCURRING AT THE OPPOSITE END SPECIFICALLY AT THE CATHETER ADAPTER THAT CONNECTS TO THE FOLEY CATHETER. WHEN THE TUBING IS CONNECTED TO A SCOPE, THERE ARE NO ISSUES. HOWEVER, ONCE THE FLEXIBLE TUBING IS REMOVED AND THE TIP IS ATTACHED TO A FOLEY CATHETER, THE CONNECTION DOES NOT REMAIN SECURE AND EASILY SEPARATES. THIS CONNECTION FAILURE HAS RESULTED IN MULTIPLE PATIENTS BEING SOAKED WITH URINE, BOTH AT THEIR FACILITY AND IN THE HOME SETTING. THERE WERE PATIENT INVOLVEMENT AND NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24241430 ↗
  • MalfunctionJan 17, 2022Contamination; Contamination

    IT WAS REPORTED THAT AN EYELASH WAS IDENTIFIED IN THE BLADDER IRRIGATION TUBING OF A LTXFR NV TUR Y-SET. THE PACKAGE WAS SEALED PRIOR TO THE CUSTOMER FINDING THE EYELASH. THERE WAS NO PATIENT INVOLVEMENT, NO ADVERSE EVENT AND NO HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13265971 ↗
  • MalfunctionJun 20, 2019Particulates; Particulates

    ICU MEDICAL RECEIVED A USER FACILITY MEDWATCH (#MW5086789) STATING THAT WHEN CHANGING THE T-U-R TUBING, THE SPIKE TIP BROKE INTO PIECES DURING BLADDER IRRIGATION. THERE WAS NO HARM TO THE PATIENT AND NO MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8719253 ↗
  • MalfunctionJun 20, 2019Particulates; Particulates

    THE EVENT INVOLVED A T-U-R TUBING THAT WAS ORIGINALLY REPORTED THROUGH A VOLUNTARY MEDWATCH REPORT #MW5086789. IT WAS REPORTED THAT THE SPIKE BROKE INTO PIECES DURING BLADDER IRRIGATION. THERE WAS NO REPORT OF HUMAN HARM AND NO MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8718009 ↗
  • MalfunctionJun 17, 2019Particulates; Particulates

    ICU MEDICAL RECEIVED A USER FACILITY MEDWATCH (#MW5086789) STATING THAT WHEN CHANGING THE T-U-R TUBING, THE SPIKE TIP BROKE INTO PIECES DURING BLADDER IRRIGATION. THERE WAS NO HARM TO THE PATIENT AND NO MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8705391 ↗
  • MalfunctionMay 3, 2019Material Fragmentation; Material Fragmentation

    THE EVENT INVOLVED AN IRRIGATION SET, LARGE BORE, TUR, UROLOGICAL CONNECTOR 96 INCH, THAT WHEN PUNCTURING THE IV BAG, PIECES OF THE SPIKE WERE SEEN FLOATING IN THE SIGHT CHAMBER. THE DEVICE WAS REPLACED AND THERAPY WAS RESUMED. THERE WAS NO ADVERSE EVENT AND NO DELAY IN CRITICAL THERAPY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8577699 ↗
  • MalfunctionMay 3, 2019Material Fragmentation; Material Fragmentation

    THE EVENT INVOLVED AN IRRIGATION SET, LARGE BORE, TUR, UROLOGICAL CONNECTOR 96 INCH, THAT WHEN PUNCTURING THE IV BAG, PIECES OF THE SPIKE WERE SEEN FLOATING IN THE SIGHT CHAMBER. THE DEVICE WAS REPLACED AND THERAPY WAS RESUMED. THERE WAS NO ADVERSE EVENT AND NO DELAY IN CRITICAL THERAPY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8577698 ↗
  • MalfunctionApr 29, 2019Material Fragmentation; Material Fragmentation

    THE EVENT INVOLVED AN IRRIGATION SET, LARGE BORE, TUR, UROLOGICAL CONNECTOR 96 INCH, THAT WHEN PUNCTURING THE IV BAG, PIECES OF THE SPIKE WERE SEEN FLOATING IN THE SIGHT CHAMBER. THE DEVICE WAS REPLACED AND THERAPY WAS RESUMED. THERE WAS NO ADVERSE EVENT AND NO DELAY IN CRITICAL THERAPY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8561689 ↗
  • MalfunctionApr 29, 2019Material Fragmentation; Material Fragmentation

    THE EVENT INVOLVED AN IRRIGATION SET, LARGE BORE, TUR, UROLOGICAL CONNECTOR 96 INCH, THAT WHEN PUNCTURING THE IV BAG, B. BRAUN¿S TITAN XL ( 3L NORMAL SALINE FOR IRRIGATION), PIECES OF THE SPIKE WERE SEEN FLOATING IN THE SIGHT CHAMBER. THE DEVICE WAS REPLACED AND THERAPY WAS RESUMED. THERE WAS NO ADVERSE EVENT AND NO DELAY IN CRITICAL THERAPY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8561685 ↗
  • MalfunctionApr 29, 2019Material Fragmentation; Material Fragmentation

    THE EVENT INVOLVED AN IRRIGATION SET, LARGE BORE, TUR, UROLOGICAL CONNECTOR 96 INCH, THAT WHEN PUNCTURING THE IV BAG, B. BRAUN¿S TITAN XL ( 3L NORMAL SALINE FOR IRRIGATION), PIECES OF THE SPIKE WERE SEEN FLOATING IN THE SIGHT CHAMBER. THE DEVICE WAS REPLACED AND THERAPY WAS RESUMED. THERE WAS NO ADVERSE EVENT AND NO DELAY IN CRITICAL THERAPY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8561676 ↗
  • MalfunctionApr 3, 2019Material Fragmentation; Material Fragmentation

    THE CUSTOMER REPORTED THAT WHILE USING AN IRRIGATION SET, PIECES OF THE SPIKE BROKE OFF INTO IRRIGATION BAG. THIS OCCURRED DURING BLADDER IRRIGATION ON A PATIENT. THE WAS NO HARM REPORTED. IT WAS FURTHER DESCRIBED THAT THE CUSTOMER SPIKED A 3000L OF NORMAL SALINE AND THAT IS WHEN THE PIECE BROKE OFF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8479888 ↗
  • MalfunctionMar 12, 2018Break; Break

    DURING A CYSTOSCOPY-URETEROSCOPY WITH RETROGRADE PYELOGRAM AND LEFT STENT INSERTION/REMOVAL PROCEDURE ON A (B)(6) MALE, THE CUSTOMER ALLEGED THE TIP OF THE IV SPIKE BROKE OFF AND WENT INTO THE PATIENT¿S BLADDER. THE SURGEON RETRIEVED THE BROKEN PIECE OF THE TIP. THE CUSTOMER REPORTED THE PATIENT WAS FINE AND THAT THERE WAS NO INJURY OR OTHERWISE ADVERSE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7331692 ↗

Other CATHETER, IRRIGATION devices

All CATHETER, IRRIGATION devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.