PakPlus®Recalls & reported problems
Immucor Gti Diagnostics, Inc. · The PakPlus assay is a qualitative solid-phase enzyme-linked immunosorbent assay (ELISA) designed to screen for antibodies to HLA Class I and platelet glycoprotein (GP) IV antigens, and to polymorphic epitopes on the platelet glycoproteins IIb/IIIa, Ib/IX, and Ia/IIa. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10888234500254), received between Feb 8, 2018 and Feb 8, 2018 — classified as 1 malfunction.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 8, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER COMPLAINT ((B)(4)) WAS RECEIVED 09 JAN 2018. ON (B)(6) 2017, CUSTOMER TESTED FRESH PATIENT SAMPLE WITH LOT 3005747 OF PAKPLUS AND RECEIVED A NEGATIVE RESULT FOR WELL A. CUSTOMER RECEIVED NEW LOT OF PAKPLUS 3006225 ON (B)(6) 2018. ON (B)(6) 2018, CUSTOMER TESTING WITH LOT 3006225 OF PAKPLUS ON A PATIENT ALIQUOT FROM (B)(6) 2017. THE RESULT FOR WELL A WAS POSITIVE. CUSTOMER ALSO TESTED ANOTHER PATIENT ALIQUOT COLLECTED FROM (B)(6) WITH LOT 3005747 AND WELL A WAS NEGATIVE. TESTING WAS REPEATED ON (B)(6) 2018 USING BOTH LOTS. LOT 3005747 RESULTED IN WELL A NEGATIVE AND LOT 3006225 RESULTED IN WELL A POSITIVE. CUSTOMER REPORTED NEGATIVE RESULTS ON THE PATIENT FROM THE INITIAL TESTING.
Patient problems: No Information
MAUDE report 7257347 ↗
Other Test,Platelet Antibody devices
- PakAuto®Immucor Gti Diagnostics, Inc. · Test,Platelet Antibody1 report
- ThromboType®Immucor Gti Diagnostics, Inc. · Test,Platelet Antibody1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page