Skip to content

Medline ENFIT1010GCRecalls & reported problems

Medline Industries, Inc. · ENFIT G-TUBE CONNECTOR, MEDLINE · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10888277314719), received between Apr 21, 2026 and Sep 23, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 4 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 13, 2026Class IIOpen, Classified

    Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.

    Z-1690-2026 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Device Dislodged or Dislocated2
  • Failure to Disconnect2
  • Break1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026Failure to Disconnect; Detachment of Device or Device Component

    It was reported that "part of the connector becomes detached and the harder plastic part is almost impossible to detach from the patient's tube."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50870293 ↗
  • MalfunctionSep 21, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50841713 ↗
  • MalfunctionSep 10, 2026Failure to Disconnect; Detachment of Device or Device Component

    It was reported that "We have had three separate occurrences with three different patients and three different nurses where part of the connector becomes detached and the harder plastic part is almost impossible to detach from the patient's tube."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50713460 ↗
  • MalfunctionSep 2, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50571579 ↗
  • MalfunctionApr 21, 2026Device Dislodged or Dislocated

    CUSTOMER REPORTED, "ENFIT G-TUBE CONNECTOR WITH HARD PLASTIC INSERT PIECE WITHIN SILICONE CONNECTOR DISLODGE AND RETAINED IN GASTROTOMY OR JEJUNOSTOMY TUBES FOR (2) PATIENTS. IN BOTH CASES, THESE PIECES NEEDED TO BE REMOVED WITH KELLY CLAMPS."

    Patient problems: Insufficient Information

    MAUDE report 24945470 ↗
  • MalfunctionApr 21, 2026Device Dislodged or Dislocated

    CUSTOMER REPORTED, "ENFIT G-TUBE CONNECTOR WITH HARD PLASTIC INSERT PIECE WITHIN SILICONE CONNECTOR DISLODGE AND RETAINED IN GASTROTOMY OR JEJUNOSTOMY TUBES FOR (2) PATIENTS. IN BOTH CASES, THESE PIECES NEEDED TO BE REMOVED WITH KELLY CLAMPS."

    Patient problems: Insufficient Information

    MAUDE report 24945231 ↗

Other Enteral specific transition connectors devices

All Enteral specific transition connectors devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page