Skip to content

Enfit G-Tube Connector, Medline recallZ-1690-2026 · Medline Industries, LP · initiated Feb 13, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.

FDA root cause category: Process change control

What the company is doing

On February 13, 2026, the firm notified affected consignees via Urgent Medical Device Recall letters delivered through First Class mail and email. Customers were instructed to destroy all affected ENFit G-Tube connectors and kits that contain the G-Tube connectors.

Recall details

Recall number
Z-1690-2026
Status
Open, Classified
Initiated by the company
Feb 13, 2026
Posted by the FDA
Mar 27, 2026
Recalling firm
Medline Industries, LP
Quantity
651,789
Product code
PIO
FDA submissions
K151628
Distribution
Worldwide - US Nationwide and the countries of Canada.

Product description

ENFIT G-TUBE CONNECTOR, MEDLINE Item No. ENFIT1010GC

Code information (models, lots, serial numbers)

UDI-DI 10888277314719 (ea) 40888277314710(case) All Lots

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98524).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page