Skip to content

MEDLINE 401606RHRecalls & reported problems

Medline Industries, Inc. · 401606 @DIAGNOSTIC EP CAT, STEERABLE, LR · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K221067); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jul 31, 2025Open, Classified

    Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

    Z-2610-2025 · Root cause: Under Investigation by firm · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K221067. One submission often covers many models, so check the product and lot details in each record.

  • Z-1321-2026 · Dec 22, 2025 · Open, Classified
    Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheters: Livewire Steerable Electrophysiology Catheter,10 Elec. 2-5-2mm 401582RH 401940RH; Livewire Steerable Electr
  • Z-1460-2025 · Feb 6, 2025 · Open, Classified
    Medline ReNewal Reprocessed Livewire Steerable Electrophysiology Catheters, Medline Item Numbers: 401582R, 401904R, 401904R, 401905R, 401933R, and 401991R

Other Catheter, recording, electrode, reprocessed devices

All Catheter, recording, electrode, reprocessed devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.