Reprocessed Livewire Steerable Electrophysiology Catheter recallZ-2610-2025 · Medline Industries, LP · initiated Jul 31, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
FDA root cause category: Under Investigation by firm
What the company is doing
On July 31, 2025, the firm notified customers via first class mail and email of the potential product issue. An updated letter was sent to customers on 8/13/25 with additional item and lot numbers. Affected products are to be returned for credit. Customers were instructed to immediately identify and quarantine all affected product within their possession. Once they have completed and returned the response form listing their affected inventory, they will receive return labels and RGA information. Customers will receive credit when the returned product is received.
Recall details
- Recall number
- Z-2610-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 31, 2025
- Posted by the FDA
- Sep 26, 2025
- Recalling firm
- Medline Industries, LP
- Quantity
- 7913 eaches
- Product code
- NLH
- FDA submissions
- K221067
- Distribution
- Domestic US distribution nationwide. International distribution pending. No international distribution.
Product description
Reprocessed Livewire Steerable Electrophysiology Catheter Item Numbers 401575RH 401582RH 401600RH 401603RH 401606RH 401652RH 401653RH 401654RH 401904RH 401905RH 401908RH 401914RH 401915RH 401918RH 401932RH 401933RH 401934RH 401938RH 401939RH 401940RH 401941RH 401991RH
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Item Number/UDI (ea)/UDI(case): 401575RH 10888277407497 20888277407494; 401582RH 10888277407510 20888277407517; 401600RH 10888277407541 20888277407548; 401603RH 10888277407558 20888277407555; 401606RH 10888277407565 20888277407562; 401652RH 10888277407572 20888277407579; 401653RH 10888277407589 20888277407586; 401654RH 10888277407596 20888277407593; 401904RH 10888277407602 20888277407609; 401905RH 10888277407619 20888277407616; 401908RH 10197344020966 20197344020963; 401914RH 10888277407527 20888177407524; 401915RH 10888277407626 20888277407623; 401918RH 10888277407640 20888277407647; 401932RH 10888277407671 20888277407678; 401933RH 10888277407688 20888277407685; 401934RH 10888277407695 20888277407692; 401938RH 10888277407701 20888277407708; 401939RH 10888277407718 20888277407715; 401940RH 10888277407480 20888277407487; 401941RH 10888277407725 20888277407722; 401991RH 10888277407756 2088277407753; LOTS EP230724 EP230731 EP230803 EP230816 EP230914 EP231017 EP231116 EP240108 EP240122 EP240212 EP240227 EP240311 EP240322 EP240411 EP240415 EP240422 EP240429 EP240507 EP240514 EP240520 EP240528 EP240607 EP240611 EP240710 EP240715 EP240723 EP240805 EP240812 EP240820 EP240829 EP240911 EP240923 EP240930 EP241008 EP241030 EP241125 EP250108
Devices named in this recall
22 devices in the FDA device registry whose identifier appears in the recall's code information.
- MEDLINE 401904RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed3 recalls1 report
- MEDLINE 401582RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed3 recalls0 reports
- MEDLINE 401940RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401914RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401652RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401905RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401932RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401933RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401991RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401575RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
- MEDLINE 401600RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
- MEDLINE 401603RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
- MEDLINE 401606RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
- MEDLINE 401653RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
- MEDLINE 401654RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
- MEDLINE 401915RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
- MEDLINE 401918RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
- MEDLINE 401934RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
- MEDLINE 401938RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
- MEDLINE 401939RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
- MEDLINE 401941RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
- MEDLINE 401908RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
Other devices under the same FDA submission
10 devices cleared under K221067 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- MEDLINE 401860RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401877RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401878RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401890RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401891RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 402004RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401863RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
- MEDLINE 401864RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
- MEDLINE 401966RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
- MEDLINE 401994RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97496).
- Z-2611-2025 · Open, ClassifiedReprocessed Webster CS Catheter 10 Electrodes for electrophysiological mapping of cardiac structures. Medline Item Numbers BD710DF282CRH BD710DF282RRH BD710FJ282CRH BD710FJ282RRH
- Z-2612-2025 · Open, ClassifiedReprocessed Inquiry Steerable Diagnostics Catheter for electrogram recording and cardiac stimulation during diagnostic electrophysiologic studies; Medline Item Numbers 81102RH 8110…
- Z-2613-2025 · Open, ClassifiedReprocessed ViewFlex Xtra ICE Catheter to visualize cardiac structures; blood flow and other devices within the heart. Medline Item Numbers D087031RH
- Z-2614-2025 · Open, ClassifiedReprocessed ACUSON AcuNav Ultrasound Catheter for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page