Skip to content

Medline ENTR99235Recalls & reported problems

Medline Industries, Inc. · DBD-PUMP,FEEDING,ENTERAL,ENTRAFLO RNTL · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10888277420014), received between Jul 10, 2024 and Nov 20, 2024 — classified as 1 death, 1 injury.

The most frequently reported specific device problem is “Excess Flow or Over-Infusion” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1
  • Excess Flow or Over-Infusion1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Aspiration/Inhalation1
  • Respiratory Failure1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 20, 2024Excess Flow or Over-Infusion; Excess Flow or Over-Infusion

    ACCORDING TO THE CUSTOMER, THE "RETAINER BRACKET WAS NOT SECURED IN THE APPROPRIATE BLUE HOLDER, TRIGGERING A FREE FLOW OF THE ISOSOURCE FEED. FREE FLOW ALARM DID NOT ENGAGE AND RESIDENT RECEIVED AN OVERFEED, WITHIN HER 1 HOUR FEED TIME."

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Aspiration/Inhalation

    MAUDE report 20735818 ↗
  • DeathJul 10, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    ACCORDING TO THE FACILITY ON (B)(6) 2024, "THE PATIENT WAS SUPPOSED TO GET 400ML IN 2 HRS FROM THE ENTRAFLO FEEDING PUMP. THE MACHINE WAS BEEPING AND THE NURSE RESPONDED TO THE ALARM. THE RESIDENT WAS CHECKED AT 7:50AM AND IT WAS NOTICED 1200CC WAS DELIVERED. THE BAG THAT WAS HUNG WAS 1.5L, ONLY 300ML LEFT WHEN CHECKED BY NURSE. THE RESIDENT RECEIVED 1200ML IN 90 MIN.

    Patient problems: Respiratory Failure

    MAUDE report 19709820 ↗

Other Pump, infusion, enteral devices

All Pump, infusion, enteral devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page