Skip to content

NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FF-410375Recalls & reported problems

Allergan, Inc. · NATRELLE 410 STYLE FF 375CC BIOCELL |textured|contour|silicone · Class 3 · Implant

All models of Natrelle Breast Implants (1167)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Natrelle 410 textured silicone breast implant is a soft, shaped breast implant made by Allergan. The 36 reports mostly describe implants that rupture or leak, with the surrounding scar tissue tightening around the implant and fluid collecting near it, sometimes requiring removal and replacement. All 36 reports describe injuries, and no deaths were reported. One open recall, started in July 2019, followed FDA notice of a higher rate of a rare breast implant-related lymphoma in patients with textured implants.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
36
Deaths reported
0
Injuries reported
36
Recalls naming it
1

Summary

The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 10888628002678), received between May 14, 2019 and Mar 26, 2026 — classified as 36 injuries.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (18 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 24, 2019Class IOpen, Classified

    The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.

    Z-2452-2019 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Appears to Trigger Rejection18
  • Material Rupture10
  • Patient-Device Incompatibility7
  • Adverse Event Without Identified Device or Use Problem2
  • No Apparent Adverse Event1

Patient problems

  • Capsular Contracture24
  • Failure of Implant7
  • Seroma7
  • No Clinical Signs, Symptoms or Conditions2
  • Abscess1
  • Cellulitis1
  • Erosion1
  • Swollen Lymph Nodes1

Recent adverse event reports

The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 26, 2026Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED "BILATERAL RUPTURES." THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24699647 ↗
  • InjuryMar 26, 2026Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED "INTRACAPSULAR RUPTURE". THIS RECORD IS FOR THE LEFT SIDE. DEVICE REMAINS IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698864 ↗
  • InjuryFeb 17, 2026Device Appears to Trigger Rejection; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED "RUPTURE", "CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN" AND "FLUID HAS ALSO BEEN IDENTIFIED ON ONE SIDE VIA IMAGING". THIS RECORD IS FOR LEFT SIDE. THE DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 24363867 ↗
  • InjuryOct 30, 2025No Apparent Adverse Event; No Apparent Adverse Event

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "REMOVAL OF TEXTURED DEVICE DUE TO THE PATIENTS CONCERNS WITH THE PRODUCT" AND "LATE ONSET SEROMA". THE DEVICE WAS EXPLANTED.

    Patient problems: Seroma

    MAUDE report 23431282 ↗
  • InjurySep 10, 2025Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED ¿RUPTURE¿. THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 23016003 ↗
  • InjuryAug 7, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED "SEROMA" AND LATER REPORTED "COLLAPSED CAPSULAR TISSUE MEASURING 0 CM IN RECONSTRUCTED DIAMETER, FIBROPLASIA, SCLEROSIS, FIBROSIS AND CALCIFICATION." AFFECTED SIDE IS RIGHT. DEVICE WAS EXPLANTED.

    Patient problems: Capsular Contracture; Seroma

    MAUDE report 22738499 ↗
  • InjuryJul 31, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE EXCHANGE FROM TEXTURED TO SMOOTH IMPLANTS DUE TO THE PATIENTS CONCERN WITH THE PRODUCT. HEALTHCARE PROFESSIONAL LATER REPORTED CAPSULAR CONTRACTURE, BAKER GRADE UNKNOWN. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 17429080 ↗
  • InjuryJul 31, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE EXCHANGE FROM TEXTURED TO SMOOTH IMPLANTS DUE TO THE PATIENTS CONCERN WITH THE PRODUCT. HEALTHCARE PROFESSIONAL LATER REPORTED CAPSULAR CONTRACTURE, BAKER GRADE UNKNOWN. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 17428186 ↗
  • InjuryMay 29, 2023Material Rupture; Material Rupture

    THE DEVICE HAS BEEN EXPLANTED AND DISCARDED.

    Patient problems: Failure of Implant

    MAUDE report 17017353 ↗
  • InjuryMay 23, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE NO COMPLAINT AGAINST DEVICE. HEALTHCARE PROFESSIONAL LATER REPORTED CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 16989035 ↗
  • InjuryMar 9, 2023Device Appears to Trigger Rejection; Material Rupture; Device Appears to Trigger Rejection; Material Rupture

    HEALTHCARE PROFESSIONAL LATER REPORTED CAPSULE APPEARED TO BE GRADE II-III AND NO RUPTURE WAS FOUND. AS NOTED IN THE OPERATIVE REPORT, PERIPROSTHETIC FLUID ON THE RIGHT SIDE WAS ASPIRATED AND SENT FOR CYTOLOGY. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture; Failure of Implant; Seroma

    MAUDE report 16516708 ↗
  • InjuryDec 16, 2022Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE EXCHANGE FROM TEXTURED TO SMOOTH BREAST IMPLANTS DUE TO THE PATIENT¿S CONCERN WITH THE PRODUCT. HEALTHCARE PROFESSIONAL LATER REPORTED RUPTURE THAT WAS NOTICED DURING EXPLANT SURGERY. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 16000524 ↗
  • InjuryJul 19, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    PATIENT REPRESENTATIVE REPORTED "CAPSULAR CONTRACTURE" BAKER GRADE UNKNOWN AND "INCREASED RISK OF ALCL." DEVICE REMAINS IMPLANTED. THIS RECORD REPRESENTS THE LEFT SIDE.

    Patient problems: Capsular Contracture

    MAUDE report 15051896 ↗
  • InjuryJul 1, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTH PROFESSIONAL REPORTED LEFT SIDE EXCHANGE FROM TEXTURED TO SMOOTH IMPLANT DUE TO THE PATIENT¿S CONCERN WITH THE PRODUCT. LATER PATIENT REPRESENTATIVE REPORTED, "CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN REQUIRING BILATERAL CAPSULECTOMY, CELLULITIS AND ABSCESS OF THE BREAST, AND INCREASED RISK OF ALCL." THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Abscess; Capsular Contracture; Cellulitis

    MAUDE report 14885361 ↗
  • InjuryOct 29, 2021Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "EXCHANGE OF TEXTURED TO SMOOTH BREAST IMPLANTS DUE TO THE PATIENT¿S CONCERN WITH THE PRODUCT AND CAPSULAR CONTRACTIVE, BAKER IV." DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 12722212 ↗
  • InjuryOct 26, 2021Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE IV AND "EXCHANGE OF TEXTURED TO SMOOTH BREAST IMPLANTS DUE TO THE PATIENT¿S CONCERN WITH THE PRODUCT." ADDITIONAL EVENT WAS REPORTED "EVIDENCE OF OLD BLOOD HEMATOMA". DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 12696273 ↗
  • InjuryOct 1, 2021Material Rupture; Material Rupture

    PATIENT REPORTED LEFT SIDE REMOVAL DUE TO THE PATIENT¿S CONCERN WITH THE PRODUCT. DEVICE HAS BEEN EXPLANTED. HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE RUPTURE.

    Patient problems: Failure of Implant

    MAUDE report 12565715 ↗
  • InjuryApr 7, 2021Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "PATIENT'S CONCERN WITH THE PRODUCT" AND CAPSULAR CONTRACTURE, BAKER GRADE UNKNOWN. ADDITIONALLY, HEALTHCARE PROFESSIONAL REPORTED AND CONFIRMED BAKER GRADE III. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 11628013 ↗
  • InjuryApr 7, 2021Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE EXCHANGE FROM TEXTURED TO SMOOTH IMPLANTS DUE TO THE PATIENT'S CONCERN WITH THE PRODUCT AND CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 11627500 ↗
  • InjuryApr 2, 2021Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE "VERY CONCERNED WITH A TEXTURED IMPLANT", "ROTATED 45 DEGREE IN A CLOCKWISE POSITION", "DEFORMITY OF RECONSTRUCTED BREAST", "ANIMATION DEFORMITY ON THE LEFT LATERAL ASPECT OF BREAST", "LEFT IMPLANT HAS BEEN EXTRUDING", AND "A SMALL DEGREE OF SEROMA WHICH IS CULTURED AND WE CAN ALSO OBSERVE INSIDE THE STRETCHING OF THE LATERAL POCKET A LITTLE BIT." DEVICE HAS BEEN EXPLANTED.

    Patient problems: Erosion; Seroma

    MAUDE report 11609713 ↗
  • InjuryFeb 26, 2021Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL ADDITIONALLY REPORTED A "PSEUDOCYST." DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 11387079 ↗
  • InjuryFeb 10, 2021Material Rupture; Material Rupture

    PATIENT REPORTED RIGHT SIDE EXCHANGE DUE TO CONCERN WITH THE PRODUCT. HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "TORN AND RUPTURED." DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 11308200 ↗
  • InjuryFeb 8, 2021Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE "PAIN", CAPSULAR CONTRACTURE, BAKER GRADE IV, AND "PATIENT'S CONCERNS WITH THE PRODUCT." DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 11294524 ↗
  • InjuryFeb 8, 2021Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "PAIN", CAPSULAR CONTRACTURE, BAKER GRADE IV, AND "PATIENT'S CONCERNS WITH THE PRODUCT." DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 11294410 ↗
  • InjuryJan 6, 2021Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 11124646 ↗

Other Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled devices

All Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page