Natrelle Breast Implants recallZ-2452-2019 · Allergan PLC · initiated Jul 24, 2019
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Allergan PLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
FDA root cause category: Unknown/Undetermined by firm
What the company is doing
UPDATE: On June 1, 2020, Allergan issued a press release outlining a dedicated multi-channel campaign to contact women who may not be aware of the Biocell recall and to improve tracking information for U. S. Breast Implant patients (link: https://news.abbvie.com/news/press-releases/allergan-aesthetics-launches-dedicated-multi-channel-campaign-to-contact-patients-who-may-not-be-aware-biocell-recall-and-to-improve-tracking-information-for-us-breast-implant-patients.htm) On July 24, 2019, Allergan issued a press release related to the recall was posted on the Allergan website (Allergan.com). Recall letters were sent on August 6, 2019 via: FedEx overnight mail with tracking to all consignees who may have product within expiry. The recall letter will inform customers to do the following: 1. If you have inventory of the recalled products, Quarantine product to prevent its use. 2. Conduct a physical count of the affected products in your possession and record the count on the enclosed Recall Stock Response Form. 3. Send the Recall Stock Response Form to Inmar Rx Solutions, Inc via fax or email within five (5) business days of receipt. To ensure we are able to account for all recalled product, it is imperative that you return the form. 4. Upon receipt of the Response Form, Inmar will issue Return Authorization label(s). 5. When returning the recalled products, attach the prepaid FedEx Authorized Return shipping label to the outside of the return carton and return the recalled product to: Inmar Rx Solutions, Inc. 4332 Empire Rd. Fort Worth, TX 76155 6. Please Do Not return any products that are not the subject of this recall. Please contact Inmar Rx Solutions, Inc. if you have any questions about these recall actions. Patient letters were mailed on August 14, 19 and 26, 2019 to all impacted patients to which Allergan had complete names and addresses in their device tracking database. The patient letters informed customers of the following: 1. How
Recall details
- Recall number
- Z-2452-2019
- Status
- Open, Classified
- Initiated by the company
- Jul 24, 2019
- Posted by the FDA
- Sep 11, 2019
- Recalling firm
- Allergan PLC
- Quantity
- 4,026,287 Breast Implants and Tissue Expanders Combined in total
- Product code
- FTR
- FDA submissions
- P040046
- Distribution
- Worldwide Distribution and US Nationwide All fifty (50) US States, the US Virgin Islands and Puerto Rico. (AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY). OUS: Bermuda, Canada, Chile, Japan, Korea, Thailand, Taiwan and Vietnam.
Product description
Natrelle 410, Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implant Biocell Low Height Low Projection, Sterile, for the following style numbers: Syle LL: LL-410135,LL-410180,LL-410210,LL-410240,LL-410300; Style LM: LM-410140, LM-410190, LM-410220, LM-410250, LM-410320; Style LF: LF-410125, LF-410150,LF-410175,LF-410205,LF-410240,LF-410270,LF-410310,LF-410390,LF-410440,LF-410490,LF-410540,LF-410595,LL-410135,LL-410180,LL-410210,LL-410240,LL-410300; style LX: LX-410145,LX-410175,LX-410195,LX-410225,LX-410255,LX-410290,LX-410330,LX-410365,LX-410405,LX-410455,LX-410515,LX-410570,LX-410625; Style ML: ML-410125,ML-410170,ML-410195,ML-410220,ML-410285; Style MM: MM-410160,MM-410185,MM-410215,MM-410245,MM-410280,MM-410320,MM-410360,MM-410400,MM-410450; Style MF: MF-410140MF-410165,MF-410195,MF-410225,MF-410255,MF-410295,MF-410335,MF-410375,MF-410420,MF-410470,MF-410525,MF-410580,MF-410640; Style MX: MX-410165,MX-410195,MX-410225,MX-410255,MX-410290,MX-410325,MX-410370,MX-410410,MX-410445,MX-410520,MX-410550,MX-410620,MX-410685; Style FL: FL-410140,FL-410190,FL-410220,FL-410250,FL-410320; Style FM:FM-410155,FM-410180,FM-410205,FM-410235,FM-410270,FM-410310,FM-410350,FM-410395,FM-410440,FM-410500,FM-410550,FM-410605,FM-410670; Style FF: FF-410160,FF-410185,FF-410220,FF-410255,FX-410290,FF-410335,FF-410375,FF-410425,FF-410475,FF-410535,FF-410595,FF-410655,FF-410740; Style FX: FX-410185FX-410215,FX-410245,FX-410280,FX-410315,FX-410360,FX-410410,FX-410450,FX-410495,FX-410560,FX-410615,FX-410690,FX-410775 These products may have been sold under the McGhan, Inamed or NATRELLE brands. Product Usage: Breast augmentation and Breast reconstruction
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
ALL LOTS Product Code/UDI Style LL: LL-410135/10888628003170,LL-410180/10888628003187,LL-410210/10888628003194,LL-410240/10888628003200,LL-410300/10888628003217; Style LM: LM-410140/10888628003224,LM-410190/10888628003231,LM-410220/10888628003248,LM-410250/10888628003255,LM-410320/10888628003262; Style LF: LF-410125/10888628003057,LF-410150/10888628003064,LF-410175/10888628003071,LF-410205/10888628003088,LF-410240/10888628003095,LF-410270/10888628003101,LF-410310/10888628003118,LF-410390/10888628003125,LF-410440/10888628003132,LF-410490/10888628003149,LF-410540/10888628003156,LF-410595/10888628003163; Style LX: LX-410145/10888628003279,LX-410175/10888628003286,LX-410195/10888628003293,LX-410225/10888628003309,LX-410255/10888628003316,LX-410290/10888628003323,LX-410330/10888628003330,LX-410365/10888628003347,LX-410405/10888628003354,LX-410455/10888628003361,LX-410515/10888628003378,LX-410570/10888628003385,LX-410625/10888628003392; Style ML: ML-410125/10888628003538,ML-410170/10888628003545,ML-410195/10888628003552,ML-410220/10888628003569,ML-410285/10888628003576; Style MM: MM-410160 10888628003583,MM-410185/10888628003590,MM-410215/10888628003606,MM-410245/10888628003613,MM-410280/10888628003620,MM-410320/10888628003637,MM-410360/10888628003644,MM-410400/10888628003651,MM-410450/10888628003668; Style MF:MF-410140/10888628003408,MF-410165/10888628003415,MF-410195/10888628003422,MF-410225/10888628003439,MF-410255/10888628003446,MF-410295/10888628003453,MF-410335/10888628003460,MF-410375/10888628003477,MF-410420/10888628003484,MF-410470/10888628003491,MF-410525/10888628003507,MF-410580/10888628003514,MF-410640/10888628003521; Style MX:MX-410165/10888628003675,MX-410195/10888628003682,MX-410225/10888628003699,MX-410255/10888628003705,MX-410290/10888628003712,MX-410325/10888628003729,MX-410370/10888628003736,MX-410410/10888628003743,MX-410445/10888628003750,MX-410520/10888628003767,MX-410550/10888628003774,MX-410620/10888628003781,MX-410685/10888628003798; Style FL: FL-410140/10888628002746,FL-410190/10888628002753,FL-410220/10888628002760,FL-410250/10888628002777,FL-410320/10888628002784; Style FM:FM-410155/10888628002791,FM-410180/10888628002807,FM-410205/10888628002814,FM-410235/10888628002821,FM-410270/10888628002838,FM-410310/10888628002845,FM-410350/10888628002852,FM-410395/10888628002869,FM-410440/10888628002876,FM-410500/10888628002883,FM-410550/10888628002890,FM-410605/10888628002906,FM-410670/10888628002913; Style FF: FF-410160/10888628002616,FF-410185/10888628002623,FF-410220/10888628002630,FF-410255/10888628002647,FF-410290/10888628002654,FF-410335/10888628002661,FF-410375/10888628002678,FF-410425/10888628002685,FF-410475/10888628002692,FF-410535/10888628002708,FF-410595/10888628002715,FF-410655/10888628002722,FF-410740/10888628002739; Style FX: FX-410185/10888628002920,FX-410215/10888628002937,FX-410245/10888628002944,FX-410280/10888628002951,FX-410315/10888628002968,FX-410360/10888628002975,FX-410410/10888628002982,FX-410450/10888628002999,FX-410495/10888628003002,FX-410560/10888628003019,FX-410615/10888628003026,FX-410690/10888628003033,FX-410775/10888628003040;
Devices named in this recall
118 devices in the FDA device registry whose identifier appears in the recall's code information; the 100 most reported are shown.
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FX-410495Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall62 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FX-410775Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall58 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FX-410560Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall1 death49 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FX-410450Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall48 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FF-410655Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall46 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MX-410445Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall46 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FF-410425Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall43 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FX-410615Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall43 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FX-410360Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall39 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FX-410690Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall39 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FX-410410Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall37 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FF-410375Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall36 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MF-410375Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall35 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MF-410640Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall34 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FF-410535Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall33 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MX-410685Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall32 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FF-410475Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall1 death31 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FF-410740Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall31 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MF-410470Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall31 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MM-410320Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall31 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MX-410410Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall30 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MX-410520Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall30 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MF-410335Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall29 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MF-410525Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall1 death29 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FF-410595Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall28 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MF-410580Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall26 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MX-410620Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall26 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MX-410325Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall25 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MF-410420Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall24 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MX-410550Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall24 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MX-410370Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall23 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FM-410310Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall22 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MF-410255Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall19 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MM-410360Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall19 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FM-410350Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall14 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FX-410315Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall14 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MF-410295Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall14 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FF-410335Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall12 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FF-410290Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall11 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MM-410400Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall11 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MM-410450Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall11 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FM-410440Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall10 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FL-410320Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall9 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FX-410280Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall9 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FM-410270Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall8 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MM-410280Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall8 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MX-410290Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall8 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FL-410250Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall7 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FM-410500Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall7 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MM-410245Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall6 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LX-410365Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall5 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MF-410225Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall5 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MM-410185Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall5 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FM-410395Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall4 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LF-410490Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall4 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LX-410405Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall4 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants ML-410285Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall4 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MX-410255Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall4 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FM-410550Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall3 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LF-410390Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall3 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LF-410595Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall3 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LL-410135Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall3 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LX-410255Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall3 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LX-410570Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall3 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LX-410625Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall3 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MF-410195Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall3 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FL-410140Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall2 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FL-410190Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall2 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FX-410245Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall2 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LF-410240Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall2 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LL-410180Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall2 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LL-410300Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall2 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LM-410250Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall2 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LX-410330Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall2 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LX-410515Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall2 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants MM-410215Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall2 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FF-410220Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall1 report
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FF-410255Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall1 report
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FM-410205Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall1 report
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FM-410235Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall1 report
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FM-410605Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall1 report
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LF-410175Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall1 report
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LF-410440Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall1 report
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LF-410540Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall1 report
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LM-410220Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall1 report
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LM-410320Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall1 report
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LX-410290Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall1 report
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants ML-410170Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall1 report
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants ML-410195Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall1 report
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FF-410160Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall0 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FF-410185Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall0 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FL-410220Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall0 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FM-410155Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall0 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FM-410180Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall0 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FM-410670Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall0 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FX-410185Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall0 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants FX-410215Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall0 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LF-410125Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall0 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LF-410150Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall0 reports
- NATRELLE 410 Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implants LF-410205Allergan, Inc. · Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 83500).
- Z-2451-2019 · Open, ClassifiedNatrelle, Saline-Filled Breast Implant Biocell Full Height with Fill Tube, Sterile, for the following style numbers: Style 163:163-360CC, 163-440CC, 163-530CC, 163-655CC, 163-780…
- Z-2453-2019 · Open, ClassifiedNatrelle Inspira, Silicone-Filled Breast Implants, Sterile, for the following style numbers: Style TRL: TRL-110TRL-125,TRL-140,TRL-170,TRL-200,TRL-230,TRL-260,TRL-290,TRL-320,TRL-…
- Z-2454-2019 · Open, ClassifiedNatrelle and McGhan Silicone-Filled Breast Implant BioCell Round Moderate Profile, Sterile, for the following style numbers: Style 110:110-090, 110-120, 110-150, 110-180, 110-210, …
- Z-2455-2019 · Open, ClassifiedNatrelle 133 Tissue Expanders (also named style 133 BIOSPAN Tissue Expander and Mcghan Magna-Site Tissue Expander) BioCell Full Height Variable Projection with Magna-Fider Xact & 2…
- Z-2456-2019 · Open, ClassifiedNatrelle133 Plus Tissue Expander With/Without Suture Tabs, BioCell Full Height Variable Projection with Magna-Fider Xact & 21G Needle Infusion Set, Sterile, for the following style…
- Z-2457-2019 · Open, ClassifiedNatrelle 133 Tissue Expanders w/Suture Tabs BIOCELL Full Height Variable Projection with MAGNA-FINDEr Xact & 21G Needle Infusion Set STYLE 133FV-T: 133FV-11-T, 133FV-12-T, 133FV-1…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page