Natrelle® Style 133 Tissue Expanders with Suture Tabs 133MX-16-TRecalls & reported problems
Allergan, Inc. · NATRELLE 133MX With Sut Tab, 850cc. · Class U · Implant
All models of Natrelle Breast Implants (1167) · Natrelle Style 133 Tissue Expanders with Suture Tabs (42)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10888628004511), received between Nov 5, 2018 and Sep 29, 2020 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jul 24, 2019Class IOpen, Classified
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
Z-2457-2019 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak3
- Improper or Incorrect Procedure or Method1
Patient problems
- Failure of Implant2
- Cellulitis1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjurySep 29, 2020Fluid/Blood Leak; Fluid/Blood Leak
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "RUPTURE". THE EXPANDER WAS IMPLANTED FOR MORE THAN SIX MONTHS. DEVICE HAS BEEN EXPLANTED.
Patient problems: Failure of Implant
MAUDE report 10601223 ↗MalfunctionAug 30, 2019Fluid/Blood Leak; Fluid/Blood Leak
HEALTHCARE PROFESSIONAL REPORTED TISSUE EXPANDER WAS "LEAKING FROM THE SAFE ZONE." DEVICE HAS BEEN EXPLANTED. THIS RECORD IS FOR THE LEFT SIDE.
Patient problems: Failure of Implant
MAUDE report 8951026 ↗InjuryNov 5, 2018Fluid/Blood Leak; Improper or Incorrect Procedure or Method; Fluid/Blood Leak; Improper or Incorrect Procedure or Method
PATIENT REPORTED RIGHT SIDE ACUTE CELLULITIS, IMPLANT IS HARD AND RED, THERE MAY BE A LEAK, A LOT OF BLEEDING, ASPIRATED 300CC OF BLOOD, YELLOW AND CLOUDY LIQUID, AND THE DOCTOR "REMOVED THE EXPANDER, RINSED AND REPLACED IT". DEVICE REMAINS IMPLANTED.
Patient problems: Cellulitis
MAUDE report 8039417 ↗
Other Expander, Skin, Inflatable devices
- B.BRAUN 7B3050B. Braun Medical Inc. · EXPANDER, SKIN, INFLATABLE1 recall40 reports
- MENTOR CPX 4 Breast Tissue Expander with Suture Tabs 3549314Mentor Texas L.P. · EXPANDER, SKIN, INFLATABLE1 recall22 reports
- MENTOR CPX4 PLUS, Smooth, Medium Height SCPX146MHMentor Texas L.P. · Expander, skin, inflatable1 recall18 reports
- Natrelle® Style 133 Tissue Expanders with Suture Tabs 133MX-13-TAllergan, Inc. · Expander, Skin, Inflatable1 recall16 reports
- MENTOR ARTOURA PLUS, Smooth, Ultra High Profile SDC135UHMentor Texas L.P. · Expander, skin, inflatable1 recall14 reports
- MENTOR ARTOURA PLUS, Smooth, High Profile SDC150HMentor Texas L.P. · Expander, skin, inflatable1 recall14 reports
- MENTOR ARTOURA PLUS, Smooth, Ultra High Profile SDC120UHMentor Texas L.P. · Expander, skin, inflatable1 recall14 reports
- Natrelle® Style 133 Tissue Expanders with Suture Tabs 133MV-14-TAllergan, Inc. · Expander, Skin, Inflatable1 recall13 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.