Natrelle 133S Tissue Expanders 133S-MV-15-TRecalls & reported problems

Allergan, Inc. · Natrelle TE Smooth 133S-MV-15-T (US) · Class U · Implant

All models of Natrelle Breast Implants (1167) · Natrelle 133S Tissue Expanders (25)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Natrelle 133S tissue expander is an inflatable implant used to stretch skin before breast reconstruction. The 29 reports mostly describe fluid or blood leaking from the expander, along with infections after surgery and the implant failing to stay in place; 27 reports involved injuries and no deaths were reported. Several mention tears, ruptures or a seam splitting open, and skin redness or tightening around the area. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
27
Recalls naming it
0

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 10888628043893), received between Jan 8, 2020 and Dec 10, 2025 — classified as 27 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (17 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak17
  • Adverse Event Without Identified Device or Use Problem7
  • Break2
  • Deflation Problem1
  • Device Appears to Trigger Rejection1
  • Device Contaminated During Manufacture or Shipping1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Failure of Implant21
  • Unspecified Infection6
  • Seroma3
  • Capsular Contracture1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 10, 2025Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED "RUPTURED AT THE PATCH". THIS RECORD IS FOR THE UNKNOWN SIDE. THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 23771936 ↗
  • InjuryNov 18, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "SOME AREAS OF CREASING OF THE EXPANDER" AND "SOME SKIN ERYTHEMA". LATER, HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURE BAKER GRADE III". PATIENT WAS TREATED WITH ACCOLATE AND IMPLANT REMOVAL. DEVICE WAS EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 23582928 ↗
  • InjuryJun 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE OF ALL LAST 3 CONSECUTIVE TISSUE EXPANDER BREAST RECONSTRUCTION DEVELOPED POST OP INFECTIONS". DEVICE WAS EXPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 22331750 ↗
  • InjuryJun 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE OF ALL LAST 3 CONSECUTIVE TISSUE EXPANDER BREAST RECONSTRUCTION DEVELOPED POST OP INFECTIONS". DEVICE REMAINS IMPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 22331738 ↗
  • InjuryMar 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED VIA NBIR LEFT SIDE INFECTION. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 21499585 ↗
  • InjuryNov 14, 2023Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED DEFLATION FOR AN UNKNOWN SIDE. DEVICE WAS NOT IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 18134800 ↗
  • InjuryOct 4, 2023Fluid/Blood Leak; Fluid/Blood Leak

    SALES REPRESENTATIVE REPORTED A "TEAR ON THE SEAM" OF THE DEVICE FOR AN UNKNOWN SIDE. DEVICE WAS NOT IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 17869913 ↗
  • InjuryMar 20, 2023Deflation Problem; Deflation Problem

    HEALTHCARE PROFESSIONAL REPORTED UNKNOWN SIDE TISSUE EXPANDER "LEAKING UPON REMOVAL". DEVICE EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 16576901 ↗
  • InjuryFeb 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE INFECTION. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Unspecified Infection

    MAUDE report 16452729 ↗
  • InjuryFeb 14, 2023Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE "DEFLATED TISSUE EXPANDERS DUE TO RADIATION". THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant; Seroma

    MAUDE report 16370943 ↗
  • InjuryFeb 14, 2023Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE "DEFLATED TISSUE EXPANDERS DUE TO RADIATION". THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant; Seroma

    MAUDE report 16370931 ↗
  • InjuryNov 4, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED " SUPERIOR-MEDIAL TAB HAD DETACHED FROM THE EXPANDER" AND SEROMA. DEVICE REMAINS IMPLANTED.

    Patient problems: Seroma

    MAUDE report 15739352 ↗
  • InjurySep 28, 2022Break; Break

    HEALTH CARE PROFESSIONAL REPORTED ONE OF THE TABS OF THE TISSUE EXPANDER WAS FREE FLOATING IN THE POCKET. DEVICE WAS NOT IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 15498521 ↗
  • MalfunctionAug 2, 2022Break; Break

    REPRESENTATIVE REPORTED ¿A TAB HAD RIPPED OFF THE TISSUE EXPANDER¿. PATIENT CONTACT WAS MADE WITH THE DEVICE. DEVICE IMPLANTED. SIDE UNKNOWN.

    Patient problems: Failure of Implant

    MAUDE report 15148024 ↗
  • InjuryMar 30, 2022Fluid/Blood Leak; Fluid/Blood Leak

    COMPANY REPRESENTATIVE REPORTED "DEFLATION" AGAINST UNKNOWN SIDE TISSUE EXPANDER. THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 13957594 ↗
  • InjuryMar 4, 2022Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 13667742 ↗
  • InjuryFeb 18, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED A INFECTED RIGHT EXPANDER. THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 13563750 ↗
  • InjuryFeb 2, 2022Fluid/Blood Leak; Improper or Incorrect Procedure or Method; Fluid/Blood Leak; Improper or Incorrect Procedure or Method

    PATIENT REPORTED RIGHT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 13424564 ↗
  • InjuryJan 27, 2022Fluid/Blood Leak; Fluid/Blood Leak

    PATIENT REPORTED RIGHT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 13367444 ↗
  • InjuryOct 6, 2021Fluid/Blood Leak; Fluid/Blood Leak

    PATIENT REPORTED UNKNOWN SIDE "TISSUES EXPANDER HAS BEEN LEAKING AND GOING FLAT.¿ DEVICE HAS BEEN IMPLANTED FOR MORE THAN SIX MONTHS. DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 12584083 ↗
  • InjuryJun 7, 2021Fluid/Blood Leak; Fluid/Blood Leak

    COMPANY REPRESENTATIVE REPORTED, UNKNOWN SIDE DEFLATION OF A TISSUE EXPANDER. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 11948027 ↗
  • MalfunctionMar 9, 2021Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    HEALTHCARE PROFESSIONAL REPORTED AN IMPLANT WITH "RANDOM MATERIAL STUCK ON IT." DEVICE WAS NOT IMPLANTED. SURGERY WAS COMPLETED WITH A BACK-UP DEVICE.

    Patient problems: Failure of Implant

    MAUDE report 11446541 ↗
  • InjuryFeb 19, 2021Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED AN UNKNOWN SIDE "RIPPED TE." DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 11351766 ↗
  • InjuryDec 7, 2020Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED FOR UNKNOWN SIDE "ONE OF MY DOCS INFORMED ME ABOUT A RUPTURED TISSUE EXPANDER." DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 10960459 ↗
  • InjuryNov 24, 2020Fluid/Blood Leak; Fluid/Blood Leak

    PATIENT REPORTED RIGHT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 10897029 ↗