Natrelle 133S Tissue Expanders 133S-MX-14-TRecalls & reported problems

Allergan, Inc. · Natrelle TE Smooth 133S-MX-14-T (US) · Class U · Implant

All models of Natrelle Breast Implants (1167) · Natrelle 133S Tissue Expanders (25)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Allergan Natrelle tissue expander is an inflatable implant used to stretch skin. Reports mostly describe the expander leaking and deflating from tiny holes or seam tears, sometimes after flipping out of position, often leading to removal, with a few infections mentioned. There are 26 reports naming its identifier, including no deaths and 21 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
25
Deaths reported
0
Injuries reported
20
Recalls naming it
0

Summary

The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 10888628044203), received between Oct 24, 2019 and Jul 2, 2025 — classified as 20 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (14 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak14
  • Adverse Event Without Identified Device or Use Problem7
  • Break4
  • Malposition of Device2
  • Defective Device1
  • Deflation Problem1
  • Device Handling Problem1

Patient problems

  • Failure of Implant16
  • Unspecified Infection4
  • Fluid Discharge2
  • Cellulitis1
  • Necrosis1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED "CELLULITIS OF RIGHT BREAST". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED.

    Patient problems: Cellulitis

    MAUDE report 22386264 ↗
  • MalfunctionJun 10, 2025Break; Malposition of Device; Break; Malposition of Device

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE EXPANDER GOT PUNCTURED AFTER FLIPPING. LATER HEALTHCARE PROFESSIONAL REPORTED " FLIPPED AFTER PLACEMENT". DEVICE WAS EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 22181062 ↗
  • InjuryJun 10, 2025Break; Malposition of Device; Device Handling Problem; Break; Malposition of Device; Device Handling Problem

    HEALTHCARE PROFESSIONAL REPORTED EXPANDER GOT PUNCTURED AFTER FLIPPING. LATER HEALTHCARE PROFESSIONAL REPORTED "FLIPPED AFTER PLACEMENT". THIS RECORD IS FOR THE LEFT SIDE. DEVICE WAS EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 22175003 ↗
  • InjuryJul 19, 2023Fluid/Blood Leak; Fluid/Blood Leak

    DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 17353932 ↗
  • InjuryMay 25, 2023Fluid/Blood Leak; Fluid/Blood Leak

    COMPANY REPRESENTATIVE REPORTED RIGHT SIDE "HAVE A PROBLEM WITH AN IMPLANT." LATER, HEALTHCARE PROFESSIONAL REPORTED "RIGHT TE HAD DEFLATED" AND "WE FOUND A VERY TINY HOLE". DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 17007627 ↗
  • InjuryNov 17, 2022Fluid/Blood Leak; Fluid/Blood Leak

    COMPANY REPRESENTATIVE REPORTED LEFT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED AND DISCARDED.

    Patient problems: Failure of Implant

    MAUDE report 15815424 ↗
  • MalfunctionNov 11, 2022Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED UNKNOWN SIDE TEAR IN THE SEAM, CAUSING THE DEVICE TO DEFLATE. DEVICE WAS EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 15779305 ↗
  • InjuryOct 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED "PATIENT INFECTION ¿ CULTURE INDICATED MOLD" FOR AN UNKNOWN SIDE. THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 15697493 ↗
  • InjuryOct 29, 2022Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED "LEAKING EXPANDER" FOR AN UNKNOWN SIDE. THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 15697292 ↗
  • InjuryMay 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED "INFECTION" AGAINST RIGHT DEVICE. THE DEVICE IS NOT IMPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 14323398 ↗
  • InjuryMar 8, 2022Fluid/Blood Leak; Fluid/Blood Leak

    COMPANY REPRESENTATIVE REPORTED "PUNCTURED" AGAINST UNKNOWN SIDE DEVICE. THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 13690613 ↗
  • InjuryAug 24, 2021Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED UNKNOWN SIDE TISSUE EXPANDER "DEFLATED WHILE PATIENT WAS IN RECOVERY." DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 12362670 ↗
  • InjuryJul 15, 2021Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED UNKNOWN SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 12176719 ↗
  • MalfunctionFeb 4, 2021Defective Device; Defective Device

    HEALTHCARE PROFESSIONAL REPORTED UNKNOWN SIDE "DEFECTIVE TISSUE EXPANDER, CIRCLE IS IN THE WRONG PLACE." DEVICE WAS NOT IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11279317 ↗
  • InjuryJan 7, 2021Deflation Problem; Fluid/Blood Leak; Deflation Problem; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED UNKNOWN SIDE "BELIEVED TO HAVE A LEAK". DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 11132291 ↗
  • InjurySep 28, 2020Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROVIDER REPORTED AN UNKNOWN SIDE "EXPANDER HAD DEFLATED." DEVICE HAS BEEN EXPLANTED AND DISCARDED AFTER SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 10592109 ↗
  • InjurySep 18, 2020Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROVIDER REPORTED "TISSUE EXPANDER NOT FILL IN CLINIC. WE HAD PUT IN 325CC IN THE TISSUE EXPANDER BUT IT WAS ALWAYS LOOKING MORE DEFLATED THAN THE OTHER SIDE." DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 10557027 ↗
  • InjurySep 4, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT REPRESENTATIVE REPORTED "LEAKAGE OF FLUID FROM THE BREAST,¿ ¿SWELLING,¿ ¿FEARS [PATIENT] WILL AGAIN DEVELOP CANCER,¿ ¿ENDURED CONTINUOUS EMOTIONAL DISTRESS,¿ ¿INCREASED RISK OF DEVELOPING BIA-ALCL,¿ ¿PAIN," "DISCOMFORT¿, ¿SORENESS,¿, AND THE DEVICE WAS IMPLANTED LONGER THAN INDICATED IN DEVICE LABELING. PATIENT REPRESENTATIVE ALSO REPORTED ¿SUFFERING,¿ THIS IS NOT DEVICE RELATED. DEVICE HAS BEEN EXPLANTED. THIS RECORD IS FOR THE LEFT SIDE.

    Patient problems: Fluid Discharge

    MAUDE report 10493101 ↗
  • InjurySep 4, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT REPRESENTATIVE REPORTED "LEAKAGE OF FLUID FROM THE BREAST,¿ ¿SWELLING,¿ ¿FEARS [PATIENT] WILL AGAIN DEVELOP CANCER,¿ ¿ENDURED CONTINUOUS EMOTIONAL DISTRESS,¿ ¿INCREASED RISK OF DEVELOPING BIA-ALCL,¿ ¿PAIN," "DISCOMFORT¿, ¿SORENESS,¿, AND THE DEVICE WAS IMPLANTED LONGER THAN INDICATED IN DEVICE LABELING. PATIENT REPRESENTATIVE ALSO REPORTED ¿SUFFERING,¿ THIS IS NOT DEVICE RELATED. DEVICE HAS BEEN EXPLANTED. THIS RECORD IS FOR THE RIGHT SIDE.

    Patient problems: Fluid Discharge

    MAUDE report 10493090 ↗
  • MalfunctionSep 1, 2020Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED VIA REGULATORY AGENCY A "PUNCTURE IN THE FILL PORT". DEVICE WAS NOT IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 10475959 ↗
  • InjuryAug 19, 2020Break; Break

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE "NECROSIS" AND "UPON OPENING THE SKIN NOTICED ONE OF THE TABS HAD COME OFF OF THE EXPANDER, SURGEON NEEDED TO SCRAP EVERYTHING, AND START OVER WITH A NEW DEVICE." DEVICE HAS BEEN EXPLANTED AND DISCARDED AFTER SURGERY.

    Patient problems: Necrosis

    MAUDE report 10424532 ↗
  • MalfunctionJul 17, 2020Break; Fluid/Blood Leak; Break; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED UNKNOWN SIDE "NATRELLE TISSUE EXPANDER WOULD NOT COMPLETELY FILL, IT UNKNOWINGLY GOT PUNCTURED, BUT THERE WERE NO RUPTURES OR LEAKS. IT JUST WASN¿T GETTING THE VOLUME NEEDED." DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 10289902 ↗
  • InjuryFeb 28, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "PSEUDOMONAS INFECTION REQUIRING REMOVAL." DEVICE REMAINS IMPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 9769605 ↗
  • InjuryFeb 24, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE "PSEUDOMONAS INFECTION." DEVICE REMAINS IMPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 9745552 ↗
  • InjuryOct 24, 2019Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE "2 MM HOLE SEVERAL CENTIMETERS MEDIAL TO THE FILL PORT." DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 9234072 ↗

Other Expander, Skin, Inflatable devices

All Expander, Skin, Inflatable devices →