Natrelle 133S Tissue Expanders 133S-MX-15-TRecalls & reported problems

Allergan, Inc. · Natrelle TE Smooth 133S-MX-15-T (US) · Class U · Implant

All models of Natrelle Breast Implants (1167) · Natrelle 133S Tissue Expanders (25)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Natrelle 133S Smooth tissue expander is an inflatable silicone implant placed under the skin to stretch tissue before breast reconstruction. The 28 reports mostly describe fluid leaking out or the expander breaking, tearing or deflating, sometimes with holes in it, and several needing removal and replacement because of infection or rupture. No deaths were reported; 22 reports involved injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
28
Deaths reported
0
Injuries reported
22
Recalls naming it
0

Summary

The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 10888628044210), received between Dec 20, 2019 and Jul 27, 2026 — classified as 22 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak19
  • Adverse Event Without Identified Device or Use Problem4
  • Break4
  • Deflation Problem1
  • Delivered as Unsterile Product1
  • Material Rupture1

Patient problems

  • Failure of Implant19
  • Unspecified Infection4
  • No Clinical Signs, Symptoms or Conditions3
  • Seroma2
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 27, 2026Fluid/Blood Leak

    Healthcare Professional reported through sales representative "several holes in it". This record is for the right side. Device was explanted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50037813 ↗
  • MalfunctionDec 3, 2025Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED "TISSUE EXPANDER LEAK". THE RECORD IS FOR A DEVICE THAT WASN'T IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23707101 ↗
  • InjurySep 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT REPORTED "INFECTION TE EXCHANGE". THIS RECORD IS FOR THE LEFT SIDE. THE DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: Unspecified Infection

    MAUDE report 23011388 ↗
  • InjurySep 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT REPORTED "INFECTION TE EXCHANGE". THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: Unspecified Infection

    MAUDE report 22985920 ↗
  • InjuryApr 15, 2025Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "DEFLATION". DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 21845534 ↗
  • InjuryNov 17, 2023Fluid/Blood Leak; Fluid/Blood Leak

    SALES REPRESENTATIVE REPORTED ON BEHALF OF THE HEALTHCARE PROFESSIONAL THAT THE DEVICE "HAD A HOLE". THIS IS FOR A UNKNOWN SIDE. DEVICE IS NOT IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 18161100 ↗
  • InjuryOct 17, 2023Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE TISSUE EXPANDER TORE/SEPARATED AT PORT AND SILICONE INTERFACE, NOTICEABLY DEFLATED. TISSUE EXPANDER HAS BEEN EXPLANTED AND REPLACED WITH IMPLANT.

    Patient problems: Failure of Implant

    MAUDE report 17950221 ↗
  • InjuryJun 30, 2023Fluid/Blood Leak; Material Rupture; Fluid/Blood Leak; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED RUPTURE. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 17240828 ↗
  • InjuryApr 24, 2023Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED A "LEFT SIDE DEFLATION" AND "SEVERAL HOLES". DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 16796104 ↗
  • InjuryDec 20, 2022Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED "LEFT BREAST TISSUE EXPANDER WAS QUITE A BIT SOFTER THAN IT HAD BEEN", "THE VOLUME SEEMED TO HAVE DECREASED", "RUPTURED LEFT BREAST TISSUE EXPANDER", AND "SOME SEROUS FLUID WITHIN THE TISSUE EXPANDER AND IT WAS NOTED TO HAVE FAILED AT A POSTERIOR SEAM OF THE 2 PIECE SHELL, AT ABOUT THE 1 O'CLOCK POSITION WHEN VIEWED ANTERIORLY". DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Seroma

    MAUDE report 16017075 ↗
  • InjuryDec 12, 2022Deflation Problem; Fluid/Blood Leak; Deflation Problem; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED UNKNOWN SIDE "NOTICED A HOLE IN THE SEAM". DEVICE WAS EXPLANTED AND REPLACED WITH A BACKUP.

    Patient problems: Failure of Implant; Seroma

    MAUDE report 15962137 ↗
  • InjuryAug 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED VIA DEVICE TRACKING "INFECTION" AGAINST LEFT DEVICE. THE DEVICE HAS BEEN REPLACED.

    Patient problems: Unspecified Infection

    MAUDE report 15324433 ↗
  • MalfunctionApr 28, 2022Break; Break

    HEALTH CARE PROFESSIONAL REPORTED RIGHT SIDE "DURING IMPLANTATION SURGERY DOCTOR NOTICED DAMAGE NEAR THE 12 O'CLOCK TAB ON THE SEAM WHERE THE MAGNETIC PORT IS LOCATED.¿ THE DEVICE CAME IN CONTACT WITH THE PATIENT BUT WAS NOT IMPLANTED. SURGERY WAS NOT COMPLETED WITH A BACK UP DEVICE BUT RATHER WITH A BREAST IMPLANT.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 14245502 ↗
  • InjuryFeb 11, 2022Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "DEFLATED TISSUE EXPANDER.¿ DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 13517252 ↗
  • InjuryFeb 9, 2022Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED "PATIENT THAT TWO DEFLATED TISSUE EXPANDERS.¿ DEVICE REMAINS IMPLANTED. LEFT SIDE.

    Patient problems: Failure of Implant

    MAUDE report 13501650 ↗
  • InjuryOct 8, 2021Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 12599143 ↗
  • InjuryMar 3, 2021Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED "TISSUE EXPANDER LEAKING." THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 11407220 ↗
  • InjuryFeb 9, 2021Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED UNKNOWN SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 11301396 ↗
  • MalfunctionDec 4, 2020Break; Break

    HEALTHCARE PROFESSIONAL REPORTED THAT WHILE IMPLANTING TISSUE EXPANDER, DEVICE RUPTURED. SURGERY WAS COMPLETED SUCCESSFULLY WITH A BACK UP DEVICE. DEVICE WAS NOT IMPLANTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10944145 ↗
  • MalfunctionNov 13, 2020Delivered as Unsterile Product; Delivered as Unsterile Product

    HEALTHCARE PROFESSIONAL REPORTED A TISSUE EXPANDER THAT "DID NOT HAVE AN INNER STERILE WRAPPING." DEVICE WAS NOT IMPLANTED.

    Patient problems: No Patient Involvement

    MAUDE report 10837306 ↗
  • InjuryNov 4, 2020Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE "TEAR IN EXPANDER ALONG SEAM." DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 10784858 ↗
  • MalfunctionNov 3, 2020Break; Break

    HEALTHCARE PROFESSIONAL REPORTED "EXPANDER WAS PUT IN THE PATIENT, BUT WOULD NOT PROPERLY FILL, IT WAS LIKE IT WAS 'USED CLOSED' ." DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 10781743 ↗
  • MalfunctionSep 29, 2020Break; Break

    HEALTHCARE PROFESSIONAL REPORTED A DEVICE "APPEARED TO BE DEFLATING DURING THE EXPANSION PROCESS", AND "THEY WERE NOT ABLE TO PUT MUCH VOLUME IN THIS EXPANDER AT THE TIME OF IMPLANTATION." DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 10599448 ↗
  • InjurySep 14, 2020Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED ¿RIGHT SIDE DEFLATION¿. DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 10531894 ↗
  • InjuryAug 28, 2020Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED "DEFLATION" WITH A TISSUE EXPANDER FOR AN UNKNOWN SIDE. DEVICE WAS EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 10468710 ↗

Other Expander, Skin, Inflatable devices

All Expander, Skin, Inflatable devices →