Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10888857136168), received between Sep 22, 2016 and Mar 9, 2022 — classified as 5 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture5
- Break1
- Migration1
Patient problems
- Pain3
- Injury2
- No Clinical Signs, Symptoms or Conditions1
- No Information1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryMar 9, 2022Migration; Migration
IT WAS REPORTED THAT APPROXIMATELY TWO-MONTHS POST-OPERATIVELY THE MESA SCREWS AT RIGHT/LEFT L2 AND LEFT L3 MIGRATED. ALL THE RODS AND SCREWS WERE REMOVED AND REPLACED AND THE CONSTRUCT WAS EXTENDED. THIS REPORT CAPTURES THE SCREW AT LEFT L2.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13714792 ↗InjuryAug 25, 2020Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT BEGAN TO EXPERIENCE "GROWING" BACK PAIN INTRA-OPERATIVELY. DURING REVISION, TWO MESA POLYAXIAL SCREWS WERE OBSERVED TO BE FRACTURED APPROXIMATELY 1CM BENEATH THE SCREW HEADS. THE SURGEON TRIED TO RETRIEVE THE SCREWS, BUT THEY WERE LEFT IN AS THE ONLY WAY TO RETRIEVE THE SCREWS WAS TO DRILL AWAY THE PEDICLE. THE SURGEON WAS ABLE TO SUCCESSFULLY COMPLETE THE SURGERY, WITH AN UNKNOWN DELAY, BY EXTENDING THE CONSTRUCT ONE LEVEL FURTHER DOWN. THIS REPORT REPRESENTS THE SECOND OF THE TWO SCREWS.
Patient problems: Pain
MAUDE report 10451968 ↗InjuryAug 25, 2020Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT BEGAN TO EXPERIENCE "GROWING" BACK PAIN INTRA-OPERATIVELY. DURING REVISION, TWO MESA POLYAXIAL SCREWS WERE OBSERVED TO BE FRACTURED APPROXIMATELY 1 CM BENEATH THE SCREW HEADS. THE SURGEON TRIED TO RETRIEVE THE SCREWS, BUT THEY WERE LEFT IN AS THE ONLY WAY TO RETRIEVE THE SCREWS WAS TO DRILL AWAY THE PEDICLE. THE SURGEON WAS ABLE TO SUCCESSFULLY COMPLETE THE SURGERY, WITH AN UNKNOWN DELAY, BY EXTENDING THE CONSTRUCT ONE LEVEL FURTHER DOWN. THIS REPORT REPRESENTS THE FIRST OF THE TWO SCREWS.
Patient problems: Pain
MAUDE report 10451966 ↗InjuryApr 14, 2020Fracture; Fracture
IT WAS REPORTED THAT TWO MESA POLYAXIAL SCREWS FRACTURED POST-OPERATIVELY. THE PATIENT WAS REVISED. THIS REPORT REPRESENTS THE FIRST OF THE TWO SCREWS.
Patient problems: Injury
MAUDE report 9960628 ↗InjuryApr 14, 2020Fracture; Fracture
IT WAS REPORTED THAT TWO MESA POLYAXIAL SCREWS FRACTURED POST-OPERATIVELY. THE PATIENT WAS REVISED. THIS REPORT REPRESENTS THE SECOND OF THE TWO SCREWS.
Patient problems: Injury
MAUDE report 9960467 ↗MalfunctionMay 26, 2017Fracture; Fracture
ON (B)(6) 2017 IT WAS REPORTED TO K2M, INC. THAT A PATIENT PRESENTED WITH SCREW FRACTURES. THE PATIENT REPORTEDLY HAD SCREW FRACTURES APPROXIMATELY 12 MONTHS POST-OP. THE PATIENT HAS NOT BEEN REVISED AT THIS TIME.
Patient problems: No Information
MAUDE report 6595523 ↗MalfunctionSep 22, 2016Break; Break
ON (B)(6) 2016 IT WAS REPORTED TO K2M, INC. THAT A SCREW FRACTURED. THE PATIENT REPORTEDLY HAD A FRACTURED SCREW APPROXIMATELY 12 MONTHS POST-OP.
Patient problems: Pain
MAUDE report 5969444 ↗
Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page