Navigated Spine Instruments 7601-N90003-KRecalls & reported problems
VB Spine LLC · Navigated Screw Inserter, Yukon · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10888857514782), received between Jan 26, 2022 and Dec 23, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Failure to Eject” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Eject4
- Difficult to Remove2
- Fracture2
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 23, 2025Fracture; Difficult to Remove; Fracture; Difficult to Remove
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23882053 ↗MalfunctionDec 23, 2025Fracture; Difficult to Remove; Fracture; Difficult to Remove
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23877174 ↗MalfunctionOct 31, 2022Failure to Eject; Failure to Eject
THIS RECORD IS A CONSOLIDATION OF Q3 2022 RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. THIS REPORT CAPTURES TWO INSTANCE OF INTRAOPERATIVE YUKON NAVIGATED SCREW INSERTER JAMMING IN WHICH A PHYSICIAN HAD DIFFICULTY DISSOCIATING THE TIP OF THE INSERTER FROM THE SCREW AFTER SCREW INSERTION. THE ASSOCIATED PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO A PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15700822 ↗MalfunctionJul 29, 2022Failure to Eject; Failure to Eject
THIS RECORD IS A CONSOLIDATION OF Q2 2022 RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. THIS REPORT CAPTURES ONE INSTANCE OF INTRAOPERATIVE YUKON SCREW INSERTER COUPLING ISSUE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A TWO-MINUTE SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15135881 ↗MalfunctionApr 29, 2022Failure to Eject; Failure to Eject
THIS RECORD IS A CONSOLIDATION OF Q1 2022 RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. THIS REPORT CAPTURES FOUR INSTANCES OF INTRAOPERATIVE YUKON NAVIGATED SCREW INSERTER TIP COUPLING ISSUES (JAMMING). IN ALL CASES, THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14253240 ↗MalfunctionJan 26, 2022Mechanical Jam; Failure to Eject; Mechanical Jam; Failure to Eject
THIS REPORT SUMMARIZES TWO MALFUNCTION EVENTS WHERE THE TIPS OF NAVIGATED YUKON OCT POLYAXIAL SCREW INSERTERS WERE STICKING WHEN TRYING TO DISENGAGE FROM THE SCREW. SURGERIES WERE SUCCESSFULLY COMPLETED WITH NO DELAY. NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13362048 ↗
Other Orthopedic stereotaxic instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page