Sunmed 7-6507-12Recalls & reported problems
Sunmed, LLC · Foley Latex 2 Way 30cc 12fr · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10889483005507), received between Apr 8, 2025 and Apr 8, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Inflation Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Inflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 8, 2025Inflation Problem; Inflation Problem
SUNMED HOLDINGS LLC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING TWO DIFFERENT PATIENTS. THIS IS THE FIRST OF TWO REPORTS. REFER TO 1314417-2025-00024 FOR THE SECOND REPORT. DURING SURGERY THE BALLOON DID NOT DEFLATE COMPLETELY; THE PROBE WAS REPLACED; THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21798236 ↗
Other Catheter, Retention Type, Balloon devices
- Bardex® All-Silicone Foley Catheter 165816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON512 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON476 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 1758SI16C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON1 death383 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON234 reports
- Bardex® Lubri-Sil® All-Silicone Temp Sensing Foley Catheter 119216MC. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON167 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175814C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON153 reports
- Bardex® All-Silicone Foley Catheter 165812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON151 reports
- SureStep™ Foley Tray System Bard® Lubri-Sil® Foley Catheter Tray A942216C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON2 recalls46 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page