Medline Industries, Inc. DYNJAA10601Recalls & reported problems
Medline Industries, Inc. · ANESTHESIA CIRC ADULT 120-LF · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10889942655434), received between Nov 25, 2025 and Nov 25, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Crack” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 10, 2025Class IOpen, Classified
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Z-0442-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Crack1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 25, 2025Crack; Crack
ACCORDING TO THE CUSTOMER, THERE HAVE BEEN MULTIPLE INCIDENTS OF THE CIRCUIT "CRACKING" DURING USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23645470 ↗
Other Circuit, breathing (w connector, adaptor, y piece) devices
- Medline Industries, Inc. DYNJ905503FMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)3 recalls0 reports
- Medline Industries, Inc. PHS972096014BMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)3 recalls0 reports
- Medline Industries, Inc. DYNJ0185290GMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)3 recalls0 reports
- Medline Industries, Inc. DYNJ905503GMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
- Medline Industries, Inc. DYNJAA262AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
- Medline Industries, Inc. DYNJ905567AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
- MEDLINE INDUSTRIES, INC. DYNJ86657AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL199AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page