Anesthesia Circuit Kits and other various kits that contain… recallZ-0442-2026 · Medline Industries, LP · initiated Oct 10, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Safety notice was provided to customers on 10/10/2025. Follow-up "URGENT MEDICAL DEVICE RECALL" notification dated 10/21/2025 was sent out on 10/21/2025. Notification was through email and first-class mail. Customers were instructed to Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product and destroy affected kits on hand. No products are to be returned to Medline. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-25-215-FG Recall Code: If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. An expanded "URGENT MEDICAL DEVICE RECALL" notification dated 10/24/2025 was provided to customers on 10/24/25, which included additional affected products. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-0442-2026
- Status
- Open, Classified
- Initiated by the company
- Oct 10, 2025
- Posted by the FDA
- Nov 13, 2025
- Recalling firm
- Medline Industries, LP
- Quantity
- 96,406 total units
- Product code
- CAI
- Distribution
- International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
Product description
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP-LF DYNJAA6555G; ANES CIRCUIT ADULT 120"-LF DYNJAA10817; ANESTHESIA 120" ADLT-LF DYNJAA10775; ANESTHESIA CIRC ADULT 120-LF DYNJAA10601; ANESTHESIA CIRCUIT - LF DYNJAA10261B; ANESTHESIA CIRCUIT 120EXP-LF DYNJAA10741A; ANESTHESIA CIRCUIT ADULT-LF DYNJAA0380A DYNJAA10641 DYNJAA10706B DYNJAA10851 DYNJAA10922B DYNJAA10971A DYNJAA11030A DYNJAA11087 DYNJAA11126; ANESTHESIA CIRCUIT-LF DYNJAA0309A DYNJAA0368A DYNJAA10368A DYNJAA10392 DYNJAA10430A DYNJAA10547A DYNJAA10639A DYNJAA10656A DYNJAA10685A DYNJAA10685B; CIRC 120" EXP 1 BV HMEF 3L GSL DYNJAA10920; EXPANDABLE CIRCUIT-LF DYNJAA11095; NM ANESTHESIA TURNOVER KIT DYNJAA0433; PROV HEALTH HEPA CIRC #2-LF DYNJAA0360; SPTMPX01IA CIRCUIT ADULT-LF DYNJAA10973A;
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
DYNJAA0309A UDI-DI 40193489400268 (case), 10193489400267 (ea) Lots 25EBU353 25GBI473; DYNJAA0360 UDI DI 40193489419000 (case), 10193489419009 (ea) Lots 25EBU304 25GBQ732; DYNJAA0368A UDI-DI 40195327092550 (case), 10195327092559 (ea) Lots 25GBH582; DYNJAA0380A UDI-DI 40195327478286 (case), 10195327478285 (ea) Lots 25FBB950 25FBT875 25GBX623; DYNJAA0433 UDI-DI 40198459320706 (case), 10198459320705 (ea) Lots 25EBH621 25EBQ391; DYNJAA10261B UDI- DI 40193489430784 (case), 10193489430783 (ea) Lots 25FBJ760; DYNJAA10368A UDI-DI 40198459320720 (case), 10198459320729 (ea) Lots 25GBY374 25IBK072; DYNJAA10392 UDI-DI 40888277962201 (case), 10888277962200 (ea) Lots 25FBN032; DYNJAA10430A UDI-DI 40193489832052 (case), 10193489832051 (ea) Lots 25FBW335 25IBH048; DYNJAA10547A UDI-DI 40193489603188 (case), 10193489603187 (ea) 25FBT853; DYNJAA10591 UDI-DI 40889942632535 (case), 10889942632534 (ea) Lots 25FBD876 25FBD877 25GBJ254 25GBJ256; DYNJAA10601 UDI-DI 40889942655435 (case), 10889942655434 (ea) Lots 25FBN184 25HBM893; DYNJAA10639A UDI-DI 40888277761958 (case), 10888277761957 (ea) Lots 25FBN043; DYNJAA10641 UDI-DI 40889942841227 (case), 10889942841226 (ea) Lots 25EBK360; DYNJAA10656A UDI-DI 40193489455527 (case), 10193489455526 (ea) Lots 25FBN055; DYNJAA10685A UDI-DI 40195327279142 (case), 10195327279141 (ea) Lots 25EBJ456; DYNJAA10685B UDI-DI 40198459372293 (case), 10198459372292 (ea) Lots 25FBL121 25FBT898 25FBT899; DYNJAA10706B UDI-DI 40193489485982 (case), 10193489485981 (ea) Lots 25FBH921 25GBK879 25GBN124 25GBN125; DYNJAA10741A UDI-DI 40193489250573 (case), 10193489250572 (ea) Lots 25EBK364 25FBQ326 25HBA472; DYNJAA10775 UDI-DI 40193489348102 (case), 10193489348101 (ea) Lots 25GBY311 25IBD803; DYNJAA10817 UDI-DI 40193489365338 (case), 10193489365337 (ea) Lots 25GBO619; DYNJAA10851 UDI-DI 40193489429467 (case), 10193489429466 (ea) Lots 25GBE789 25IBL075; DYNJAA10920 UDI-DI 40193489463386 (case), 10193489463385 (ea) Lots 25FBT844 25GBP694; DYNJAA10922B UDI-DI 40198459301880 (case), 10198459301889 (ea) Lots 25FBQ442 25GBO675; DYNJAA10971A UDI-DI 40195327298167 (case), 10195327298166 (ea) Lots 25GBA751; DYNJAA10973A UDI-DI 40195327372423 (case), 10195327372422 (ea) Lots 25FBQ349; DYNJAA11030A UDI-DI 40195327341030 (case), 10195327341039 (ea) Lots 25FBQ346; DYNJAA11087 UDI-DI 40195327694402 (case), 10195327694401 (ea) Lots 25FBG942 25FBG944; DYNJAA11095 UDI-DI 40195327520992 (case), 10195327520991 (ea) Lots 25GBY361; DYNJAA11126 UDI-DI 40195327695508 (case), 10195327695507 (ea) Lots 25EBV629 25GBL211; DYNJAA6555G UDI-DI 40195327214648 (case), 10195327214647 (ea) Lots 25EBG518 25FBS747 25FBS748 25HBE691
Devices named in this recall
30 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJAA0380AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall5 reports
- Medline Industries, Inc. DYNJAA10641Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall5 reports
- MEDLINE INDUSTRIES, INC. DYNJAA10685BMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall3 reports
- Medline Industries, Inc. DYNJAA10261BMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall1 report
- Medline Industries, Inc. DYNJAA10430AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall1 report
- Medline Industries, Inc. DYNJAA6555GMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall1 report
- Medline Industries, Inc. DYNJAA10601Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall1 report
- Medline Industries, Inc. DYNJAA10741AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA10775Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA10817Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA0309AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA0360Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA10851Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA10656AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA10920Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA10706BMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA10547AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA0368AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA10685AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA10971AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA11030AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA10973AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA11095Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA11087Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA11126Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA0433Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA10368AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA10639AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA10392Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
- Medline Industries, Inc. DYNJAA10591Medline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97882).
- Z-0443-2026 · Open, ClassifiedAnesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; BREATHING KIT-LF DYNJ10550
- Z-0444-2026 · Open, ClassifiedAnesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; OPEN HEART SUPPLEMENT DYNJ909116D; VASCULAR DYNJ909124F
- Z-0445-2026 · Open, ClassifiedAnesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; ENT DYNJ909128I; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D
- Z-0446-2026 · Open, ClassifiedDYNJ909119G ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909127F CYSTO DYNJ909129D GENERAL LAPAROSCOPY-LF
- Z-0447-2026 · Open, ClassifiedConvenience Kit, BREAST HERNIA DYNJ909125G
- Z-0448-2026 · Open, ClassifiedConvenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET UP DYNJ909308D HIP PINNING
- Z-0449-2026 · Open, ClassifiedAnesthesia circuit kit; CIRCUIT,ANES,EXTENDFLEX,120,LF DYNJAA10230
- Z-0450-2026 · Open, ClassifiedConvenience kits; DYNJ901046M GYN ROBOTIC DYNJ909120D ROBOTIC PROSTATE
- Z-0451-2026 · Open, ClassifiedConvenience kits; DYNJ901047R ORTHO SPINE FUMICH DYNJ902308K TOTAL JOINT DYNJ909111F DR SALMA'S SPINE DYNJ909117G PATTERSON TOTAL JOINT DYNJ909113F KYPHOPLASTY
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page