Medline Industries, Inc. DYNJAA10641Recalls & reported problems

Medline Industries, Inc. · ANESTHESIA CIRCUIT ADULT-LF · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10889942841226), received between Nov 5, 2025 and Nov 6, 2025 — classified as 5 injuries.

The most frequently reported specific device problem is “Gas/Air Leak” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 5 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 10, 2025Class IOpen, Classified

    Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

    Z-0442-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Gas/Air Leak5
  • Crack1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 6, 2025Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE CUSTOMER, "THE ANESTHESIA CIRCUITS APPEAR TO BE EITHER CUT OR DAMAGED" CAUSING THE "MACHINES TO ALARM WITH A PRESSURE LEAK". THE CUSTOMER DID NOT REPORT ANY PATIENT IMPACT OR ADVERSE HEALTH EFFECT RELATED TO THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23491190 ↗
  • InjuryNov 6, 2025Crack; Gas/Air Leak; Crack; Gas/Air Leak

    ACCORDING TO THE CUSTOMER, THE TUBING "CRACKED" AND "LEAKED DURING USE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23487695 ↗
  • InjuryNov 6, 2025Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE CUSTOMER, "THE ANESTHESIA CIRCUITS APPEAR TO BE EITHER CUT OR DAMAGED" CAUSING THE "MACHINES TO ALARM WITH A PRESSURE LEAK". THE CUSTOMER DID NOT REPORT ANY PATIENT IMPACT OR ADVERSE HEALTH EFFECT RELATED TO THE REPORTED ISSUE. THE CUSTOMER DID NOT REPORT ANY SERIOUS INJURY, MEDICAL INTERVENTION, OR FOLLOW-UP CARE RELATED TO THE REPORTED INCIDENT. NO ADDITIONAL DETAILS ARE AVAILABLE RELATED TO THE CUSTOMER REPORTED ISSUE. IT HAS BEEN DETERMINED THAT THE REPORTED EVENT COULD CAUSE OR CONTRIBUTE TO SERIOUS INJURY IF IT WERE TO OCCUR AGAIN. IN AN ABUNDANCE OF CAUTION, THIS MEDWATCH IS BEING FILED. IF ANY FURTHER RELEVANT INFORMATION IS IDENTIFIED OR OBTAINED, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23486268 ↗
  • InjuryNov 6, 2025Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE CUSTOMER, THE TUBING "CRACKED" AND "LEAKED DURING USE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23486263 ↗
  • InjuryNov 5, 2025Gas/Air Leak; Material Split, Cut or Torn; Gas/Air Leak; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE ANESTHESIA CIRCUIT APPEARED CUT OR DAMAGED, TRIGGERING MACHINE ALARMS FOR PRESSURE LEAKS UPON PATIENT CONNECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23478438 ↗

Other Circuit, breathing (w connector, adaptor, y piece) devices

All Circuit, breathing (w connector, adaptor, y piece) devices →

Sources and updates