Malibu™ 12-1060Recalls & reported problems

Seaspine Orthopedics Corporation · PRECONTOURED ROD, 5.5 x 60mm · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10889981029975), received between Mar 16, 2020 and Mar 16, 2020 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Corroded” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Corroded2
  • Material Integrity Problem2

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 16, 2020Corroded; Material Integrity Problem; Corroded; Material Integrity Problem

    ON (B)(6) 2020 SEASPINE WAS INFORMED OF A MALIBU SPINAL SYSTEM REMOVAL SURGERY. THE INDEX SURGERY WAS (B)(6) 2008. THE PRIMARY REASON FOR THE REMOVAL WAS PATHOLOGY AT AN ADJACENT LEVEL AND NOT DUE TO FAILURE OF THE IMPLANT. DURING THE REMOVAL, CORROSION WAS DISCOVERED WHERE THE ROD AND SCREW INTERFACED. THE DISTRIBUTOR EXPRESSED THE SURGEON'S CONCERN REGARDING THE MALIBU SCREW RECALL THAT TOOK PLACE SHORTLY AFTER THE INDEX SURGERY. THE REFERENCED RECALL WAS DUE TO THE MANUFACTURER USING THE WRONG METAL INSIDE THE TULIP HEAD OF THE SCREW.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9835527 ↗
  • MalfunctionMar 16, 2020Corroded; Material Integrity Problem; Corroded; Material Integrity Problem

    ON (B)(6) 2020 SEASPINE WAS INFORMED OF A MALIBU SPINAL SYSTEM REMOVAL SURGERY. THE INDEX SURGERY WAS (B)(6) 2008. THE PRIMARY REASON FOR THE REMOVAL WAS PATHOLOGY AT AN ADJACENT LEVEL AND NOT DUE TO FAILURE OF THE IMPLANT. DURING THE REMOVAL, CORROSION WAS DISCOVERED WHERE THE ROD AND SCREW INTERFACED. THE DISTRIBUTOR EXPRESSED THE SURGEON'S CONCERN REGARDING THE MALIBU SCREW RECALL THAT TOOK PLACE SHORTLY AFTER THE INDEX SURGERY. THE REFERENCED RECALL WAS DUE TO THE MANUFACTURER USING THE WRONG METAL INSIDE THE TULIP HEAD OF THE SCREW.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9835516 ↗

Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices

All Orthosis, spinal pedicle fixation, for degenerative disc disease devices →