Daytona® Small Stature 79-0002Recalls & reported problems

Seaspine Orthopedics Corporation · CoCr Locking Cap 4.5mm, Small Stature, T25, BT · Class 2

All models of Daytona Small Stature (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
13
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10889981185640), received between Sep 3, 2024 and Sep 3, 2024 — classified as 13 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (13 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 13 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 29, 2024Class IIOpen, Classified

    Due to manufacturing defects observed on Locking Caps that affect the ability for components to mate.

    Z-0751-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Insufficient Device Problem Information13

Patient problems

  • Insufficient Information13

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 5 OF 5 UNITS OF PART 79-0002, LOT MM1002742E.

    Patient problems: Insufficient Information

    MAUDE report 20130654 ↗
  • MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 4 OF 5 UNITS OF PART 79-0002, LOT MM1002742E.

    Patient problems: Insufficient Information

    MAUDE report 20130648 ↗
  • MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 3 OF 5 UNITS OF PART 79-0002, LOT MM1002742E.

    Patient problems: Insufficient Information

    MAUDE report 20130647 ↗
  • MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 2 OF 5 UNITS OF PART 79-0002, LOT MM1002742E.

    Patient problems: Insufficient Information

    MAUDE report 20130646 ↗
  • MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 1 OF 5 UNITS OF PART 79-0002, LOT MM1002742E.

    Patient problems: Insufficient Information

    MAUDE report 20130641 ↗
  • MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 7 OF 8 UNITS OF PART 79-0002, LOT MM1003065E.

    Patient problems: Insufficient Information

    MAUDE report 20130639 ↗
  • MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 6 OF 8 UNITS OF PART 79-0002, LOT MM1003065E.

    Patient problems: Insufficient Information

    MAUDE report 20130638 ↗
  • MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 8 OF 8 UNITS OF PART 79-0002, LOT MM1003065E.

    Patient problems: Insufficient Information

    MAUDE report 20130637 ↗
  • MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 4 OF 8 UNITS OF PART 79-0002, LOT MM1003065E.

    Patient problems: Insufficient Information

    MAUDE report 20130633 ↗
  • MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 5 OF 8 UNITS OF PART 79-0002, LOT MM1003065E.

    Patient problems: Insufficient Information

    MAUDE report 20130632 ↗
  • MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 2 OF 8 UNITS OF PART 79-0002, LOT MM1003065E.

    Patient problems: Insufficient Information

    MAUDE report 20130631 ↗
  • MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 1 OF 8 UNITS OF PART 79-0002, LOT MM1003065E.

    Patient problems: Insufficient Information

    MAUDE report 20130630 ↗
  • MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 3 OF 8 UNITS OF PART 79-0002, LOT MM1003065E.

    Patient problems: Insufficient Information

    MAUDE report 20130628 ↗

Other Thoracolumbosacral pedicle screw system devices

All Thoracolumbosacral pedicle screw system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page