Daytona® Small Stature 79-0002Recalls & reported problems
Seaspine Orthopedics Corporation · CoCr Locking Cap 4.5mm, Small Stature, T25, BT · Class 2
All models of Daytona Small Stature (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10889981185640), received between Sep 3, 2024 and Sep 3, 2024 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (13 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 29, 2024Class IIOpen, Classified
Due to manufacturing defects observed on Locking Caps that affect the ability for components to mate.
Z-0751-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Insufficient Device Problem Information13
Patient problems
- Insufficient Information13
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 5 OF 5 UNITS OF PART 79-0002, LOT MM1002742E.
Patient problems: Insufficient Information
MAUDE report 20130654 ↗MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 4 OF 5 UNITS OF PART 79-0002, LOT MM1002742E.
Patient problems: Insufficient Information
MAUDE report 20130648 ↗MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 3 OF 5 UNITS OF PART 79-0002, LOT MM1002742E.
Patient problems: Insufficient Information
MAUDE report 20130647 ↗MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 2 OF 5 UNITS OF PART 79-0002, LOT MM1002742E.
Patient problems: Insufficient Information
MAUDE report 20130646 ↗MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 1 OF 5 UNITS OF PART 79-0002, LOT MM1002742E.
Patient problems: Insufficient Information
MAUDE report 20130641 ↗MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 7 OF 8 UNITS OF PART 79-0002, LOT MM1003065E.
Patient problems: Insufficient Information
MAUDE report 20130639 ↗MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 6 OF 8 UNITS OF PART 79-0002, LOT MM1003065E.
Patient problems: Insufficient Information
MAUDE report 20130638 ↗MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 8 OF 8 UNITS OF PART 79-0002, LOT MM1003065E.
Patient problems: Insufficient Information
MAUDE report 20130637 ↗MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 4 OF 8 UNITS OF PART 79-0002, LOT MM1003065E.
Patient problems: Insufficient Information
MAUDE report 20130633 ↗MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 5 OF 8 UNITS OF PART 79-0002, LOT MM1003065E.
Patient problems: Insufficient Information
MAUDE report 20130632 ↗MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 2 OF 8 UNITS OF PART 79-0002, LOT MM1003065E.
Patient problems: Insufficient Information
MAUDE report 20130631 ↗MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 1 OF 8 UNITS OF PART 79-0002, LOT MM1003065E.
Patient problems: Insufficient Information
MAUDE report 20130630 ↗MalfunctionSep 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S DAYTONA SMALL STATURE SPINAL SYSTEM. DURING THE PROCEDURE THE SURGEON HAD ISSUES WITH THE LOCKING CAPS THREADING ONTO THE SCREWS, RESULTING IN A 30-45 MINUTE SURGICAL DELAY. NO PATIENT INJURY WAS REPORTED. 13 UNITS OF PART 79-0002 AND 3 UNITS OF PART 76-0030 WERE RETURNED. SEE RELATED REPORTS. THIS REPORT IS FOR 3 OF 8 UNITS OF PART 79-0002, LOT MM1003065E.
Patient problems: Insufficient Information
MAUDE report 20130628 ↗
Other Thoracolumbosacral pedicle screw system devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page