buddy lite AC Blood/Fluid Warmer 905-00022Recalls & reported problems

Belmont Instrument Corporation · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10896128002449), received between May 18, 2019 and May 14, 2021 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Smoking” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Smoking2
  • Sparking2
  • Electrical Shorting1
  • Excessive Heating1
  • Leak/Splash1
  • Temperature Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Information2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2021Smoking; Sparking; Smoking; Sparking

    BELMONT MEDICAL TECHNOLOGIES RECEIVED THE FOLLOWING MEDWATCH REPORT (UF/IMPORTER REPORT # (B)(4)) FROM THE DEPARTMENT OF HEALTH & HUMAN SERVICES: "THE AC ADAPTER OF THE BELMONT BUDDY LITE FLUID WARMER BECAME WET WITH BLOOD. WHILE TRYING TO CLEAN THE DEVICE, THE USER USED CAVICIDE AND SPRAYED THE AC ADAPTER WITHOUT UNPLUGGING IT FROM THE ELECTRICAL OUTLET, CAUSING A SHORT CIRCUIT ON THE POWER SUPPLY THAT GENERATED SPARKS AND SMOKE IN THE OPERATING ROOM (OR). THE OR PROCEDURE WAS ALREADY COMPLETED AT THE TIME OF THE INCIDENT, NO PATIENT OR USER WERE HARMED BY THIS SITUATION. THE AC ADAPTER AND BELMONT BUDDY LITE WERE REMOVED FROM SERVICE. OR LEADERSHIP CONDUCTED A PROCEDURE REVIEW WITH OR PERSONNEL TO REVIEW PROPER CLEANING PROCEDURES OF EQUIPMENT THAT ARE CONTAMINATED WITH BIOLOGICAL FLUIDS."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11829429 ↗
  • MalfunctionApr 5, 2021Leak/Splash; Smoking; Sparking; Electrical Shorting

    THE AC ADAPTER OF THE BELMONT BUDDY LITE FLUID WARMER BECAME WET WITH BLOOD. WHILE TRYING TO CLEAN THE DEVICE, THE USER USED CAVICIDE AND SPRAYED THE AC ADAPTER WITHOUT UNPLUGGING IT FROM THE ELECTRICAL OUTLET, CAUSING A SHORT CIRCUIT ON THE POWER SUPPLY THAT GENERATED SPARKS AND SMOKE IN THE OPERATING ROOM (OR). THE OPERATING ROOM PROCEDURE WAS ALREADY COMPLETED AT THE TIME OF THE INCIDENT, NO PATIENT OR USER WERE HARMED BY THIS SITUATION. THE AC ADAPTER AND BELMONT BUDDY LITE WERE REMOVED FROM SERVICE. OPERATING ROOM LEADERSHIP CONDUCTED A PROCEDURE REVIEW WITH OPERATING ROOM PERSONNEL TO REVIEW PROPER CLEANING PROCEDURES OF EQUIPMENT THAT ARE CONTAMINATED WITH BIOLOGICAL FLUIDS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11612730 ↗
  • MalfunctionJun 14, 2019Temperature Problem; Temperature Problem

    BELMONT'S CLINICAL SPECIALIST RECEIVED A COMPLAINT FROM THE USER FACILITY AND RELAYED THE FOLLOWING REPORT: "THERE WAS ANOTHER FIRE WITH THE BUDDY LIGHT TONIGHT, IN ANOTHER OR. I QUESTIONED THE COMMENT AND RECEIVED BACK ON THE 16TH 'GOOD AFTERNOON, ATRIUM IS PERFORMING A RISK ASSESSMENT ON THE EVENT. I BELIEVE THE UNIT IS WITH GE/BIOMED. I AM AWAITING THEIR REVIEW MYSELF.'

    Patient problems: No Information

    MAUDE report 8701497 ↗
  • MalfunctionMay 18, 2019Excessive Heating; Excessive Heating

    THE USER FACILITY REPORTED THAT THE BUDDY LITE AC CARTRIDGE OVERHEATED.

    Patient problems: No Information

    MAUDE report 8622419 ↗

Other Warmer, Blood, Non-Electromagnetic Radiation devices