English, Rapid Infuser RI-2, 230V, 750 ml/min 903-00039ARecalls & reported problems
Belmont Instrument Corporation · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10896128002777), received between Jun 19, 2019 and Mar 21, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Complete Loss of Power” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Complete Loss of Power1
- Device Sensing Problem1
- Electrical /Electronic Property Problem1
- Excessive Heating1
- Increase in Pressure1
- No Tactile Prompts/Feedback1
- Output Problem1
- Pressure Problem1
Patient problems
- No Known Impact Or Consequence To Patient4
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- No Information1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 21, 2024Pressure Problem; Excessive Heating; Pressure Problem; Excessive Heating
THE USER FACILITY REPORTED OVER TEMPERATURE AND POSSIBLE PRESSURE ERRORS DURING USE OF SURGERY CASE. THE PATIENT INVOLVED IN THE INCIDENT PASSED DUE TO CARDIAC ARREST BUT BELMONT IS NOT BEING NOTED AS CONTRIBUTING TO THE PATIENT DEATH.
Patient problems: Insufficient Information
MAUDE report 18956710 ↗MalfunctionJul 22, 2020Device Sensing Problem; Device Sensing Problem
BELMONT'S DISTRIBUTOR RECEIVED A COMPLAINT FROM THE USER FACILITY AND RELAYED THE FOLLOWING REPORT: "DURING A MAJOR RESUSCITATION THIS MORNING, THE DEVICE FAILED ON STARTUP. IT WENT STRAIGHT TO THE CALIBRATION/SET-UP SCREEN AND THE EXIT SERVICE BUTTON DID NOT RESPOND. THEY REBOOTED THE RI-2 COMPUTER AND CHANGED THE AC POWER SOURCE REPEATEDLY HOWEVER IT DID NOT RESOLVE THE ERROR."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10311705 ↗MalfunctionJul 22, 2020Increase in Pressure; Temperature Problem; Increase in Pressure; Temperature Problem
BELMONT'S DISTRIBUTOR RECEIVED A COMPLAINT FROM THE USER FACILITY AND RELAYED THE FOLLOWING REPORT: "REPEATED INSTANCES OF 'HIGH PRESSURE DETECTED' AND 'TEMPERATURE ALARM' EVEN WHEN NOT BEING USED. OUR ENGINEER HAS ASSESSED IT AND FOUND THE FOLLOWING: CUSTOMER HAS HAD REPEATED TEMPERATURE AND PRESSURE SENSORS, EVEN WHEN THE UNIT HAS NOT BEEN IN USE (SET INSERTED, BUT NOT RUNNING). UNREPORTED UNTIL NOW. UNIT RECENTLY SENT TO GETZ ((B)(6) 2020) FOR ERROR 101 AND PASSED WARRANTY PM WITH NO ISSUES."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10311394 ↗MalfunctionJun 18, 2020Complete Loss of Power; Complete Loss of Power
BELMONT'S DISTRIBUTOR RECEIVED A COMPLAINT FROM THE USER FACILITY AND RELAYED THE FOLLOWING REPORT: "THE LIVER TRANSPLANT SURGERY WAS SCHEDULED ON SATURDAY, MAY 16TH. THE UNIT WAS PREPARED AND THE DISPOSABLES (903-00004, 903-00006, AND 903-00018) WAS ATTACHED AND PRIMED AS USUAL. THE UNIT WAS IN OPERATION AND THE SURGERY WAS GOING WELL UNTIL THIS PROBLEM CAME UP. THE UNIT HAD POWER FAILURE AND IMMEDIATELY SHUT DOWN THE SYSTEM. THE STAFFS TRIED TO RESTART THE SYSTEM, BUT THE POWER WAS COMPLETELY GONE. THE STAFF CONTACTED US THE EMERGENCY ASSISTANCE AND WE INSTALLED THE REPLACEMENT."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10171887 ↗MalfunctionFeb 12, 2020Output Problem; No Tactile Prompts/Feedback; Output Problem; No Tactile Prompts/Feedback
BELMONT'S DISTRIBUTER RECEIVED A COMPLAINT FROM THE USER FACILITY AND RELAYED THE FOLLOWING REPORT. "THE TOUCH SCREEN INTERMITTENTLY STOP FUNCTION. A PATIENT GOT TRANSFUSED IN VARIOUS SPEED IN A PERIOD OF 90 MINUTES BECAUSE OF A BLEEDING AORTA. BUT ON THE LAST PHASE OF THESE 90 MINUTES THEY SCALED IT DOWN TO 50 ML/MIN, AFTER THIS IT CAME AN ERROR (101) THAT WAS POSSIBLE TO OVERRIDE, WHICH THEY DID AND THEN IT WENT UP TO 750 ML/MIN AUTOMATICALLY AND IT WASN'T POSSIBLE TO TURN THE SPEED DOWN. IT WAS CONSTANTLY ON 750 ML/MIN. WHEN THEY TRIED TO PRESS "STOP", THE MACHINE THE STOPPED AS LONG AS THEY HOLD THE BUTTON IN, BUT IT CONTINUED AT 750 ML/MIN WHEN THEY RELEASED THE BUTTON. FROM THERE THEY DIDN'T WASNT TO RISK THE PATIENT LIFE AND THEY HAD TO GET THE OTHER BELMONT THEY HAVE AT THEIR HOSPITAL."
Patient problems: No Clinical Signs, Symptoms or Conditions; No Information
MAUDE report 9703206 ↗MalfunctionJun 19, 2019Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE USER FACILITY REPORTED TO OUR DISTRIBUTOR THAT, DURING A PROCEDURE, THERE WAS A SPARK AND THE UNIT STOPPED WORKING. THE DOCTOR HAD THE STAFF SWITCH TO AN FMS 2000 AND THE OPERATION WAS COMPLETED. THE PATIENT WAS NOT INJURED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8713824 ↗
Other Warmer, Thermal, Infusion Fluid devices
- Level 1 HL-90Icu Medical, Inc. · Warmer, thermal, infusion fluid991 reports
- Level 1 HL-390Icu Medical, Inc. · Warmer, thermal, infusion fluid634 reports
- Level 1 L-70Icu Medical, Inc. · Warmer, thermal, infusion fluid1 recall385 reports
- 3-Spike Disposable Set (1 Piece) 903-00006PBelmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 recall1 death97 reports
- Rapid Infuser RI-2, 120V, 750 ml/min 903-00039Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid132 reports
- Level 1 L-70NIIcu Medical, Inc. · Warmer, thermal, infusion fluid114 reports
- Rapid Infuser RI-2, 120V, 1000ml/min 903-00037Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 death76 reports
- Hyperthermia Pump Procedure Kit 902-00045Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 recall4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page